Stress-Controlled Torsional Fatigue Assessment of Selected Contemporary Endodontic Instruments
Clinical Oral Investigations · 14.03.2026 · 30 (4)
DOI: 10.1007/s00784-026-06811-0 ↗Kurz gefasst
Fachartikel in der Zeitschrift Clinical Oral Investigations, online erschienen am 14.03.2026. Verfasst von Kushagra Ohri und 2 weiteren Autor:innen. Beteiligte Einrichtung: School of Dentistry, The University of Queensland, Brisbane, Australia. Hauptthemen laut MEDLINE: Dental Instruments. Die Arbeit nennt 31 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: The University of Queensland.
Abstract (Originalsprache)
Objectives
To compare stress-controlled torsional fatigue resistance of TruNatomy (TN), ProTaper Next (PTN), and ProTaper Ultimate (PTUL) instruments of comparable sizes at both room and body temperature. Materials and Methods Torsional strength and related twist angles were determined at both temperatures in a torque testing fixture. Stress-controlled torsional fatigue resistance (TFR) was determined by cycling instruments in a custom fixture to pre-set twist angles corresponding to 90% torsional strength until fracture. Data were analyzed using two-way ANOVA and Tukey post-hoc tests with the significance level set at 0.05.
Results
Overall, mean (± standard deviation) TFR ranged from 192 (± 66) to 841 (± 201) cycles; PTUL had the highest TFR and TN the lowest; the differences among the three instruments were statistically significant. At 37 °C compared to room temperature, PTUL had a similar TFR, while PTN had a higher and TN a lower one. Applying customized stress-controlled conditions resulted in differences in TFR among the 3 instrument types. Different environmental temperatures had varying effects on the tested instruments manufactured from differently heat-treated nickel-titanium (NiTi) alloy.
Conclusions
Understanding TFR is relevant for clinical use of NiTi instruments for continuous rotation and reciprocation, specifically for motor settings. Within the limitations and specific testing parameters of this study, stress-controlled TFR varied among the three tested instruments. Different environmental temperatures had varying effects on the tested instruments manufactured from differently heat-treated nickel-titanium (NiTi) alloy. Variations in instrument design and the NiTi alloy type used may account for different torsional fatigue performance and suggest specific clinical usage parameters. Clinical Significance During canal preparation, instruments are stressed under cyclic load, a condition which is replicated in this experiment. The data is relevant for motor presets for continuous rotation and reciprocating motion as well as handling of contemporary heat-treated Nickel-titanium instruments.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- School of Dentistry, The University of Queensland, Brisbane, Australia.
- School of Dentistry, The University of Queensland, Brisbane, Australia. o.peters@uq.edu.au.
- Oral Health Centre, School of Dentistry, The University of Queensland, 288 Herston Road, Herston, Qld, 4006, Australia. o.peters@uq.edu.au.
Förderung
- The University of Queensland
Zitieren
APA
Ohri, K., Chien, P. Y., & Peters, O. A. (2026). Stress-Controlled Torsional Fatigue Assessment of Selected Contemporary Endodontic Instruments. Clinical Oral Investigations, 30(4). https://doi.org/10.1007/s00784-026-06811-0
Vancouver
Ohri K, Chien PY, Peters OA. Stress-Controlled Torsional Fatigue Assessment of Selected Contemporary Endodontic Instruments. Clinical Oral Investigations. 2026;30(4). doi:10.1007/s00784-026-06811-0
BibTeX
@article{Ohri2026,
author = {Ohri, Kushagra and Chien, Philip Yuan-Ho and Peters, Ove A},
title = {Stress-Controlled Torsional Fatigue Assessment of Selected Contemporary Endodontic Instruments},
journal = {Clinical Oral Investigations},
year = {2026},
volume = {30},
number = {4},
doi = {10.1007/s00784-026-06811-0}
}
EUVIMED