Analysis of root causes of adverse events with Mazor X and Mazor X stealth edition-guided spine surgeries: an FDA MAUDE database study
Analyse der Ursachen von unerwünschten Ereignissen bei mit Mazor X und Mazor X Stealth Edition geführten Wirbelsäulenoperationen: eine FDA-MAUDE-Datenbankstudie
European Spine Journal · 2026-08 · 35 (8) · 4797-4806
DOI: 10.1007/s00586-025-09683-2 ↗Auf einen Blick
In 330 in MAUDE gemeldeten Ereignissen bei Mazor X und Mazor X Stealth Edition waren externe Ursachen am häufigsten; am häufigsten wurde Anwenderfehler genannt. Die Berichte deuten auf unterschiedliche Patient:innenfolgen hin.
Fragestellung
Welche Ursachen, Wirbelverteilung und Patient:innenfolgen sind bei unerwünschten Ereignissen mit Mazor X und Mazor X Stealth Edition beschrieben?
Methode
Analyse von 330 unerwünschten Ereignissen aus der MAUDE-Datenbank für Mazor X und Mazor X Stealth Edition; Zuordnung nach Ursachen, Wirbelverteilung, chirurgischer Verzögerung, Roboterausfall und Patient:innenfolgen.
Ergebnisse
Externe Ursachen waren am häufigsten (MX 35 %, MXSE 52 %), meist Anwenderfehler. Abweichungen betrafen vor allem L3, L4 und L5; sie verliefen meist medial oder lateral. Bei MX waren motorische Defizite, bei MXSE sensible Defizite am häufigsten; bei MX gab es 2 Todesfälle.
Einschränkungen
Die MAUDE-Datenbank erlaubt keine Berechnung von Inzidenz oder Prävalenz.
Schlussfolgerung der Autor:innen
Die MAUDE-Berichte deuten darauf hin, dass anwenderbezogene Fehler den Großteil der unerwünschten Ereignisse ausmachen; die Autor:innen betonen die Bedeutung gezielter Chirurg:innen-Schulung.
Kernaussagen
- Externe Ursachen waren am häufigsten; Anwenderfehler dominierten.
- L3, L4 und L5 waren am häufigsten betroffen.
- MX: motorische Defizite und 2 Todesfälle; MXSE: sensorische Defizite.
Deutsche Zusammenfassung: KI-gestützt aus dem Original-Abstract erstellt; alle Zahlen wurden automatisch mit dem Original abgeglichen. Maßgeblich ist der Original-Abstract.
Kurz gefasst
Fachartikel in der Zeitschrift European Spine Journal, online erschienen am 22.12.2025. Verfasst von Apratim Maity und 6 weiteren Autor:innen. Beteiligte Einrichtung: Donald and Barbara Zucker School of Medicine at Hofstra/Northwell, Lake Success, USA. amaity@northwell.edu. Hauptthemen laut MEDLINE: Robotic Surgical Procedures, Spine, Postoperative Complications, Spinal Fusion. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 27 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Themen: Patientensicherheit
Abstract (Originalsprache)
Purpose
This study aimed to analyze adverse events associated with Mazor X surgical guidance robots, focusing on parameters such as root causes, vertebral distribution, surgical delay, robot abandonment, and more.
Methods
330 adverse events reported within the Manufacturer and User Facility Device Experience (MAUDE) database for the Mazor X (MX) and Mazor X Stealth Edition (MXSE) robot guidance systems were analyzed and categorized based on the root causes, vertebral distribution, surgical delay, robot abandonment, and patient outcomes.
Results
Both the MX and MXSE adverse events were primarily due to external root causes (35% and 52%, respectively), with user error being the most common. Deviations were most likely to affect the L3, L4, and L5 vertebra, and deviate either medially or laterally. Motor deficits were the most common adverse patient outcome for MX, while sensory deficits were the most common for MXSE. There were two deaths associated with the MX, compared to zero deaths with the MXSE.
Conclusion
While the MAUDE database cannot be used to calculate the incidence or prevalence of adverse events, it provides insight into the types of adverse events that tend to occur with medical devices. MAUDE reports suggest that user-related errors account for most of the adverse events that occur, underscoring the need for focused surgeon training. Our study highlights the importance of understanding both the types of adverse events and their root causes to promote patient safety and reduce complication rates.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Donald and Barbara Zucker School of Medicine at Hofstra/Northwell, Lake Success, USA. amaity@northwell.edu.
- Donald and Barbara Zucker School of Medicine at Hofstra/Northwell, Lake Success, USA.
- Donald and Barbara Zucker School of Medicine at Hofstra/Northwell, Lake Success, USA. eorsino1@pride.hofstra.edu.
Zitieren
APA
Maity, A., Brown, E. D. L., Ward, M., Orsino, E. A., Schneider, D. M., Sciubba, D. M., & Lo, S. L. (2026). Analysis of root causes of adverse events with Mazor X and Mazor X stealth edition-guided spine surgeries: an FDA MAUDE database study. European Spine Journal, 35(8), 4797-4806. https://doi.org/10.1007/s00586-025-09683-2
Vancouver
Maity A, Brown EDL, Ward M, Orsino EA, Schneider DM, Sciubba DM, et al. Analysis of root causes of adverse events with Mazor X and Mazor X stealth edition-guided spine surgeries: an FDA MAUDE database study. European Spine Journal. 2026;35(8):4797-4806. doi:10.1007/s00586-025-09683-2
BibTeX
@article{Maity2026,
author = {Maity, Apratim and Brown, Ethan D L and Ward, Max and Orsino, Emily A and Schneider, Daniel M and Sciubba, Daniel M and Lo, Sheng-Fu Larry},
title = {Analysis of root causes of adverse events with Mazor X and Mazor X stealth edition-guided spine surgeries: an FDA MAUDE database study},
journal = {European Spine Journal},
year = {2026},
volume = {35},
number = {8},
pages = {4797--4806},
doi = {10.1007/s00586-025-09683-2}
}