Adverse events associated with robotic-assistance in total hip arthroplasty: an analysis based on the FDA MAUDE database
Unerwünschte Ereignisse bei roboterassistierter Totalendoprothese der Hüfte: Analyse auf Basis der FDA-MAUDE-Datenbank
HIP International · 2024-11 · 34 (6) · 688-694
DOI: 10.1177/11207000241263315 ↗Auf einen Blick
In der MAUDE-Datenbank wurden bei roboterassistierter THA häufig unerwünschte Ereignisse berichtet, am häufigsten ein intraoperativer Hardware-Defekt. Auch chirurgische Verzögerungen kamen vor.
Fragestellung
Wie variieren unerwünschte Ereignisse bei roboterassistierter THA in der FDA-MAUDE-Datenbank?
Methode
Analyse von Medical Device Reports in der MAUDE-Datenbank zwischen 2017 und 2021; Auswertung nach Art und Konsequenzen der Ereignisse.
Ergebnisse
521 MDRs mit 546 Ereignissen. Häufigstes Ereignis: intraoperativer Hardware-Defekt (304/546, 55,7 %), am häufigsten gebrochener Impaction Handle/Platform (110, 20,1 %). Fehlplatzierung der Pfanne: 63 (11,5 %). Roboterabbruch: 71/521 (13,0 %). OP-Verzögerung: 146/521 (28 %) mit mittlerer Verzögerung 17,9 Minuten (1–60).
Schlussfolgerung der Autor:innen
Die Identifikation möglicher Komplikationen bei roboterassistierter THA ist ein wichtiger erster Schritt zur Vermeidung unerwünschter Ereignisse und OP-Verzögerungen. Die Datenbankanalyse gibt einen Überblick über das Spektrum der Komplikationen.
Kernaussagen
- 521 MDRs mit 546 Ereignissen in der MAUDE-Datenbank
- Häufigstes Ereignis: intraoperativer Hardware-Defekt
- OP-Verzögerungen und Roboterabbruch wurden berichtet
Deutsche Zusammenfassung: KI-gestützt aus dem Original-Abstract erstellt; alle Zahlen wurden automatisch mit dem Original abgeglichen. Maßgeblich ist der Original-Abstract.
Kurz gefasst
Fachartikel in der Zeitschrift HIP International, online erschienen am 04.08.2024. Verfasst von S Bradley Graefe und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Orthopaedics and Rehabilitation, Penn State College of Medicine, Hershey, PA, USA. Hauptthemen laut MEDLINE: Arthroplasty, Replacement, Hip, Robotic Surgical Procedures, Databases, Factual, United States Food and Drug Administration. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 25 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Themen: Patientensicherheit
Abstract (Originalsprache)
Background
The Food and Drug Administration (FDA) maintains the Manufacturer and User Facility Device Experience (MAUDE) database for reporting adverse events associated with medical devices, including emerging technologies, such as robotic-assisted total hip arthroplasty (THA). Aim of this study was to evaluate the variation of adverse events associated with robotics in THA.
Methods
Medical device reports (MDRs) within the MAUDE database were identified between 2017 and 2021. For MDR identification the product class “orthopaedic stereotaxic equipment” and terms associated with THA were used. Individual adverse events were identified and organised by type and consequences, such as patient injury, surgical delay, or conversion to the manual technique.
Results
521 MDRs constituting 546 discrete events were found. The most common reported complication was intraoperative hardware failure (304/546, 55.7%), among which the most common failure was a broken impaction handle/platform (110, 20.1%). Inaccurate cup placement was the second most common reported complication (63, 11.5%). Abandoning the robot occurred in 13.0% (71/521) of reports. A surgical delay was noted in 28% (146/521) of reports, with an average delay of 17.9 (range 1–60) minutes.
Conclusions
Identifying complications that may occur with robotics in THA is an important first step in preventing adverse events and surgical delays. Database analysis provide an overview of the range of complications.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Orthopaedics and Rehabilitation, Penn State College of Medicine, Hershey, PA, USA.
ORCID-Profile: Hannah H Nam, Mark L Dunleavy
Zitieren
APA
Graefe, S. B., Kirchner, G. J., Pahapill, N. K., Nam, H. H., Dunleavy, M. L., & Haines, N. (2024). Adverse events associated with robotic-assistance in total hip arthroplasty: an analysis based on the FDA MAUDE database. HIP International, 34(6), 688-694. https://doi.org/10.1177/11207000241263315
Vancouver
Graefe SB, Kirchner GJ, Pahapill NK, Nam HH, Dunleavy ML, Haines N. Adverse events associated with robotic-assistance in total hip arthroplasty: an analysis based on the FDA MAUDE database. HIP International. 2024;34(6):688-694. doi:10.1177/11207000241263315
BibTeX
@article{Graefe2024,
author = {Graefe, S Bradley and Kirchner, Gregory J and Pahapill, Natalie K and Nam, Hannah H and Dunleavy, Mark L and Haines, Nikkole},
title = {Adverse events associated with robotic-assistance in total hip arthroplasty: an analysis based on the FDA MAUDE database},
journal = {HIP International},
year = {2024},
volume = {34},
number = {6},
pages = {688--694},
doi = {10.1177/11207000241263315}
}