Semi-automatic robotic puncture system based on deformable soft tissue point cloud registration
International Journal of Computer Assisted Radiology and Surgery · 2024
DOI: 10.1007/s11548-024-03247-3 ↗Kurz gefasst
Fachartikel in der Zeitschrift International Journal of Computer Assisted Radiology and Surgery, 2024. Verfasst von Bo Zhang und 9 weiteren Autor:innen. Beteiligte Einrichtung: WuXi AMIT Intelligent Medical Technology Col., Ltd., Wuxi, 214000, China. Hauptthemen laut MEDLINE: Robotic Surgical Procedures, Punctures, Liver, Surgery, Computer-Assisted. Die Arbeit wurde bisher 6-mal zitiert (Crossref) und nennt 17 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Purpose
Traditional surgical puncture robot systems based on computed tomography (CT) and infrared camera guidance have natural disadvantages for puncture of deformable soft tissues such as the liver. Liver movement and deformation caused by breathing are difficult to accurately assess and compensate by current technical solutions. We propose a semi-automatic robotic puncture system based on real-time ultrasound images to solve this problem.
Method
Real-time ultrasound images and their spatial position information can be obtained by robot in this system. By recognizing target tissue in these ultrasound images and using reconstruction algorithm, 3D real-time ultrasound tissue point cloud can be constructed. Point cloud of the target tissue in the CT image can be obtained by using developed software. Through the point cloud registration method based on feature points, two point clouds above are registered. The puncture target will be automatically positioned, then robot quickly carries the puncture guide mechanism to the puncture site and guides the puncture. It takes about just tens of seconds from the start of image acquisition to completion of needle insertion. Patient can be controlled by a ventilator to temporarily stop breathing, and patient’s breathing state does not need to be the same as taking CT scan.
Results
The average operation time of 24 phantom experiments is 64.5 s, and the average error between the needle tip and the target point after puncture is 0.8 mm. Two animal puncture surgeries were performed, and the results indicated that the puncture errors of these two experiments are 1.76 mm and 1.81 mm, respectively.
Conclusion
Robot system can effectively carry out and implement liver tissue puncture surgery, and the success rate of phantom experiments and experiments is 100%. It also shows that the puncture robot system has high puncture accuracy, short operation time, and great clinical value.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- WuXi AMIT Intelligent Medical Technology Col., Ltd., Wuxi, 214000, China.
- Future Robotics Organization, Waseda University, Tokyo, 1620044, Japan.
- WuXi AMIT Intelligent Medical Technology Col., Ltd., Wuxi, 214000, China. chenkui@amit.net.cn.
- The First Affiliated Hospital of Sun Yat-sen University, Guangzhou, 510000, China.
Zitieren
APA
Zhang, B., Chen, K., Yao, Y., Wu, B., Li, Q., Zhang, Z., Fan, P., Wang, W., Lin, M., & Fujie, M. G. (2024). Semi-automatic robotic puncture system based on deformable soft tissue point cloud registration. International Journal of Computer Assisted Radiology and Surgery. https://doi.org/10.1007/s11548-024-03247-3
Vancouver
Zhang B, Chen K, Yao Y, Wu B, Li Q, Zhang Z, et al. Semi-automatic robotic puncture system based on deformable soft tissue point cloud registration. International Journal of Computer Assisted Radiology and Surgery. 2024. doi:10.1007/s11548-024-03247-3
BibTeX
@article{Zhang2024,
author = {Zhang, Bo and Chen, Kui and Yao, Yuhang and Wu, Bo and Li, Qiang and Zhang, Zheming and Fan, Peihua and Wang, Wei and Lin, Manxia and Fujie, Masakatsu G},
title = {Semi-automatic robotic puncture system based on deformable soft tissue point cloud registration},
journal = {International Journal of Computer Assisted Radiology and Surgery},
year = {2024},
doi = {10.1007/s11548-024-03247-3}
}
EUVIMED