Towards assessing and improving the reliability of ultrashort echo time quantitative magnetization transfer (UTE-qMT) MRI of cortical bone: In silico and ex vivo study
Magnetic Resonance Materials in Physics, Biology and Medicine · 2024
DOI: 10.1007/s10334-024-01190-7 ↗Kurz gefasst
Fachartikel in der Zeitschrift Magnetic Resonance Materials in Physics, Biology and Medicine, 2024. Verfasst von Soo Hyun Shin und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiology, University of California San Diego, 9452 Medical Center Drive, La Jolla, CA, USA. shs033@health.ucsd.edu. Hauptthemen laut MEDLINE: Phantoms, Imaging, Magnetic Resonance Imaging, Cortical Bone, Algorithms. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 53 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: NIAMS NIH HHS, BLRD VA, NIH HHS.
Abstract (Originalsprache)
Objective
To assess and improve the reliability of the ultrashort echo time quantitative magnetization transfer (UTE-qMT) modeling of the cortical bone. Materials and Methods Simulation-based digital phantoms were created that mimic the UTE-qMT properties of cortical bones. A wide range of SNR from 25 to 200 was simulated by adding different levels of noise to the synthesized MT-weighted images to assess the effect of SNR on UTE-qMT fitting results. Tensor-based denoising algorithm was applied to improve the fitting results. These results from digital phantom studies were validated via ex vivo rat leg bone scans.
Results
The selection of initial points for nonlinear fitting and the number of data points tested for qMT analysis have minimal effect on the fitting result. Magnetization exchange rate measurements are highly dependent on the SNR of raw images, which can be substantially improved with an appropriate denoising algorithm that gives similar fitting results from the raw images with an 8-fold higher SNR.
Discussion
The digital phantom approach enables the assessment of the reliability of bone UTE-qMT fitting by providing the known ground truth. These findings can be utilized for optimizing the data acquisition and analysis pipeline for UTE-qMT imaging of cortical bones.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Radiology, University of California San Diego, 9452 Medical Center Drive, La Jolla, CA, USA. shs033@health.ucsd.edu.
- Department of Radiology, University of California San Diego, 9452 Medical Center Drive, La Jolla, CA, USA.
- Radiology Service, Veterans Affairs San Diego Healthcare System, 3350 La Jolla Village Drive, San Diego, CA, USA.
- Shu Chien-Gene Lay Department of Bioengineering, University of California San Diego, La Jolla, CA, USA.
- Department of Radiology, University of California San Diego, 9452 Medical Center Drive, La Jolla, CA, USA. ericchangmd@gmail.com.
- Radiology Service, Veterans Affairs San Diego Healthcare System, 3350 La Jolla Village Drive, San Diego, CA, USA. ericchangmd@gmail.com.
Förderung
- NIAMS NIH HHS
- BLRD VA
- NIH HHS
- U.S. Department of Veterans Affairs
- CSRD VA
- University of California, San Diego
- National Institutes of Health
Zitieren
APA
Shin, S. H., Moazamian, D., Tang, Q., Jerban, S., Ma, Y., Du, J., & Chang, E. Y. (2024). Towards assessing and improving the reliability of ultrashort echo time quantitative magnetization transfer (UTE-qMT) MRI of cortical bone: In silico and ex vivo study. Magnetic Resonance Materials in Physics, Biology and Medicine. https://doi.org/10.1007/s10334-024-01190-7
Vancouver
Shin SH, Moazamian D, Tang Q, Jerban S, Ma Y, Du J, et al. Towards assessing and improving the reliability of ultrashort echo time quantitative magnetization transfer (UTE-qMT) MRI of cortical bone: In silico and ex vivo study. Magnetic Resonance Materials in Physics, Biology and Medicine. 2024. doi:10.1007/s10334-024-01190-7
BibTeX
@article{Shin2024,
author = {Shin, Soo Hyun and Moazamian, Dina and Tang, Qingbo and Jerban, Saeed and Ma, Yajun and Du, Jiang and Chang, Eric Y},
title = {Towards assessing and improving the reliability of ultrashort echo time quantitative magnetization transfer (UTE-qMT) MRI of cortical bone: In silico and ex vivo study},
journal = {Magnetic Resonance Materials in Physics, Biology and Medicine},
year = {2024},
doi = {10.1007/s10334-024-01190-7}
}
EUVIMED