EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Beyond S-cross section design: mechanical properties, metallurgical features and shaping ability of two nickel-titanium rotary instruments in double curved canals

Larissa Tres, Gabriel Barcelos Só, Raimundo Sales de Oliveira Neto, Marco Antônio Húngaro Duarte, Carolina Horn Troian-Michel, Ricardo Abreu da Rosa, Murilo Priori Alcalde, Marcus Vinícius Reis Só

Clinical Oral Investigations · 29.05.2026 · 30 (6)

DOI: 10.1007/s00784-026-06953-1 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Clinical Oral Investigations, online erschienen am 29.05.2026. Verfasst von Larissa Tres und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Conservative Dentistry, School of Dentistry, Rio Grande do Sul Federal University (UFRGS), R. Ramiro Barcelos, 2492 - Santa Cecília, Po…. Hauptthemen laut MEDLINE: Root Canal Preparation, Nickel, Dental Instruments, Titanium. Die Arbeit nennt 31 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Universidade Federal Do Rio Grande Do Sul.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objectives

This study evaluated the metallurgical properties and shaping ability of two nickel-titanium instruments in artificial double-curvature canals.

Methods

20 NiTi instruments (25 mm in length) with S -shaped cross-section design and different NiTi alloy were selected for analysis ( n = 10, SPIN 25.06; n = 10 Rotate 25.06). The instruments were evaluated with respect to geometric design, differential scanning calorimetry, and shaping ability. Cyclic fatigue was also evaluated after three simulated clinical uses, comparing it to that of new instruments. Data were analyzed using parametric and nonparametric tests according to data distribution, with a significance level set at 5%.

Results

The instruments exhibited similar design features and different phase transformation temperatures. Rotate 25.06 presented lower metal mass volume and cross-sectional area than Spin 25.06 ( P < 0.05). Rotate 25.06 resulted in lower canal deviation and better center ability than Spin 25.06 ( P < 0.05). Spin 25.06 instruments demonstrated greater cyclic fatigue resistance ( P < 0.05); however, after three simulated clinical uses, both instruments exhibited similar clinical fatigue resistance ( P > 0.05).

Conclusion

Although Spin 25.06 showed greater resistance to cyclic fatigue and better phase transformation temperature, Rotate 25.06 achieved superior shaping performance, promoting less deviation and better centering ability in the artificial double-curved root canals. Clinical relevance Selection of NiTi instruments must consider cyclic fatigue resistance, shaping ability and metallurgical behavior. Our in vitro findings highlight the need of balancing instruments’ properties to reduce procedural errors and optimize clinical outcomes.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Root Canal PreparationNickelDental InstrumentsTitaniumEquipment DesignSurface PropertiesCalorimetry, Differential ScanningHumansMaterials TestingDental Alloys

Stichwörter der Autor:innen

Cyclic fatigueDifferential scanning calorimetryEndodonticsMicrocomputed tomographyScanning electron microscopy

Substanzen

NickelTitaniumDental Alloys

Autor:innen und Institutionen

  • Department of Conservative Dentistry, School of Dentistry, Rio Grande do Sul Federal University (UFRGS), R. Ramiro Barcelos, 2492 - Santa Cecília, Porto Alegre, RS, 90035-004, Brazil.
  • Department of Stomatology, Faculty of Dentistry, Federal University of Santa Maria (UFSM), Santa Maria, RS, Brazil.
  • Department of Operative Dentistry, Endodontics and Dental Materials, Bauru School of Dentistry, University of São Paulo, São Paulo, Brazil.
  • Department of Conservative Dentistry, School of Dentistry, Rio Grande do Sul Federal University (UFRGS), R. Ramiro Barcelos, 2492 - Santa Cecília, Porto Alegre, RS, 90035-004, Brazil. carolinaendo@gmail.com.

Förderung

  • Universidade Federal Do Rio Grande Do Sul

Zitieren

APA

Tres, L., Só, G. B., de Oliveira Neto, R. S., Duarte, M. A. H., Troian-Michel, C. H., da Rosa, R. A., Alcalde, M. P., & Só, M. V. R. (2026). Beyond S-cross section design: mechanical properties, metallurgical features and shaping ability of two nickel-titanium rotary instruments in double curved canals. Clinical Oral Investigations, 30(6). https://doi.org/10.1007/s00784-026-06953-1

Vancouver

Tres L, Só GB, de Oliveira Neto RS, Duarte MAH, Troian-Michel CH, da Rosa RA, et al. Beyond S-cross section design: mechanical properties, metallurgical features and shaping ability of two nickel-titanium rotary instruments in double curved canals. Clinical Oral Investigations. 2026;30(6). doi:10.1007/s00784-026-06953-1

BibTeX
@article{Tres2026,
  author = {Tres, Larissa and Só, Gabriel Barcelos and de Oliveira Neto, Raimundo Sales and Duarte, Marco Antônio Húngaro and Troian-Michel, Carolina Horn and da Rosa, Ricardo Abreu and Alcalde, Murilo Priori and Só, Marcus Vinícius Reis},
  title = {Beyond S-cross section design: mechanical properties, metallurgical features and shaping ability of two nickel-titanium rotary instruments in double curved canals},
  journal = {Clinical Oral Investigations},
  year = {2026},
  volume = {30},
  number = {6},
  doi = {10.1007/s00784-026-06953-1}
}
Kostenlos starten →