Effects of Long Term Fatigue Cycling on In Situ Fenestrations of Polyethylene Terephthalate and Expanded Polytetrafluorethylene Thoracic Aortic Stent grafts: An Experimental Study
European Journal of Vascular and Endovascular Surgery · 2024-05 · 67 (5) · 819-826
DOI: 10.1016/j.ejvs.2023.08.060 ↗Kurz gefasst
Fachartikel in der Zeitschrift European Journal of Vascular and Endovascular Surgery, 2024-05. Verfasst von Yuan He und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Vascular Surgery, The First Affiliated Hospital, School of Medicine, Zhejiang University, Hangzhou City, 310003, China. Hauptthemen laut MEDLINE: Blood Vessel Prosthesis, Stents, Polyethylene Terephthalates, Prosthesis Design. Die Arbeit wurde bisher 10-mal zitiert (Crossref) und nennt 26 Literaturangaben. Gefördert von: Medical Science and Technology Project of Zhejiang Province, Health and Family Planning Commission of Zhejiang Province, Science and Technology Department of Zhejiang Province.
Abstract (Originalsprache)
OBJECTIVE: In this study, the long term durability of fenestrations after in situ fenestration (ISF) of five commercial thoracic aortic stent grafts was evaluated in an in vitro experiment after a simulated 10 year period. METHODS: Five different thoracic aortic stent grafts (Relay, Valiant, Hercules, TAG, and Ankura, with a diameter of 34 mm) received both needle and laser ISF in vitro. A Viabahn (11 × 50 mm) was released in each fenestration as a bridging stent graft. Long term fatigue tests (simulating 10 years) of each of the fenestrated stent grafts were then conducted in a flow fatigue test system. The area, shape, margin, and the long and short axis of all the fenestrations were evaluated with light microscopy before and after the fatigue test. The leakage from the fenestration junction before and after the long term fatigue was also measured. RESULTS: The experimental results showed no obvious difference between needle and laser fenestrations. The long axes of all the fenestrations remained unchanged, while the short axes increased after the fatigue test, which was significant in Relay, Valiant, and Hercules polyethylene terephthalate stent grafts. The shape scores of fenestrations improved after the fatigue test in Valiant and Hercules, remained unchanged in Relay and Ankura, and worsened in the TAG. After the fatigue cycling, the average leakage from the fenestration junction decreased in all the stent grafts, and the Ankura had the maximum decline rate. CONCLUSION: The ISF technique was durable over a simulated 10 year period. The fenestrations were positively remodelled to be more circular, and the leakage from the junction decreased after long term fatigue testing.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Vascular Surgery, The First Affiliated Hospital, School of Medicine, Zhejiang University, Hangzhou City, 310003, China.
- Department of Vascular Surgery, The First Affiliated Hospital, School of Medicine, Zhejiang University, Hangzhou City, 310003, China. Electronic address: 1198050@zju.edu.cn.
- Department of Vascular Surgery, The First Affiliated Hospital, School of Medicine, Zhejiang University, Hangzhou City, 310003, China. Electronic address: lidonglin@zju.edu.cn.
Förderung
- Medical Science and Technology Project of Zhejiang Province
- Health and Family Planning Commission of Zhejiang Province
- Science and Technology Department of Zhejiang Province
- Zhejiang Province Natural Science Foundation
Zitieren
APA
He, Y., Wang, Y., Zhou, X., Wu, Z., Zhang, H., & Li, D. (2024). Effects of Long Term Fatigue Cycling on In Situ Fenestrations of Polyethylene Terephthalate and Expanded Polytetrafluorethylene Thoracic Aortic Stent grafts: An Experimental Study. European Journal of Vascular and Endovascular Surgery, 67(5), 819-826. https://doi.org/10.1016/j.ejvs.2023.08.060
Vancouver
He Y, Wang Y, Zhou X, Wu Z, Zhang H, Li D. Effects of Long Term Fatigue Cycling on In Situ Fenestrations of Polyethylene Terephthalate and Expanded Polytetrafluorethylene Thoracic Aortic Stent grafts: An Experimental Study. European Journal of Vascular and Endovascular Surgery. 2024;67(5):819-826. doi:10.1016/j.ejvs.2023.08.060
BibTeX
@article{He2024,
author = {He, Yuan and Wang, Yishu and Zhou, Xiang and Wu, Ziheng and Zhang, Hongkun and Li, Donglin},
title = {Effects of Long Term Fatigue Cycling on In Situ Fenestrations of Polyethylene Terephthalate and Expanded Polytetrafluorethylene Thoracic Aortic Stent grafts: An Experimental Study},
journal = {European Journal of Vascular and Endovascular Surgery},
year = {2024},
volume = {67},
number = {5},
pages = {819--826},
doi = {10.1016/j.ejvs.2023.08.060}
}
EUVIMED