Application of high‐density 2D receiver coil arrays for improved SNR in prostate MRI
Magnetic Resonance in Medicine · 2025
DOI: 10.1002/mrm.30289 ↗Kurz gefasst
Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, 2025. Verfasst von Stephen J Riederer und 9 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiology, Mayo Clinic, Rochester, Minnesota, USA. Hauptthemen laut MEDLINE: Magnetic Resonance Imaging, Phantoms, Imaging, Prostate, Signal-To-Noise Ratio. Die Arbeit nennt 36 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: NCRR NIH HHS, NIBIB NIH HHS, National Institute of Biomedical Imaging and Bioengineering.
Abstract (Originalsprache)
Purpose
To study if adaptive image receive (AIR) receiver coil elements can be configured into a 2D array with high (>45% by diameter) element‐to‐element overlap, allowing improved SNR at depth (0.7–1.5× element diameter) versus conventional (20%) overlap.
Methods
An anterior array composed of twenty 10‐cm diameter elements with 45% overlap arranged into a 4 × 5 grid and a similar 3 × 7 twenty‐one‐element posterior array were constructed. SNR and g‐factor were measured in a pelvic phantom using the new high‐density (HD) arrays (41 total elements) and compared to vendor AIR‐based arrays (30 total elements) with conventional overlap. T 2 ‐weighted fast‐spin‐echo (T2SE) images acquired using both arrays were compared in 20 subjects. SNR was estimated in vivo. Results were compared blindly by three uroradiologists using a five‐point scale. Images using the HD arrays were also compared to a set of images acquired over a range of acceleration factors ( R = 2.0, 2.5, 3.0) with the conventional arrays.
Results
SNR within the phantom was on average 15% higher for R = 1.0, 1.5, and 2.0 using the HD arrays. Across the 20 subjects SNR within the prostate was 11% higher and assessed radiologically as significantly higher ( p < 0.001) for the HD versus conventional arrays. At all acceleration factors the new HD arrays outperformed the conventional arrays ( p ≤ 0.01), allowing increased R for similar SNR.
Conclusion
AIR elements can be configured into 2D arrays with high (45%) element‐to‐element overlap, consistently providing increased SNR at depth versus arrays with conventional (20%) overlap. The SNR improvement allows increased acceleration in T2SE prostate MRI.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Radiology, Mayo Clinic, Rochester, Minnesota, USA.
- Grinnell College, Grinnell, Iowa, USA.
- Department of Radiology, Mayo Clinic, Phoenix, Arizona, USA.
- GE HealthCare, Aurora, Ohio, USA.
ORCID-Profile: Stephen J Riederer, Eric A Borisch
Förderung
- NCRR NIH HHS
- NIBIB NIH HHS
- National Institute of Biomedical Imaging and Bioengineering
Zitieren
APA
Riederer, S. J., Borisch, E. A., Du, Q., Froemming, A. T., Hulshizer, T. C., Kawashima, A., McGee, K. P., Robb, F., Rossman, P. J., & Takahashi, N. (2025). Application of high‐density 2D receiver coil arrays for improved SNR in prostate MRI. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.30289
Vancouver
Riederer SJ, Borisch EA, Du Q, Froemming AT, Hulshizer TC, Kawashima A, et al. Application of high‐density 2D receiver coil arrays for improved SNR in prostate MRI. Magnetic Resonance in Medicine. 2025. doi:10.1002/mrm.30289
BibTeX
@article{Riederer2025,
author = {Riederer, Stephen J and Borisch, Eric A and Du, Qingshuo and Froemming, Adam T and Hulshizer, Thomas C and Kawashima, Akira and McGee, Kiaran P and Robb, Fraser and Rossman, Phillip J and Takahashi, Naoki},
title = {Application of high‐density 2D receiver coil arrays for improved SNR in prostate MRI},
journal = {Magnetic Resonance in Medicine},
year = {2025},
doi = {10.1002/mrm.30289}
}
EUVIMED