Variable flip angle T1 mapping for quantitative liver imaging at 0.55 T
Magnetic Resonance Materials in Physics, Biology and Medicine · 19.09.2026
DOI: 10.1007/s10334-026-01424-w ↗Kurz gefasst
Fachartikel in der Zeitschrift Magnetic Resonance Materials in Physics, Biology and Medicine, online erschienen am 19.09.2026. Verfasst von Majd Helo und 2 weiteren Autor:innen. Beteiligte Einrichtung: Medical Image and Data Analysis (MIDAS.Lab), Department of Diagnostic and Interventional Radiology, University of Tuebingen, Tuebingen, Germany. majd…. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 49 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Eberhard Karls Universität Tübingen.
Abstract (Originalsprache)
Purpose
Investigating the feasibility of variable flip angle (VFA) T 1 mapping for quantitative liver imaging at 0.55 T and evaluating the impact of RF spoiling correction and deep learning-based (DL) reconstruction on accuracy.
Methods
A VFA protocol was developed and combined with DL reconstruction to improve quantitative robustness at 0.55 T. Phantom experiments validated T 1 estimates against multi-TI inversion-recovery spin-echo reference measurements. Bloch simulations characterized RF spoiling-induced bias and derived 0.55 T-specific calibration coefficients. Twelve healthy subjects were scanned to assess field-strength-dependent effects of B 1 + inhomogeneity.
Results
In vivo, DL reconstruction reduced noise-related variability of voxel-wise hepatic T 1 estimates compared to conventional reconstructions in the same subject. At 0.55 T, B 1 + maps showed the highest transmit homogeneity, with liver flip angle variation of 98.2 ± 2.9%, compared with 94.6 ± 7.4% at 1.5 T and 100.7 ± 9.8% at 3 T. Accordingly, B 1 + correction had minimal impact on hepatic T 1 distributions at 0.55 T, whereas more pronounced effects were observed at higher field strengths. Phantom experiments demonstrated improved agreement of VFA-derived T 1 values with the IR-SE reference after 0.55 T-specific RF spoiling correction, supporting the accuracy of the corrected VFA approach. Bland–Altman analysis confirmed that RF spoiling correction had a greater impact on phantom T 1 accuracy than B 1 + correction.
Conclusion
At 0.55 T, RF spoiling and reconstruction strategies are the main determinants of quantitative accuracy, while transmit field inhomogeneity plays a minor role. Combined RF spoiling correction and DL reconstruction enables robust VFA T 1 mapping without B 1 + correction. Reduced low-field T 1 values improve VFA model conditioning and reduce sensitivity to flip angle deviations, supporting VFA-based T 1 mapping for low-field MRI.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Medical Image and Data Analysis (MIDAS.Lab), Department of Diagnostic and Interventional Radiology, University of Tuebingen, Tuebingen, Germany. majd.helo@siemens-healthineers.com.
- Research and Clinical Translation, Magnetic Resonance, Siemens Healthineers AG, Erlangen, Germany. majd.helo@siemens-healthineers.com.
- Research and Clinical Translation, Magnetic Resonance, Siemens Healthineers AG, Erlangen, Germany.
- Medical Image and Data Analysis (MIDAS.Lab), Department of Diagnostic and Interventional Radiology, University of Tuebingen, Tuebingen, Germany.
Förderung
- Eberhard Karls Universität Tübingen
Zitieren
APA
Helo, M., Nickel, M. D., & Küstner, T. (2026). Variable flip angle T1 mapping for quantitative liver imaging at 0.55 T. Magnetic Resonance Materials in Physics, Biology and Medicine. https://doi.org/10.1007/s10334-026-01424-w
Vancouver
Helo M, Nickel MD, Küstner T. Variable flip angle T1 mapping for quantitative liver imaging at 0.55 T. Magnetic Resonance Materials in Physics, Biology and Medicine. 2026. doi:10.1007/s10334-026-01424-w
BibTeX
@article{Helo2026,
author = {Helo, Majd and Nickel, Marcel Dominik and Küstner, Thomas},
title = {Variable flip angle T1 mapping for quantitative liver imaging at 0.55 T},
journal = {Magnetic Resonance Materials in Physics, Biology and Medicine},
year = {2026},
doi = {10.1007/s10334-026-01424-w}
}
EUVIMED