Motion robust coronary MR angiography using zigzag centric ky–kz trajectory and high-resolution deep learning reconstruction
Magnetic Resonance Materials in Physics, Biology and Medicine · 2024
DOI: 10.1007/s10334-024-01172-9 ↗Kurz gefasst
Fachartikel in der Zeitschrift Magnetic Resonance Materials in Physics, Biology and Medicine, 2024. Verfasst von Hideki Ota und 8 weiteren Autor:innen. Beteiligte Einrichtung: Department of Advanced Radiological Imaging Collaborative Research, Graduate School of Medicine, Tohoku University, Sendai, Japan. Hauptthemen laut MEDLINE: Magnetic Resonance Angiography, Deep Learning, Signal-To-Noise Ratio, Image Processing, Computer-Assisted. Die Arbeit wurde bisher 7-mal zitiert (Crossref) und nennt 39 Literaturangaben. Gefördert von: NHLBI NIH HHS, NIA NIH HHS, Canon Medical Systems Corporation.
Abstract (Originalsprache)
PURPOSE: To develop a new MR coronary angiography (MRCA) technique by employing a zigzag fan-shaped centric ky-kz k-space trajectory combined with high-resolution deep learning reconstruction (HR-DLR). METHODS: All imaging data were acquired from 12 healthy subjects and 2 patients using two clinical 3-T MR imagers, with institutional review board approval. Ten healthy subjects underwent both standard 3D fast gradient echo (sFGE) and centric ky-kz k-space trajectory FGE (cFGE) acquisitions to compare the scan time and image quality. Quantitative measures were also performed for signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) as well as sharpness of the vessel. Furthermore, the feasibility of the proposed cFGE sequence was assessed in two patients. For assessing the feasibility of the centric ky-kz trajectory, the navigator-echo window of a 30-mm threshold was applied in cFGE, whereas sFGE was applied using a standard 5-mm threshold. Image quality of MRCA using cFGE with HR-DLR and sFGE without HR-DLR was scored in a 5-point scale (non-diagnostic = 1, fair = 2, moderate = 3, good = 4, and excellent = 5). Image evaluation of cFGE, applying HR-DLR, was compared with sFGE without HR-DLR. Friedman test, Wilcoxon signed-rank test, or paired t tests were performed for the comparison of related variables. RESULTS: The actual MRCA scan time of cFGE with a 30-mm threshold was acquired in less than 5 min, achieving nearly 100% efficiency, showcasing its expeditious and robustness. In contrast, sFGE was acquired with a 5-mm threshold and had an average scan time of approximately 15 min. Overall image quality for MRCA was scored 3.3 for sFGE and 2.7 for cFGE without HR-DLR but increased to 3.6 for cFGE with HR-DLR and (p < 0.05). The clinical result of patients obtained within 5 min showed good quality images in both patients, even with a stent, without artifacts. Quantitative measures of SNR, CNR, and sharpness of vessel presented higher in cFGE with HR-DLR. CONCLUSION: Our findings demonstrate a robust, time-efficient solution for high-quality MRCA, enhancing patient comfort and increasing clinical throughput.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Advanced Radiological Imaging Collaborative Research, Graduate School of Medicine, Tohoku University, Sendai, Japan.
- Department of Diagnostic Radiology, Tohoku University Hospital, Sendai, Japan.
- Department of Radiology, National Cerebral and Cardiovascular Center, Suita, Osaka, Japan.
- Department of Radiology, University of California, San Diego, La Jolla, CA, USA.
- Canon Medical Systems Corporation, Otawara, Tochigi, Japan.
- Department of Cardiology, University of California, San Diego, La Jolla, CA, USA.
Förderung
- NHLBI NIH HHS
- NIA NIH HHS
- Canon Medical Systems Corporation
- NIH HHS
- NIH
Zitieren
APA
Ota, H., Morita, Y., Vucevic, D., Higuchi, S., Takagi, H., Kutsuna, H., Yamashita, Y., Kim, P., & Miyazaki, M. (2024). Motion robust coronary MR angiography using zigzag centric ky–kz trajectory and high-resolution deep learning reconstruction. Magnetic Resonance Materials in Physics, Biology and Medicine. https://doi.org/10.1007/s10334-024-01172-9
Vancouver
Ota H, Morita Y, Vucevic D, Higuchi S, Takagi H, Kutsuna H, et al. Motion robust coronary MR angiography using zigzag centric ky–kz trajectory and high-resolution deep learning reconstruction. Magnetic Resonance Materials in Physics, Biology and Medicine. 2024. doi:10.1007/s10334-024-01172-9
BibTeX
@article{Ota2024,
author = {Ota, Hideki and Morita, Yoshiaki and Vucevic, Diana and Higuchi, Satoshi and Takagi, Hidenobu and Kutsuna, Hideaki and Yamashita, Yuichi and Kim, Paul and Miyazaki, Mitsue},
title = {Motion robust coronary MR angiography using zigzag centric ky–kz trajectory and high-resolution deep learning reconstruction},
journal = {Magnetic Resonance Materials in Physics, Biology and Medicine},
year = {2024},
doi = {10.1007/s10334-024-01172-9}
}
EUVIMED