DOI: 10.1007/s00464-026-12972-6 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Surgical Endoscopy, online erschienen am 01.07.2026. Verfasst von Yung Lee und 6 weiteren Autor:innen. Beteiligte Einrichtung: Digestive Disease Institute, Cleveland Clinic, 9500 Euclid Ave A30, Cleveland, OH, 44195, USA. Hauptthemen laut MEDLINE: Endoscopic Mucosal Resection, Robotic Surgical Procedures, Gastroscopes. Die Arbeit nennt 24 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background and aims Endoscopic submucosal dissection (ESD) enables en bloc resection of early gastrointestinal neoplasia but remains technically demanding because of limited traction, lack of triangulation, and high operator workload, contributing to a steep learning curve. We evaluated a novel endoluminal robotic platform designed to provide surgical-like bimanual triangulation, stable traction/counter-traction, and improved ergonomics while preserving compatibility with standard gastroscopes.

Methods

The robotic system (Intilume System, Agilis Robotics, Hong Kong, SAR) uses an external positioning cart to drive two 3.5-mm flexible robotic instruments mounted onto a native gastroscope via a cap-and-sheath interface and controlled by compact pen-style motion-tracking controllers enabling seated operation. Available instruments included a bipolar T-knife and a tissue grasper. In a randomized crossover ex vivo porcine stomach study, two gastroenterology fellows without prior ESD experience and two interventional endoscopy fellows with early experience each performed four ESDs (two robotic, two conventional), for a total of 16 procedures. Outcomes included en bloc resection, procedure time, specimen surface area, dissection speed, tissue injury, and operator workload assessed using NASA-TLX, with exploratory OSATS and GEARS evaluations.

Results

All procedures were completed (8 robotic, 8 conventional). Robotic ESD achieved 100% en bloc resection versus 75% with conventional ESD ( p = 0.47), was significantly faster (14.1 ± 4.3 vs 21.6 ± 7.6 min; p = 0.028), and demonstrated higher dissection speed (36.5 ± 23.4 vs 16.3 ± 10.1 mm 2 /min; p = 0.05). No muscular injuries occurred with robotic ESD compared with 5/8 conventional cases; no full-thickness injuries occurred. Operator workload was markedly lower with robotic ESD (NASA-TLX 34.7 ± 24.1 vs 75.0 ± 15.4; p = 0.002).

Conclusion

In a randomized ex vivo model, a bimanual endoluminal robotic platform compatible with standard endoscopes demonstrated promising improvements in ESD efficiency, tissue control, and operator ergonomics compared with conventional ESD. These preliminary findings support further development and progression to first-in-human feasibility evaluation, with potential to facilitate ESD skill acquisition pending clinical validation.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Endoscopic Mucosal ResectionRobotic Surgical ProceduresGastroscopesAnimalsFeasibility StudiesErgonomicsSwineEquipment DesignHumansCross-Over Studies

Stichwörter der Autor:innen

Endoluminal roboticsEndoroboticsEndoscopic submucosal dissectionErgonomicsOperator work loadRobot-assisted endoscopy

Autor:innen und Institutionen

  • Digestive Disease Institute, Cleveland Clinic, 9500 Euclid Ave A30, Cleveland, OH, 44195, USA.
  • University Hospitals Cleveland Medical Center, Case Western Reserve University, Cleveland, OH, USA.
  • Digestive Disease Institute, Cleveland Clinic, 9500 Euclid Ave A30, Cleveland, OH, 44195, USA. bhatta3@ccf.org.

Zitieren

APA

Lee, Y., Park, J. S., Firkins, S., Satoi, S., Fnu, V., Cho, K. J., & Bhatt, A. (2026). A novel bimanual endoluminal robotic platform compatible with standard gastroscopes for endoscopic submucosal dissection: preclinical feasibility and ergonomic evaluation. Surgical Endoscopy, 40(8), 7234-7242. https://doi.org/10.1007/s00464-026-12972-6

Vancouver

Lee Y, Park JS, Firkins S, Satoi S, Fnu V, Cho KJ, et al. A novel bimanual endoluminal robotic platform compatible with standard gastroscopes for endoscopic submucosal dissection: preclinical feasibility and ergonomic evaluation. Surgical Endoscopy. 2026;40(8):7234-7242. doi:10.1007/s00464-026-12972-6

BibTeX
@article{Lee2026,
  author = {Lee, Yung and Park, Ji Seok and Firkins, Stephen and Satoi, Sera and Fnu, Vikash and Cho, Kyungran Justina and Bhatt, Amit},
  title = {A novel bimanual endoluminal robotic platform compatible with standard gastroscopes for endoscopic submucosal dissection: preclinical feasibility and ergonomic evaluation},
  journal = {Surgical Endoscopy},
  year = {2026},
  volume = {40},
  number = {8},
  pages = {7234--7242},
  doi = {10.1007/s00464-026-12972-6}
}
Kostenlos starten →