Phantom‐based assessment of motion and needle targeting accuracy of robotic devices for magnetic resonance imaging‐guided needle biopsy
The International Journal of Medical Robotics and Computer Assisted Surgery · 2023
DOI: 10.1002/rcs.2526 ↗Kurz gefasst
Fachartikel in der Zeitschrift The International Journal of Medical Robotics and Computer Assisted Surgery, 2023. Verfasst von Anastasia Antoniou und 9 weiteren Autor:innen. Beteiligte Einrichtung: Department of Electrical Engineering, Computer Engineering, and Informatics, Cyprus University of Technology, Limassol, Cyprus. Hauptthemen laut MEDLINE: Robotic Surgical Procedures. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 59 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Research and Innovation Foundation, FUSROBOT, SOUNDPET.
Abstract (Originalsprache)
Background
The current study proposes simple methods for assessing the performance of robotic devices intended for Magnetic Resonance Imaging (MRI)‐guided needle biopsy.
Methods
In‐house made agar‐based breast phantoms containing biopsy targets served as the main tool in the evaluation process of an MRI compatible positioning device comprising a needle navigator. The motion accuracy of mechanical stages was assessed by calliper measurements. Laboratory evaluation of needle targeting included a repeatability phantom test and a laser‐based method. The accuracy and repeatability of needle targeting was also assessed by MRI.
Results
The maximum error of linear motion for steps up to 10 mm was 0.1 mm. Needle navigation relative to the phantom and alignment with the various biopsy targets were performed successfully in both the laboratory and MRI settings. The proposed biopsy phantoms offered tissue‐like signal in MRI and good haptic feedback during needle insertion.
Conclusions
The proposed methods could be valuable in the process of validating the accuracy of MRI‐guided biopsy robotic devices in both laboratory and real environments.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Electrical Engineering, Computer Engineering, and Informatics, Cyprus University of Technology, Limassol, Cyprus.
- R&D, Medsonic Ltd, Limassol, Cyprus.
- Department of Interventional Radiology, German Oncology Center, Limassol, Cyprus.
ORCID-Profile: Nikolas Evripidou, Marinos Giannakou, Christakis Damianou
Förderung
- Research and Innovation Foundation
- FUSROBOT
- SOUNDPET
- MRBREASTBIO
Zitieren
APA
Antoniou, A., Nikolaou, A., Evripidou, N., Georgiou, A., Filippou, A., Zinonos, V., Giannakou, M., Chrysanthou, A., Ioannides, C., & Damianou, C. (2023). Phantom‐based assessment of motion and needle targeting accuracy of robotic devices for magnetic resonance imaging‐guided needle biopsy. The International Journal of Medical Robotics and Computer Assisted Surgery. https://doi.org/10.1002/rcs.2526
Vancouver
Antoniou A, Nikolaou A, Evripidou N, Georgiou A, Filippou A, Zinonos V, et al. Phantom‐based assessment of motion and needle targeting accuracy of robotic devices for magnetic resonance imaging‐guided needle biopsy. The International Journal of Medical Robotics and Computer Assisted Surgery. 2023. doi:10.1002/rcs.2526
BibTeX
@article{Antoniou2023,
author = {Antoniou, Anastasia and Nikolaou, Anastasia and Evripidou, Nikolas and Georgiou, Andreas and Filippou, Antria and Zinonos, Vasiliki and Giannakou, Marinos and Chrysanthou, Antreas and Ioannides, Cleanthis and Damianou, Christakis},
title = {Phantom‐based assessment of motion and needle targeting accuracy of robotic devices for magnetic resonance imaging‐guided needle biopsy},
journal = {The International Journal of Medical Robotics and Computer Assisted Surgery},
year = {2023},
doi = {10.1002/rcs.2526}
}
EUVIMED