EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Phantom‐based assessment of motion and needle targeting accuracy of robotic devices for magnetic resonance imaging‐guided needle biopsy

Anastasia Antoniou, Anastasia Nikolaou, Nikolas Evripidou iD, Andreas Georgiou, Antria Filippou, Vasiliki Zinonos, Marinos Giannakou iD, Antreas Chrysanthou, Cleanthis Ioannides, Christakis Damianou iD

The International Journal of Medical Robotics and Computer Assisted Surgery · 2023

DOI: 10.1002/rcs.2526 ↗

Kurz gefasst

Fachartikel in der Zeitschrift The International Journal of Medical Robotics and Computer Assisted Surgery, 2023. Verfasst von Anastasia Antoniou und 9 weiteren Autor:innen. Beteiligte Einrichtung: Department of Electrical Engineering, Computer Engineering, and Informatics, Cyprus University of Technology, Limassol, Cyprus. Hauptthemen laut MEDLINE: Robotic Surgical Procedures. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 59 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Research and Innovation Foundation, FUSROBOT, SOUNDPET.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

The current study proposes simple methods for assessing the performance of robotic devices intended for Magnetic Resonance Imaging (MRI)‐guided needle biopsy.

Methods

In‐house made agar‐based breast phantoms containing biopsy targets served as the main tool in the evaluation process of an MRI compatible positioning device comprising a needle navigator. The motion accuracy of mechanical stages was assessed by calliper measurements. Laboratory evaluation of needle targeting included a repeatability phantom test and a laser‐based method. The accuracy and repeatability of needle targeting was also assessed by MRI.

Results

The maximum error of linear motion for steps up to 10 mm was 0.1 mm. Needle navigation relative to the phantom and alignment with the various biopsy targets were performed successfully in both the laboratory and MRI settings. The proposed biopsy phantoms offered tissue‐like signal in MRI and good haptic feedback during needle insertion.

Conclusions

The proposed methods could be valuable in the process of validating the accuracy of MRI‐guided biopsy robotic devices in both laboratory and real environments.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Robotic Surgical ProceduresHumansImage-Guided BiopsyNeedlesPhantoms, ImagingMagnetic Resonance ImagingBiopsy, Needle

Stichwörter der Autor:innen

MRI-guidedagar phantomsbreast biopsymotion accuracyneedle targetingrobotic device

Autor:innen und Institutionen

  • Department of Electrical Engineering, Computer Engineering, and Informatics, Cyprus University of Technology, Limassol, Cyprus.
  • R&D, Medsonic Ltd, Limassol, Cyprus.
  • Department of Interventional Radiology, German Oncology Center, Limassol, Cyprus.

ORCID-Profile: Nikolas Evripidou, Marinos Giannakou, Christakis Damianou

Förderung

  • Research and Innovation Foundation
  • FUSROBOT
  • SOUNDPET
  • MRBREASTBIO

Zitieren

APA

Antoniou, A., Nikolaou, A., Evripidou, N., Georgiou, A., Filippou, A., Zinonos, V., Giannakou, M., Chrysanthou, A., Ioannides, C., & Damianou, C. (2023). Phantom‐based assessment of motion and needle targeting accuracy of robotic devices for magnetic resonance imaging‐guided needle biopsy. The International Journal of Medical Robotics and Computer Assisted Surgery. https://doi.org/10.1002/rcs.2526

Vancouver

Antoniou A, Nikolaou A, Evripidou N, Georgiou A, Filippou A, Zinonos V, et al. Phantom‐based assessment of motion and needle targeting accuracy of robotic devices for magnetic resonance imaging‐guided needle biopsy. The International Journal of Medical Robotics and Computer Assisted Surgery. 2023. doi:10.1002/rcs.2526

BibTeX
@article{Antoniou2023,
  author = {Antoniou, Anastasia and Nikolaou, Anastasia and Evripidou, Nikolas and Georgiou, Andreas and Filippou, Antria and Zinonos, Vasiliki and Giannakou, Marinos and Chrysanthou, Antreas and Ioannides, Cleanthis and Damianou, Christakis},
  title = {Phantom‐based assessment of motion and needle targeting accuracy of robotic devices for magnetic resonance imaging‐guided needle biopsy},
  journal = {The International Journal of Medical Robotics and Computer Assisted Surgery},
  year = {2023},
  doi = {10.1002/rcs.2526}
}
Kostenlos starten →