Quantitative image quality metrics enable resource-efficient quality control of clinically applied AI-based reconstructions in MRI
Magnetic Resonance Materials in Physics, Biology and Medicine · 2025
DOI: 10.1007/s10334-025-01253-3 ↗Kurz gefasst
Fachartikel in der Zeitschrift Magnetic Resonance Materials in Physics, Biology and Medicine, 2025. Verfasst von Owen A White und 13 weiteren Autor:innen. Beteiligte Einrichtung: MRI Unit, The Royal Marsden NHS Foundation Trust, London, UK. owen.white@rmh.nhs.uk. Hauptthemen laut MEDLINE: Magnetic Resonance Imaging, Image Processing, Computer-Assisted, Artificial Intelligence, Image Interpretation, Computer-Assisted. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 57 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: UK Government's Department for Science, Innovation & Technology, Cancer Research UK, NIHR Biomedical Research Centre, Royal Marsden NHS Foundation Trust/Institute of Cancer Research.
Abstract (Originalsprache)
Objective
AI-based MRI reconstruction techniques improve efficiency by reducing acquisition times whilst maintaining or improving image quality. Recent recommendations from professional bodies suggest centres should perform quality assessments on AI tools. However, monitoring long-term performance presents challenges, due to model drift or system updates. Radiologist-based assessments are resource-intensive and may be subjective, highlighting the need for efficient quality control (QC) measures. This study explores using image quality metrics (IQMs) to assess AI-based reconstructions. Materials and methods 58 patients undergoing standard-of-care rectal MRI were imaged using AI-based and conventional T2-weighted sequences. Paired and unpaired IQMs were calculated. Sensitivity of IQMs to detect retrospective perturbations in AI-based reconstructions was assessed using control charts, and statistical comparisons between the four MR systems in the evaluation were performed. Two radiologists evaluated the image quality of the perturbed images, giving an indication of their clinical relevance.
Results
Paired IQMs demonstrated sensitivity to changes in AI-reconstruction settings, identifying deviations outside ± 2 standard deviations of the reference dataset. Unpaired metrics showed less sensitivity. Paired IQMs showed no difference in performance between 1.5 T and 3 T systems ( p > 0.99), whilst minor but significant ( p < 0.0379) differences were noted for unpaired IQMs.
Discussion
IQMs are effective for QC of AI-based MR reconstructions, offering resource-efficient alternatives to repeated radiologist evaluations. Future work should expand this to other imaging applications and assess additional measures.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- MRI Unit, The Royal Marsden NHS Foundation Trust, London, UK. owen.white@rmh.nhs.uk.
- Division of Radiotherapy and Imaging, The Institute of Cancer Research, London, UK. owen.white@rmh.nhs.uk.
- MRI Unit, The Royal Marsden NHS Foundation Trust, London, UK.
- Division of Radiotherapy and Imaging, The Institute of Cancer Research, London, UK.
- National Physical Laboratory, Teddington, UK.
ORCID-Profile: Joshua Shur, Francesca Castagnoli, Geoff Charles-Edwards, Brandon Whitcher, David J Collins, Matthew T D Cashmore, Matt G Hall, Spencer A Thomas, Andrew Thompson, Ciara A Harrison, Georgina Hopkinson, Dow-Mu Koh, Jessica M Winfield
Förderung
- UK Government's Department for Science, Innovation & Technology
- Cancer Research UK
- NIHR Biomedical Research Centre, Royal Marsden NHS Foundation Trust/Institute of Cancer Research
- UK Government’s Department for Science, Innovation & Technology
Zitieren
APA
White, O. A., Shur, J., Castagnoli, F., Charles-Edwards, G., Whitcher, B., Collins, D. J., Cashmore, M. T. D., Hall, M. G., Thomas, S. A., Thompson, A., Harrison, C. A., Hopkinson, G., Koh, D. M., & Winfield, J. M. (2025). Quantitative image quality metrics enable resource-efficient quality control of clinically applied AI-based reconstructions in MRI. Magnetic Resonance Materials in Physics, Biology and Medicine. https://doi.org/10.1007/s10334-025-01253-3
Vancouver
White OA, Shur J, Castagnoli F, Charles-Edwards G, Whitcher B, Collins DJ, et al. Quantitative image quality metrics enable resource-efficient quality control of clinically applied AI-based reconstructions in MRI. Magnetic Resonance Materials in Physics, Biology and Medicine. 2025. doi:10.1007/s10334-025-01253-3
BibTeX
@article{White2025,
author = {White, Owen A and Shur, Joshua and Castagnoli, Francesca and Charles-Edwards, Geoff and Whitcher, Brandon and Collins, David J and Cashmore, Matthew T D and Hall, Matt G and Thomas, Spencer A and Thompson, Andrew and Harrison, Ciara A and Hopkinson, Georgina and Koh, Dow-Mu and Winfield, Jessica M},
title = {Quantitative image quality metrics enable resource-efficient quality control of clinically applied AI-based reconstructions in MRI},
journal = {Magnetic Resonance Materials in Physics, Biology and Medicine},
year = {2025},
doi = {10.1007/s10334-025-01253-3}
}
EUVIMED