A novel internal target volume definition based on velocity and time of respiratory target motion for external beam radiotherapy
Radiological Physics and Technology · 2024
DOI: 10.1007/s12194-024-00837-3 ↗Kurz gefasst
Fachartikel in der Zeitschrift Radiological Physics and Technology, 2024. Verfasst von Masashi Yamanaka und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Medical Physics, Shonan Kamakura General Hospital, 1370-1 Okamoto, Kamakura City, Kanagawa, 247-8533, Japan. Hauptthemen laut MEDLINE: Four-Dimensional Computed Tomography, Phantoms, Imaging, Respiration, Lung Neoplasms. Die Arbeit nennt 39 Literaturangaben. Gefördert von: AMED.
Abstract (Originalsprache)
This study aimed to develop a novel internal target volume (ITV) definition for respiratory motion targets, considering target motion velocity and time. The proposed ITV was evaluated in respiratory-gated radiotherapy. An ITV modified with target motion velocity and time (ITVvt) was defined as an ITV that includes a target motion based on target motion velocity and time. The target motion velocity was calculated using four-dimensional computed tomography (4DCT) images. The ITVvts were created from phantom and clinical 4DCT images. The phantom 4DCT images were acquired using a solid phantom that moved with a sinusoidal waveform (peak-to-peak amplitudes of 10 and 20 mm and cycles of 2-6 s). The clinical 4DCT images were obtained from eight lung cancer cases. In respiratory-gated radiotherapy, the ITVvt was compared with conventional ITVs for beam times of 0.5-2 s within the gating window. The conventional ITV was created by adding a uniform margin as the maximum motion within the gating window. In the phantom images, the maximum volume difference between the ITVvt and conventional ITV was -81.9%. In the clinical images, the maximum volume difference was -53.6%. Shorter respiratory cycles and longer BTs resulted in smaller ITVvt compared with the conventional ITV. Therefore, the proposed ITVvt plan could be used to reduce treatment volumes and doses to normal tissues.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Medical Physics, Shonan Kamakura General Hospital, 1370-1 Okamoto, Kamakura City, Kanagawa, 247-8533, Japan.
- Department of Medical Physics and Engineering, Osaka University Graduate School of Medicine, 1-7 Yamadaoka, Suita-Shi, Osaka, 565-0871, Japan.
- Department of Medical Physics and Engineering, Osaka University Graduate School of Medicine, 1-7 Yamadaoka, Suita-Shi, Osaka, 565-0871, Japan. nishio@sahs.med.osaka-u.ac.jp.
- Mizuho Research & Technologies, Ltd., 2-3, Kanda-Nishikicho, Chiyoda-Ku, Tokyo, 101-8443, Japan.
Förderung
- AMED
Zitieren
APA
Yamanaka, M., Nishio, T., Iwabuchi, K., & Nagata, H. (2024). A novel internal target volume definition based on velocity and time of respiratory target motion for external beam radiotherapy. Radiological Physics and Technology. https://doi.org/10.1007/s12194-024-00837-3
Vancouver
Yamanaka M, Nishio T, Iwabuchi K, Nagata H. A novel internal target volume definition based on velocity and time of respiratory target motion for external beam radiotherapy. Radiological Physics and Technology. 2024. doi:10.1007/s12194-024-00837-3
BibTeX
@article{Yamanaka2024,
author = {Yamanaka, Masashi and Nishio, Teiji and Iwabuchi, Kohei and Nagata, Hironori},
title = {A novel internal target volume definition based on velocity and time of respiratory target motion for external beam radiotherapy},
journal = {Radiological Physics and Technology},
year = {2024},
doi = {10.1007/s12194-024-00837-3}
}
EUVIMED