Deep learning-based segmentation of ultra-low-dose CT images using an optimized nnU-Net model

Yazdan Salimi, Zahra Mansouri, Chang Sun, Amirhossein Sanaat, Mohammadhossein Yazdanpanah, Hossein Shooli, René Nkoulou, Sana Boudabbous, Habib Zaidi

La radiologia medica · 18.03.2025 · 130 (5) · 723-739

DOI: 10.1007/s11547-025-01989-x ↗

Kurz gefasst

Fachartikel in der Zeitschrift La radiologia medica, online erschienen am 18.03.2025. Verfasst von Yazdan Salimi und 8 weiteren Autor:innen. Beteiligte Einrichtung: Division of Nuclear Medicine and Molecular Imaging, Geneva University Hospital, CH-1211, Geneva, Switzerland. Hauptthemen laut MEDLINE: Deep Learning, Tomography, X-Ray Computed, Radiation Dosage, Image Processing, Computer-Assisted. Die Arbeit wurde bisher 22-mal zitiert (Crossref) und nennt 46 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: H2020 Euratom, University of Geneva.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

Low-dose CT protocols are widely used for emergency imaging, follow-ups, and attenuation correction in hybrid PET/CT and SPECT/CT imaging. However, low-dose CT images often suffer from reduced quality depending on acquisition and patient attenuation parameters. Deep learning (DL)-based organ segmentation models are typically trained on high-quality images, with limited dedicated models for noisy CT images. This study aimed to develop a DL pipeline for organ segmentation on ultra-low-dose CT images. Materials and methods 274 CT raw datasets were reconstructed using Siemens ReconCT software with ADMIRE iterative algorithm, generating full-dose (FD-CT) and simulated low-dose (LD-CT) images at 1%, 2%, 5%, and 10% of the original tube current. Existing FD-nnU-Net models segmented 22 organs on FD-CT images, serving as reference masks for training new LD-nnU-Net models using LD-CT images. Three models were trained for bony tissue (6 organs), soft-tissue (15 organs), and body contour segmentation. The segmented masks from LD-CT were compared to FD-CT as standard of reference. External datasets with actual LD-CT images were also segmented and compared.

Results

FD-nnU-Net performance declined with reduced radiation dose, especially below 10% (5 mAs). LD-nnU-Net achieved average Dice scores of 0.937 ± 0.049 (bony tissues), 0.905 ± 0.117 (soft-tissues), and 0.984 ± 0.023 (body contour). LD models outperformed FD models on external datasets.

Conclusion

Conventional FD-nnU-Net models performed poorly on LD-CT images. Dedicated LD-nnU-Net models demonstrated superior performance across cross-validation and external evaluations, enabling accurate segmentation of ultra-low-dose CT images. The trained models are available on our GitHub page.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Deep LearningTomography, X-Ray ComputedRadiation DosageImage Processing, Computer-AssistedRadiographic Image Interpretation, Computer-AssistedHumansAlgorithms

Stichwörter der Autor:innen

Deep learningOrgan segmentationRadiation doseUltra-low-dose CTnnU-Net

Autor:innen und Institutionen

  • Division of Nuclear Medicine and Molecular Imaging, Geneva University Hospital, CH-1211, Geneva, Switzerland.
  • School of Information and Communication Engineering, Beijing University of Posts and Telecommunications, Beijing, 100876, China.
  • Department of Radiology, Shiraz University of Medical Sciences, Shiraz, Iran.
  • Department of Radiology, Bushehr University of Medical Sciences, Bushehr, Iran.
  • Division of Radiology, Geneva University Hospital, CH-1211, Geneva, Switzerland.
  • Division of Nuclear Medicine and Molecular Imaging, Geneva University Hospital, CH-1211, Geneva, Switzerland. habib.zaidi@hcuge.ch.

Förderung

  • H2020 Euratom
  • University of Geneva

Zitieren

APA

Salimi, Y., Mansouri, Z., Sun, C., Sanaat, A., Yazdanpanah, M., Shooli, H., Nkoulou, R., Boudabbous, S., & Zaidi, H. (2025). Deep learning-based segmentation of ultra-low-dose CT images using an optimized nnU-Net model. La radiologia medica, 130(5), 723-739. https://doi.org/10.1007/s11547-025-01989-x

Vancouver

Salimi Y, Mansouri Z, Sun C, Sanaat A, Yazdanpanah M, Shooli H, et al. Deep learning-based segmentation of ultra-low-dose CT images using an optimized nnU-Net model. La radiologia medica. 2025;130(5):723-739. doi:10.1007/s11547-025-01989-x

BibTeX
@article{Salimi2025,
  author = {Salimi, Yazdan and Mansouri, Zahra and Sun, Chang and Sanaat, Amirhossein and Yazdanpanah, Mohammadhossein and Shooli, Hossein and Nkoulou, René and Boudabbous, Sana and Zaidi, Habib},
  title = {Deep learning-based segmentation of ultra-low-dose CT images using an optimized nnU-Net model},
  journal = {La radiologia medica},
  year = {2025},
  volume = {130},
  number = {5},
  pages = {723--739},
  doi = {10.1007/s11547-025-01989-x}
}
Kostenlos starten →