Fachartikel in der Zeitschrift NMR in Biomedicine, online erschienen am 17.03.2026. Verfasst von Nicolás Garrido und 5 weiteren Autor:innen. Beteiligte Einrichtung: Institute for Biological and Medical Engineering, Pontificia Universidad Católica de Chile, Santiago, Chile. Hauptthemen laut MEDLINE: Imaging, Three-Dimensional, Magnetic Resonance Imaging, Contrast Media, Knee Joint. Die Arbeit nennt 43 Literaturangaben. Gefördert von: Beca ANID de Doctorado Nacional Folio, ANID Fondo de Equipamiento Científico y Tecnológico Mayor Fondequip, ANID Millennium Institute iHEALTH.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Quantitative MRI ( T 1 – T 2 ) has been shown to be useful for early diagnosis of osteoarthritis (OA) in the knee at 1.5 and 3 T. However, limited research has been performed at 0.55 T. In this study, we aim to develop and evaluate the feasibility of a 3D joint T 1 – T 2 mapping estimation and synthetic contrasts radial sequence for quantitative and anatomical water‐fat imaging of the knee at 0.55 T. The proposed sequence consists of a free‐running, 3D‐golden angle radial trajectory preceded with inversion recovery and ‐preparation pulses. Reconstruction of images is performed with parallel imaging and high‐dimensional low‐rank patch‐based regularization. A water–fat separation technique is employed to obtain water and fat images, and T 1 – T 2 maps can be obtained for water and the echo images via Bloch‐equation dictionary matching. The sequence was validated on a standardized phantom and in 16 healthy subjects. Phantom results are in good agreement with SE references, with biases of ms for and ms for . In vivo, the average mean in the articular cartilage was ms and ms, compared to ms, ms for the reference sequences. Synthetic contrast images produced with the sequence are comparable to the references in terms of contrast to noise ratio (CNR) and relative contrast metrics. The feasibility of 3D joint T 1 – T 2 mapping of the knee at 0.55 T in a single scan of ~3 min was demonstrated, showing promising results for efficient knee imaging. Studies in a larger cohort of patients with osteoarthritis at 0.55 T are now warranted.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Imaging, Three-DimensionalMagnetic Resonance ImagingContrast MediaKnee JointKneeHumansPhantoms, ImagingMaleAdultFemaleTime FactorsReproducibility of Results

Stichwörter der Autor:innen

T1 mappingT2 mappingknee MRIlow‐fieldmulti‐parametric

Substanzen

Contrast Media

Autor:innen und Institutionen

  • Institute for Biological and Medical Engineering, Pontificia Universidad Católica de Chile, Santiago, Chile.
  • Millennium Institute for Intelligent Healthcare Engineering, Pontificia Universidad Católica de Chile, Santiago, Chile.
  • School of Biomedical Engineering and Imaging Sciences, King's College London, London, UK.
  • School of Engineering, Department of Electrical Engineering, Pontificia Universidad Católica de Chile, Santiago, Chile.
  • Institute for Advanced Study at Technical University of Munich, Munich, Germany.

ORCID-Profile: Nicolás Garrido, Diego Pedraza, Carlos Castillo-Passi

Förderung

  • Beca ANID de Doctorado Nacional Folio
  • ANID Fondo de Equipamiento Científico y Tecnológico Mayor Fondequip
  • ANID Millennium Institute iHEALTH
  • ANID FONDECYT
  • ANID Center of Interventional Medicine for Precision and Advanced Cellular Therapy (IMPACT)

Zitieren

APA

Garrido, N., Pedraza, D., Castillo-Passi, C., Barrera, D., Prieto, C., & Botnar, R. (2026). 3D Radial Knee Imaging at 0.55 T: Simultaneous T 1 – T 2 Mapping and Synthetic Contrast Generation in Under 4 min. NMR in Biomedicine, 39(4). https://doi.org/10.1002/nbm.70265

Vancouver

Garrido N, Pedraza D, Castillo-Passi C, Barrera D, Prieto C, Botnar R. 3D Radial Knee Imaging at 0.55 T: Simultaneous T 1 – T 2 Mapping and Synthetic Contrast Generation in Under 4 min. NMR in Biomedicine. 2026;39(4). doi:10.1002/nbm.70265

BibTeX
@article{Garrido2026,
  author = {Garrido, Nicolás and Pedraza, Diego and Castillo-Passi, Carlos and Barrera, Dabne and Prieto, Claudia and Botnar, René},
  title = {3D Radial Knee Imaging at 0.55 T: Simultaneous T 1 – T 2 Mapping and Synthetic Contrast Generation in Under 4 min},
  journal = {NMR in Biomedicine},
  year = {2026},
  volume = {39},
  number = {4},
  doi = {10.1002/nbm.70265}
}
Kostenlos starten →