Comparison of three arm-positioning techniques for minimizing motion artifacts in breast magnetic resonance imaging: a prospective volunteer study

Norikazu Koori, Minami Suzuki, Nobukiyo Yoshida, Shohei Yamamoto, Shin Miyakawa, Kota Sasaki, Hiroki Nosaka, Kenji Yasue, Masato Takahashi, Koichi Hanada, Hajime Ichikawa, Kazuyuki Sumi und 2 weitere

Breast Cancer · 2026-07 · 33 (4) · 853-861

DOI: 10.1007/s12282-026-01860-5 ↗

Kurz gefasst

Vergleichsstudie in der Zeitschrift Breast Cancer, online erschienen am 26.05.2026. Verfasst von Norikazu Koori und 13 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiological Technology, Faculty of Medical Technology, Niigata University of Health and Welfare, 1398 Shimami-cho, Niigata city, Niiga…. Hauptthemen laut MEDLINE: Artifacts, Magnetic Resonance Imaging, Patient Positioning, Breast. Die Arbeit nennt 58 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

BACKGROUND: Breast magnetic resonance imaging (MRI) assesses signal intensity changes across sequential image series (kinetic curves). Motion artifacts are common and often degrade image quality. However, the optimal patient positioning to minimize motion during breast MRI remains unclear. This study aimed to identify the most effective arm-positioning method for reducing motion artifacts during breast MRI. METHODS: Six female volunteers underwent MRI at seven time points using enhanced T1 high-resolution isotropic volume excitation, a three-dimensional T1-weighted gradient-echo sequence with fat suppression. Each phase lasted 60 s and comprised seven serial images. Three arm-positions were tested: both-arms-raised, right-arm-raised, and both-arms-down. Motion was assessed using the visual evaluation score (VES) and the distance discordance metric (DDM). RESULTS: VES results were higher with both-arms-raised (2.52 ± 0.43; median: 2.60) than with arms-down (2.39 ± 0.61; median: 2.40) or the right-arm-raised (2.33 ± 0.57; median: 2.40). A significant difference in VES was observed between the both-arms-raised and right-arm-raised groups (P = .04). DDM values were lowest with both-arms-raised (1.02 ± 0.64; median: 0.85), followed by arm-down (1.89 ± 1.19; median: 1.50) and right-arm raised (1.65 ± 0.98; median: 1.56). Significant differences in DDM were observed between the both-arms-raised position and the other two positions (P < .05 and P < .001, respectively). CONCLUSION: Positioning patients with both-arms-raised effectively minimized motion-related artifacts during breast MRI. Conversely, the right-arm-raised position was associated with increased artifacts and should be avoided to maintain image quality.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

ArtifactsMagnetic Resonance ImagingPatient PositioningBreastHumansFemaleProspective StudiesAdultMotionArm

Stichwörter der Autor:innen

Breast MRIDistance discordance metricMagnetic resonance imagingMotion artifactsPositioning method

Autor:innen und Institutionen

  • Department of Radiological Technology, Faculty of Medical Technology, Niigata University of Health and Welfare, 1398 Shimami-cho, Niigata city, Niigata, 950-3198, Japan. Norikazu-koori@nuhw.ac.jp.
  • Department of Radiology, Tsuchiura Kyodo General Hospital, 4-1-1 Otsuno, Tsuchiura, Ibaraki, 300-0028, Japan.
  • Department of Radiological Technology, Faculty of Medical Technology, Niigata University of Health and Welfare, 1398 Shimami-cho, Niigata city, Niigata, 950-3198, Japan.
  • Department of Radiological Sciences, Ibaraki Prefectural University of Health Sciences, 4669-2, Ami, Ibaraki, 300-0394, Japan.
  • Department of Radiology, Fujita Health University Bantane Hospital, Nagoya, Aichi, 454-8509, Japan.

Zitieren

APA

Koori, N., Suzuki, M., Yoshida, N., Yamamoto, S., Miyakawa, S., Sasaki, K., Nosaka, H., Yasue, K., Takahashi, M., Hanada, K., Ichikawa, H., Sumi, K., Takeda, K., & Fuse, H. (2026). Comparison of three arm-positioning techniques for minimizing motion artifacts in breast magnetic resonance imaging: a prospective volunteer study. Breast Cancer, 33(4), 853-861. https://doi.org/10.1007/s12282-026-01860-5

Vancouver

Koori N, Suzuki M, Yoshida N, Yamamoto S, Miyakawa S, Sasaki K, et al. Comparison of three arm-positioning techniques for minimizing motion artifacts in breast magnetic resonance imaging: a prospective volunteer study. Breast Cancer. 2026;33(4):853-861. doi:10.1007/s12282-026-01860-5

BibTeX
@article{Koori2026,
  author = {Koori, Norikazu and Suzuki, Minami and Yoshida, Nobukiyo and Yamamoto, Shohei and Miyakawa, Shin and Sasaki, Kota and Nosaka, Hiroki and Yasue, Kenji and Takahashi, Masato and Hanada, Koichi and Ichikawa, Hajime and Sumi, Kazuyuki and Takeda, Kazuya and Fuse, Hiraku},
  title = {Comparison of three arm-positioning techniques for minimizing motion artifacts in breast magnetic resonance imaging: a prospective volunteer study},
  journal = {Breast Cancer},
  year = {2026},
  volume = {33},
  number = {4},
  pages = {853--861},
  doi = {10.1007/s12282-026-01860-5}
}
Kostenlos starten →