Device-related adverse event reports in robotic hernia repair: a disproportionality analysis of the FDA MAUDE database

Yiping Chen, Yuhai Hu, Chunzhong Lin

Journal of Robotic Surgery · 06.11.2026 · 20 (1)

DOI: 10.1007/s11701-026-04159-x ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Robotic Surgery, online erschienen am 06.11.2026. Verfasst von Yiping Chen und 2 weiteren Autor:innen. Beteiligte Einrichtung: Department of Hepatopancreatobiliary Surgery, the First Affiliated Hospital, Fujian Medical University, Fuzhou, 350005, China. cypgg@fjmu.edu.cn. Hauptthemen laut MEDLINE: Robotic Surgical Procedures, Herniorrhaphy, Postoperative Complications. Die Arbeit nennt 45 Literaturangaben. Gefördert von: Natural Science Foundation of Fujian Province.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Robotic hernia repair has spread without demonstrated superiority, and its device-related adverse event reports are unstudied. We compared them with reports from other robotic procedures, prespecifying cholecystectomy as the primary comparator. Using complete US Food and Drug Administration (FDA) Manufacturer and User Facility Device Experience files (cutoff 8 October 2026), we identified operations by prespecified narrative rules and merged 2022-2026 reports into events. Reporting odds ratios (ROR) with 95% confidence intervals (CI) covered 18 reporter-coded outcomes, three of them hypotheses: jaw entrapment (clip appliers excluded), fragmentation or detachment, and tissue injury. Sensitivity and post hoc analysis varied comparator, device category and event merging; two surgeons double-coded 50 reports. We analyzed 3,686 hernia repair and 3,010 cholecystectomy events. Hernia repair reports more often described entrapment (7.6% vs. 1.9%; ROR 4.19, 95% CI 2.87-6.10), fragmentation (17.0% vs. 10.4%; 1.75, 1.52-2.03) and tissue injury (5.1% vs. 2.8%; 1.87, 1.44-2.43). Against colorectal surgery, none was a signal. Within device categories, entrapment remained elevated (pooled ROR 2.44, 1.64-3.65), whereas fragmentation reversed (0.61, 0.48-0.77) and tissue injury varied by category (1.23, 0.92-1.64). Estimates were stable across event-merging rules. All 34 sampled cohort reports were confirmed; estimated cohort sensitivity was 94.0%. Relative to cholecystectomy, robotic hernia repair reports were enriched for jaw entrapment, which persisted within device categories but not against colorectal surgery; fragmentation and tissue-injury signals depended on device category. These hypothesis-generating findings describe report composition, not incidence, clinical risk or device causation.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Robotic Surgical ProceduresHerniorrhaphyPostoperative ComplicationsUnited StatesHumansUnited States Food and Drug AdministrationDatabases, FactualCholecystectomy

Stichwörter der Autor:innen

Adverse event reporting systemsCholecystectomyEquipment failureHerniorrhaphyPharmacovigilanceRobotic surgical procedures

Autor:innen und Institutionen

  • Department of Hepatopancreatobiliary Surgery, the First Affiliated Hospital, Fujian Medical University, Fuzhou, 350005, China. cypgg@fjmu.edu.cn.
  • Department of Hepatopancreatobiliary Surgery, Binhai Campus of the First Affiliated Hospital, National Regional Medical Center, Fujian Medical University, Fuzhou, 350212, China. cypgg@fjmu.edu.cn.
  • Department of Hepatopancreatobiliary Surgery, the First Affiliated Hospital, Fujian Medical University, Fuzhou, 350005, China.
  • Department of Hepatopancreatobiliary Surgery, Binhai Campus of the First Affiliated Hospital, National Regional Medical Center, Fujian Medical University, Fuzhou, 350212, China.

Förderung

  • Natural Science Foundation of Fujian Province

Zitieren

APA

Chen, Y., Hu, Y., & Lin, C. (2026). Device-related adverse event reports in robotic hernia repair: a disproportionality analysis of the FDA MAUDE database. Journal of Robotic Surgery, 20(1). https://doi.org/10.1007/s11701-026-04159-x

Vancouver

Chen Y, Hu Y, Lin C. Device-related adverse event reports in robotic hernia repair: a disproportionality analysis of the FDA MAUDE database. Journal of Robotic Surgery. 2026;20(1). doi:10.1007/s11701-026-04159-x

BibTeX
@article{Chen2026,
  author = {Chen, Yiping and Hu, Yuhai and Lin, Chunzhong},
  title = {Device-related adverse event reports in robotic hernia repair: a disproportionality analysis of the FDA MAUDE database},
  journal = {Journal of Robotic Surgery},
  year = {2026},
  volume = {20},
  number = {1},
  doi = {10.1007/s11701-026-04159-x}
}
Kostenlos starten →