EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Uncertainty quantification for patient-specific domain in virtual aortic procedures: application to thoracic endovascular aortic repair

Vittorio Lissoni, Anna Ramella, Giulia Luraghi, Puck Stassen, Wouter Huberts, Santi Trimarchi, Francesco Migliavacca, Jose Felix Rodriguez Matas

Biomechanics and Modeling in Mechanobiology · 2026

DOI: 10.1007/s10237-025-02036-4 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Biomechanics and Modeling in Mechanobiology, 2026. Verfasst von Vittorio Lissoni und 7 weiteren Autor:innen. Beteiligte Einrichtung: Computational Biomechanics Laboratory-LaBS, Department of Chemistry, Materials and Chemical Engineering "Giulio Natta", Politecnico di Milano, Piazza…. Hauptthemen laut MEDLINE: Endovascular Procedures, Aorta, Thoracic. Die Arbeit nennt 43 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: European Research Council, Politecnico di Milano.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Abstracts Simulating medical procedures requires accounting for inherent uncertainty in many numerical model parameters, such as material properties. Evaluating the impact of these uncertainties is crucial for identifying parameters needing precise definition and correctly interpreting simulation results. This study explores how uncertainties in modelling the aorta affect finite element outcomes of a thoracic endovascular aortic repair (TEVAR) procedure. Based on literature data, aortic wall thickness and mechanical properties were identified as the most uncertain. The aorta was modelled using shell elements with homogeneous thickness and assumed to behave as a linear elastic isotropic material. A design of experiments approach was used for uncertainty quantification and sensitivity analysis: wall thickness and Young’s modulus were varied over 11 levels in a full factorial design, resulting in 121 simulations. Uncertainty was quantified using statistical metrics such as mean, standard deviation, coefficient of variation, and 95% confidence intervals. Results indicate wall thickness significantly affects aortic wall stress (σ aorta ), with minimal influence on stent stress (σ stent ) and device opening area (OA). Conversely, Young’s modulus has limited impact on σ aorta but affects σ stent and OA to a greater extent. The highest uncertainty was observed in σ aorta (~ 25% coefficient of variation), while σ stent and OA showed lower variability (2.6% and 6.9%, respectively). These findings suggest that, in this model, accurate wall thickness definition is more critical than precise Young’s modulus for reducing uncertainty in wall stress predictions. Therefore, literature-based averages for Young’s modulus may be sufficient for simulating this procedure.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Endovascular ProceduresAorta, ThoracicHumansUncertaintyFinite Element AnalysisStress, MechanicalElastic ModulusStentsModels, CardiovascularComputer SimulationEndovascular Aneurysm Repair

Stichwörter der Autor:innen

Digital twin in healthcareFinite element analysisIn silico medicineTEVARVVUQ

Autor:innen und Institutionen

  • Computational Biomechanics Laboratory-LaBS, Department of Chemistry, Materials and Chemical Engineering "Giulio Natta", Politecnico di Milano, Piazza L. da Vinci 32, 20133, Milan, Italy.
  • Cardiovascular Biomechanics, Biomedical Engineering, Eindhoven University of Technology, Eindhoven, Netherlands.
  • Foundation IRCCS Ca' Granda Ospedale Maggiore Policlinico, Milano, Italy.
  • Department of Clinical Sciences and Community Health, University of Milan, Milano, Italy.
  • Computational Biomechanics Laboratory-LaBS, Department of Chemistry, Materials and Chemical Engineering "Giulio Natta", Politecnico di Milano, Piazza L. da Vinci 32, 20133, Milan, Italy. francesco.migliavacca@polimi.it.
  • Foundation IRCCS Ca' Granda Ospedale Maggiore Policlinico, Milano, Italy. francesco.migliavacca@polimi.it.

Förderung

  • European Research Council
  • Politecnico di Milano

Zitieren

APA

Lissoni, V., Ramella, A., Luraghi, G., Stassen, P., Huberts, W., Trimarchi, S., Migliavacca, F., & Matas, J. F. R. (2026). Uncertainty quantification for patient-specific domain in virtual aortic procedures: application to thoracic endovascular aortic repair. Biomechanics and Modeling in Mechanobiology. https://doi.org/10.1007/s10237-025-02036-4

Vancouver

Lissoni V, Ramella A, Luraghi G, Stassen P, Huberts W, Trimarchi S, et al. Uncertainty quantification for patient-specific domain in virtual aortic procedures: application to thoracic endovascular aortic repair. Biomechanics and Modeling in Mechanobiology. 2026. doi:10.1007/s10237-025-02036-4

BibTeX
@article{Lissoni2026,
  author = {Lissoni, Vittorio and Ramella, Anna and Luraghi, Giulia and Stassen, Puck and Huberts, Wouter and Trimarchi, Santi and Migliavacca, Francesco and Matas, Jose Felix Rodriguez},
  title = {Uncertainty quantification for patient-specific domain in virtual aortic procedures: application to thoracic endovascular aortic repair},
  journal = {Biomechanics and Modeling in Mechanobiology},
  year = {2026},
  doi = {10.1007/s10237-025-02036-4}
}
Kostenlos starten →