EUVIMEDEUVIMED entdecken
EvaluationsstudieOpen Access · CC BY 4.0

Performance Evaluation of Image Segmentation Using Dual-Energy Spectral CT Images with Deep Learning Image Reconstruction: A Phantom Study

Haoyan Li iD, Zhenpeng Chen iD, Shuaiyi Gao, Jiaqi Hu, Zhihao Yang, Yun Peng, Jihang Sun

Tomography · 2025

DOI: 10.3390/tomography11050051 ↗

Kurz gefasst

Evaluationsstudie in der Zeitschrift Tomography, 2025. Verfasst von Haoyan Li und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiology, Beijing Children's Hospital, Capital Medical University, National Center for Children's Health, No. 56, Nanlishi Road, Xiche…. Hauptthemen laut MEDLINE: Deep Learning, Tomography, X-Ray Computed, Image Processing, Computer-Assisted, Radiographic Image Interpretation, Computer-Assisted. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 55 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Natural Science Foundation of Xinjiang Uygur Autonomous Region., Beijing Municipal Administration of Hospitals Incubating Program, Xinjiang Hospital of Beijing Children' s Hospital science program.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objectives

To evaluate the medical image segmentation performance of monochromatic images in various energy levels.

Methods

The low-density module (25 mm in diameter, 6 Hounsfield Unit (HU) in density difference from background) from the ACR464 phantom was scanned at both 10 mGy and 5 mGy dose levels. Virtual monoenergetic images (VMIs) at different energy levels of 40, 50, 60, 68, 74, and 100 keV were generated. The images at 10 mGy reconstructed with 50% adaptive statistical iterative reconstruction veo (ASIR-V50%) were used to train an image segmentation model based on U-Net. The evaluation set used 5 mGy VMIs reconstructed with various reconstruction algorithms: FBP, ASIR-V50%, ASIR-V100%, deep learning image reconstruction (DLIR) with low (DLIR-L), medium (DLIR-M), and high (DLIR-H) strength levels. U-Net was employed as a tool to compare algorithm performance. Image noise and segmentation metrics, such as the DICE coefficient, intersection over union (IOU), sensitivity, and Hausdorff distance, were calculated to assess both image quality and segmentation performance.

Results

DLIR-M and DLIR-H consistently achieved lower image noise and better segmentation performance, with the highest results observed at 60 keV, and DLIR-H had the lowest image noise across all energy levels. The performance metrics, including IOU, DICE, and sensitivity, were ranked in descending order with energy levels of 60 keV, 68 keV, 50 keV, 74 keV, 40 keV, and 100 keV. Specifically, at 60 keV, the average IOU values for each reconstruction method were 0.60 for FBP, 0.67 for ASIR-V50%, 0.68 for ASIR-V100%, 0.72 for DLIR-L, 0.75 for DLIR-M, and 0.75 for DLIR-H. The average DICE values were 0.75, 0.80, 0.82, 0.83, 0.85, and 0.86. The sensitivity values were 0.93, 0.91, 0.96, 0.95, 0.98, and 0.98.

Conclusions

For low-density, non-enhancing objects under a low dose, the 60 keV VMIs performed better in automatic segmentation. DLIR-M and DLIR-H algorithms delivered the best results, whereas DLIR-H provided the lowest image noise and highest sensitivity.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Deep LearningTomography, X-Ray ComputedImage Processing, Computer-AssistedRadiographic Image Interpretation, Computer-AssistedRadiography, Dual-Energy Scanned ProjectionPhantoms, ImagingHumansAlgorithmsRadiation Dosage

Stichwörter der Autor:innen

U-Net modeldeep learningdual-energy CTimage enhancementimage reconstructionmulti-slice CT

Autor:innen und Institutionen

  • Department of Radiology, Beijing Children's Hospital, Capital Medical University, National Center for Children's Health, No. 56, Nanlishi Road, Xicheng District, Beijing 100045, China.
  • Qingdao Academy of Chinese Medical Sciences, Shandong University of Traditional Chinese Medicine, Qingdao 266112, China.
  • Children's Hospital of Xinjiang Uygur Autonomous Region, Xinjiang Hospital of Beijing Children' s Hospital, Urumqi 830054, China.

ORCID-Profile: Haoyan Li, Zhenpeng Chen

Förderung

  • Natural Science Foundation of Xinjiang Uygur Autonomous Region.
  • Beijing Municipal Administration of Hospitals Incubating Program
  • Xinjiang Hospital of Beijing Children' s Hospital science program
  • Natural Science Foundation of Xinjiang Uygur Autonomous Region
  • State Key Laboratory of Advanced Medical Materials and Devices, ShanghaiTech University
  • Xinjiang Hospital of Beijing Children’s Hospital

Zitieren

APA

Li, H., Chen, Z., Gao, S., Hu, J., Yang, Z., Peng, Y., & Sun, J. (2025). Performance Evaluation of Image Segmentation Using Dual-Energy Spectral CT Images with Deep Learning Image Reconstruction: A Phantom Study. Tomography. https://doi.org/10.3390/tomography11050051

Vancouver

Li H, Chen Z, Gao S, Hu J, Yang Z, Peng Y, et al. Performance Evaluation of Image Segmentation Using Dual-Energy Spectral CT Images with Deep Learning Image Reconstruction: A Phantom Study. Tomography. 2025. doi:10.3390/tomography11050051

BibTeX
@article{Li2025,
  author = {Li, Haoyan and Chen, Zhenpeng and Gao, Shuaiyi and Hu, Jiaqi and Yang, Zhihao and Peng, Yun and Sun, Jihang},
  title = {Performance Evaluation of Image Segmentation Using Dual-Energy Spectral CT Images with Deep Learning Image Reconstruction: A Phantom Study},
  journal = {Tomography},
  year = {2025},
  doi = {10.3390/tomography11050051}
}
Kostenlos starten →