Mechanical and Metallurgical In Vitro Evaluation of Electropolished Versus Non‐Electropolished Rotary and Reciprocating Instruments
International Endodontic Journal · 2025
DOI: 10.1111/iej.70009 ↗Kurz gefasst
Vergleichsstudie in der Zeitschrift International Endodontic Journal, 2025. Verfasst von Victor T L Vieira und 9 weiteren Autor:innen. Beteiligte Einrichtung: School of Dentistry, Grande Rio University (UNIGRANRIO), Rio de Janeiro, Brazil. Hauptthemen laut MEDLINE: Dental Instruments, Root Canal Preparation. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 30 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Fundação Carlos Chagas Filho de Amparo à Pesquisa do Estado do Rio de Janeiro, Conselho Nacional de Desenvolvimento Científico e Tecnológico.
Abstract (Originalsprache)
Aim
To evaluate the effect of electropolishing on the mechanical properties of One RECI and One Curve mini nickel‐titanium (NiTi) instruments by comparing electropolished and non‐electropolished versions of each instrument type. Methodology Electropolished and non‐electropolished One RECI (reciprocating) and One Curve mini (rotary) NiTi instruments, all manufactured with identical geometry and heat treatment, were evaluated. Instrument design was analysed by light microscopy and scanning electron microscopy, while metallurgical characterisation was performed using energy‐dispersive X‐ray spectroscopy (EDS) and differential scanning calorimetry (DSC). Mechanical performance was assessed through torsional resistance, bending and buckling load, surface microhardness, and cutting efficiency. Statistical comparisons were performed using the independent samples t‐test or the Mann–Whitney U ‐test, with significance set at p < 0.05.
Results
Design and metallurgical analyses confirmed that electropolished and non‐electropolished instruments within each group were equivalent in terms of geometry, cross‐sectional design, tip configuration, elemental composition, and phase transformation temperatures. Electropolishing significantly enhanced flexibility in both instrument types, as indicated by reduced bending loads and lower buckling resistance ( p < 0.05). However, torsional strength was significantly reduced in the electropolished One RECI instruments, reflected by lower maximum torque and angle of rotation prior to fracture ( p < 0.05). No significant torsional differences were observed in the One Curve mini group ( p > 0.05). Surface microhardness and cutting efficiency remained unaffected by electropolishing in both systems ( p > 0.05).
Conclusions
Electropolishing improved the flexibility of both One RECI and One Curve mini NiTi instruments without compromising their surface microhardness or cutting efficiency. However, its impact on torsional resistance was system‐dependent, resulting in reduced strength only in the reciprocating One RECI instruments.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- School of Dentistry, Grande Rio University (UNIGRANRIO), Rio de Janeiro, Brazil.
- Deparment of Endodontics, Fluminense Federal University, Niterói, Brazil.
- Department of Endodontics, Rio de Janeiro State University, Rio de Janeiro, Brazil.
- Faculdade de Medicina Dentária, Universidade de Lisboa, Lisboa, Portugal.
- Grupo de Investigação em Bioquimica e Biologia Oral, Unidade de Investigação em Ciências Orais e Biomédicas (UICOB), Faculdade de Medicina Dentária, Universidade de Lisboa, Lisboa, Portugal.
- Centro de Estudo de Medicina Dentária Baseada na Evidência (CEMDBE) - Cochrane Portugal, Faculdade de Medicina Dentária, Universidade de Lisboa, Lisboa, Portugal.
ORCID-Profile: Emmanuel J N L Silva, Jorge N R Martins, Gustavo De-Deus
Förderung
- Fundação Carlos Chagas Filho de Amparo à Pesquisa do Estado do Rio de Janeiro
- Conselho Nacional de Desenvolvimento Científico e Tecnológico
Zitieren
APA
Vieira, V. T. L., Silva, E. J. N. L., Martins, J. N. R., Jaime, A., Puentes, C. G., Torres, A. R., Fernandes, F. M. B., Versiani, M. A., De-Deus, G., & Belladonna, F. G. (2025). Mechanical and Metallurgical In Vitro Evaluation of Electropolished Versus Non‐Electropolished Rotary and Reciprocating Instruments. International Endodontic Journal. https://doi.org/10.1111/iej.70009
Vancouver
Vieira VTL, Silva EJNL, Martins JNR, Jaime A, Puentes CG, Torres AR, et al. Mechanical and Metallurgical In Vitro Evaluation of Electropolished Versus Non‐Electropolished Rotary and Reciprocating Instruments. International Endodontic Journal. 2025. doi:10.1111/iej.70009
BibTeX
@article{Vieira2025,
author = {Vieira, Victor T L and Silva, Emmanuel J N L and Martins, Jorge N R and Jaime, Alejandro and Puentes, Carlos Garcia and Torres, António Roma and Fernandes, Francisco Manuel Braz and Versiani, Marco A and De-Deus, Gustavo and Belladonna, Felipe G},
title = {Mechanical and Metallurgical In Vitro Evaluation of Electropolished Versus Non‐Electropolished Rotary and Reciprocating Instruments},
journal = {International Endodontic Journal},
year = {2025},
doi = {10.1111/iej.70009}
}
EUVIMED