Fachartikel in der Zeitschrift International Journal of Computer Assisted Radiology and Surgery, online erschienen am 05.07.2021. Verfasst von Bogdan Maris und 10 weiteren Autor:innen. Beteiligte Einrichtung: Department of Computer Science, University of Verona, Verona, Italy. bogdan.maris@univr.it. Hauptthemen laut MEDLINE: Software. Die Arbeit wurde bisher 26-mal zitiert (Crossref) und nennt 15 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: ERC-2019-POC.
Abstract (Originalsprache)
Purpose
We present the validation of PROST, a robotic device for prostate biopsy. PROST is designed to minimize human error by introducing some autonomy in the execution of the key steps of the procedure, i.e., target selection, image fusion and needle positioning. The robot allows executing a targeted biopsy through ultrasound (US) guidance and fusion with magnetic resonance (MR) images, where the target was defined.
Methods
PROST is a parallel robot with 4 degrees of freedom (DOF) to orient the needle and 1 DOF to rotate the US probe. We reached a calibration error of less than 2 mm, computed as the difference between the needle positioning in robot coordinates and in the US image. The autonomy of the robot is given by the image analysis software, which employs deep learning techniques, the integrated image fusion algorithms and automatic computation of the needle trajectory. For safety reasons, the insertion of the needle is assigned to the doctor.
Results
System performance was evaluated in terms of positioning accuracy. Tests were performed on a 3D printed object with nine 2-mm spherical targets and on an anatomical commercial phantom that simulates human prostate with three lesions and the surrounding structures. The average accuracy reached in the laboratory experiments was $$ 1.30 \pm 0.44\, \text {mm}$$ 1.30 ± 0.44 mm in the first test and $$ 1.54 \pm 0.34\, \text {mm}$$ 1.54 ± 0.34 mm in the second test.
Conclusions
We introduced a first prototype of a prostate biopsy robot that has the potential to increase the detection of clinically significant prostate cancer and, by including some level of autonomy, to simplify the procedure, to reduce human errors and shorten training time. The use of a robot for the biopsy of the prostate will create the possibility to include also a treatment, such as focal ablation, to be delivered through the same system.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Computer Science, University of Verona, Verona, Italy. bogdan.maris@univr.it.
- Department of Computer Science, University of Verona, Verona, Italy.
- Department of Urology, University of Modena and Reggio Emilia, Modena, Italy.
- Department of Urology, University of Verona, Verona, Italy.
Förderung
- ERC-2019-POC
Zitieren
APA
Maris, B., Tenga, C., Vicario, R., Palladino, L., Murr, N., De Piccoli, M., Calanca, A., Puliatti, S., Micali, S., Tafuri, A., & Fiorini, P. (2021). Toward autonomous robotic prostate biopsy: a pilot study. International Journal of Computer Assisted Radiology and Surgery, 16(8), 1393-1401. https://doi.org/10.1007/s11548-021-02437-7
Vancouver
Maris B, Tenga C, Vicario R, Palladino L, Murr N, De Piccoli M, et al. Toward autonomous robotic prostate biopsy: a pilot study. International Journal of Computer Assisted Radiology and Surgery. 2021;16(8):1393-1401. doi:10.1007/s11548-021-02437-7
BibTeX
@article{Maris2021,
author = {Maris, Bogdan and Tenga, Chiara and Vicario, Rudy and Palladino, Luigi and Murr, Noe and De Piccoli, Michela and Calanca, Andrea and Puliatti, Stefano and Micali, Salvatore and Tafuri, Alessandro and Fiorini, Paolo},
title = {Toward autonomous robotic prostate biopsy: a pilot study},
journal = {International Journal of Computer Assisted Radiology and Surgery},
year = {2021},
volume = {16},
number = {8},
pages = {1393--1401},
doi = {10.1007/s11548-021-02437-7}
}