Torsional resistance of additively, subtractively, and conventionally manufactured occlusal devices
Clinical Oral Investigations · 2025
DOI: 10.1007/s00784-025-06362-w ↗Kurz gefasst
Fachartikel in der Zeitschrift Clinical Oral Investigations, 2025. Verfasst von Tina Maleki und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Prosthetic Dentistry, University Hospital, LMU Munich, Goethestrasse 70, 80336, Munich, Germany. Tina.Maleki@med.uni-muenchen.de. Hauptthemen laut MEDLINE: Occlusal Splints, Torsion, Mechanical, Orthodontic Appliance Design, Dental Materials. Die Arbeit nennt 42 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Klinikum der Universität München.
Abstract (Originalsprache)
Objectives
To investigate the torque load (TL) and angular rotation (AR) of additively, subtractively and conventionally manufactured occlusal devices. Materials and methods Specimens ( N = 120) were fabricated from four additive manufacturing resins (GR-10 guide, ProArt Print Splint clear, V-Print Splint, V-Print Splint comfort), five subtractively (BioniCut, EldyPlus, ProArt CAD Splint clear, Temp Premium Flexible, Thermeo) and one conventionally manufactured (Pro Base Cold) materials. The TL and AR were tested initially (24 h, 37 °C, H 2 O) as well as after thermal cycling (5,000 thermal cycles, 5/55°C). Data were analyzed using Kolmogorov–Smirnov, one-way ANOVA, Scheffé post-hoc, t-test, chi-square and Ciba-Geigy table ( p < 0.05).
Results
Initially, the mean TL values ranged from 63.7 to 104 Ncm for additively, 39.2 to 265 Ncm for subtractively, and 204 Ncm for conventionally manufactured materials. The initial mean AR values were 41.7 to 143 deg for additively, 38.4 to 138 deg for subtractively, and 29.3 deg for conventionally manufactured materials. After thermal cycling, the mean TL values ranged from 45.7 to 88.1 Ncm for additively, 31.2 to 246 Ncm for subtractively, and 138 Ncm for conventionally manufactured materials. The mean AR values after aging ranged from 19.9 to 124 deg for additively, 48.1 to 131 deg for subtractively, and 19.5 deg for conventionally manufactured materials.
Conclusions
The torsional resistance of additively, subtractively, and conventionally manufactured materials for occlusal devices varies and is affected by aging processes. Clinical Relevance Material selection for occlusal devices should be guided by the clinical needs of the patients.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Prosthetic Dentistry, University Hospital, LMU Munich, Goethestrasse 70, 80336, Munich, Germany. Tina.Maleki@med.uni-muenchen.de.
- Department of Prosthetic Dentistry, University Hospital, LMU Munich, Goethestrasse 70, 80336, Munich, Germany.
- Department of Conservative Dentistry and Periodontology, University Hospital, LMU Munich, Goethestrasse 70, 80336, Munich, Germany.
Förderung
- Klinikum der Universität München
Zitieren
APA
Maleki, T., Coldea, A., Meinen, J., Reymus, M., Edelhoff, D., & Stawarczyk, B. (2025). Torsional resistance of additively, subtractively, and conventionally manufactured occlusal devices. Clinical Oral Investigations. https://doi.org/10.1007/s00784-025-06362-w
Vancouver
Maleki T, Coldea A, Meinen J, Reymus M, Edelhoff D, Stawarczyk B. Torsional resistance of additively, subtractively, and conventionally manufactured occlusal devices. Clinical Oral Investigations. 2025. doi:10.1007/s00784-025-06362-w
BibTeX
@article{Maleki2025,
author = {Maleki, Tina and Coldea, Andrea and Meinen, John and Reymus, Marcel and Edelhoff, Daniel and Stawarczyk, Bogna},
title = {Torsional resistance of additively, subtractively, and conventionally manufactured occlusal devices},
journal = {Clinical Oral Investigations},
year = {2025},
doi = {10.1007/s00784-025-06362-w}
}
EUVIMED