Simulated vs. real visual performance of an EDOF intraocular lens using adaptive optics technology

Silvia Carrasco-Rojo iD, Luca Giacopinelli iD, Inas Baoud Ould Haddi iD, María García-Montero iD, Pedro Arriola-Villalobos iD, Nuria Garzón iD

Graefe's Archive for Clinical and Experimental Ophthalmology · 2026

DOI: 10.1007/s00417-026-07121-7 ↗

Kurz gefasst

Vergleichsstudie in der Zeitschrift Graefe's Archive for Clinical and Experimental Ophthalmology, 2026. Verfasst von Silvia Carrasco-Rojo und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department Optometry and Vision, Complutense University of Madrid, Madrid, Spain. Hauptthemen laut MEDLINE: Contrast Sensitivity, Depth Perception, Lens Implantation, Intraocular, Optics and Photonics. Die Arbeit nennt 26 Literaturangaben. Gefördert von: Agencia Estatal de Investigación.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

PURPOSE: To evaluate the predictive accuracy of an adaptive optics visual simulator (SimVis Gekko™) in replicating the optical performance of a nondiffractive extended depth-of-focus (EDOF) intraocular lens (IOL), AcrySof® IQ Vivity™ (Alcon, Fort Worth, USA), prior to cataract surgery. METHODS: Prospective, single-center, observational, comparative study. A total of 16 eyes from 16 patients with early-stage cataract underwent simulation of the AcrySof® IQ Vivity™ IOL using the SimVis Gekko™ device prior to phacoemulsification. Postoperative outcomes were assessed 1 month after bilateral implantation of the IOL. Monocular defocus curves, visual acuity (VA) at far, intermediate and near distances, and contrast sensitivity were compared between the simulated and actual postoperative results. Correlation between VA and pupil size was analyzed. RESULTS: The simulated defocus curve closely matched the postoperative performance, with VA differences of less than one line for vergences between − 1.50 D and + 1.00 D. The only point with more than 1 line of difference occurred at − 3.00 D. Simulated contrast sensitivity was slightly lower than postoperative values but followed a similar trend across spatial frequencies. Distance and intermediate visual acuity were significantly better postoperatively than under simulation, while near VA showed no significant differences. A significant Pearson correlation was found between pupil size and near visual acuity. CONCLUSION: The SimVis Gekko™ device demonstrated high accuracy in simulating the visual outcomes of the AcrySof® IQ Vivity™ IOL, suggesting it may be a valuable tool for preoperative patient counseling and personalized lens selection in clinical practice.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Contrast SensitivityDepth PerceptionLens Implantation, IntraocularOptics and PhotonicsPseudophakiaRefraction, OcularVisual AcuityAgedFemaleHumansMaleMiddle AgedFollow-Up StudiesLenses, IntraocularPhacoemulsificationProspective StudiesProsthesis Design

Stichwörter der Autor:innen

Adaptative opticsExtended depth of focusIntraocular lensPresbyopiaVisual simulator

Autor:innen und Institutionen

  • Department Optometry and Vision, Complutense University of Madrid, Madrid, Spain.
  • Department Optometry and Vision, Complutense University of Madrid, Madrid, Spain. ibaoud@ucm.es.
  • Faculty of Optics and Optometry, Complutense University of Madrid, Av. Arcos del Jalón 118, Madrid, E-28037, Spain. ibaoud@ucm.es.
  • Department of Ophthalmology, Clinico San Carlos Hospital, Madrid, Spain.
  • Department of Immunology, Ophthalmology and ENT, Faculty of Medicine, Complutense University of Madrid, IdISSC, Madrid, Spain.

ORCID-Profile: Silvia Carrasco-Rojo, Luca Giacopinelli, Inas Baoud Ould Haddi, María García-Montero, Pedro Arriola-Villalobos, Nuria Garzón

Förderung

  • Agencia Estatal de Investigación

Zitieren

APA

Carrasco-Rojo, S., Giacopinelli, L., Baoud Ould Haddi, I., García-Montero, M., Arriola-Villalobos, P., & Garzón, N. (2026). Simulated vs. real visual performance of an EDOF intraocular lens using adaptive optics technology. Graefe's Archive for Clinical and Experimental Ophthalmology. https://doi.org/10.1007/s00417-026-07121-7

Vancouver

Carrasco-Rojo S, Giacopinelli L, Baoud Ould Haddi I, García-Montero M, Arriola-Villalobos P, Garzón N. Simulated vs. real visual performance of an EDOF intraocular lens using adaptive optics technology. Graefe's Archive for Clinical and Experimental Ophthalmology. 2026. doi:10.1007/s00417-026-07121-7

BibTeX
@article{CarrascoRojo2026,
  author = {Carrasco-Rojo, Silvia and Giacopinelli, Luca and Baoud Ould Haddi, Inas and García-Montero, María and Arriola-Villalobos, Pedro and Garzón, Nuria},
  title = {Simulated vs. real visual performance of an EDOF intraocular lens using adaptive optics technology},
  journal = {Graefe's Archive for Clinical and Experimental Ophthalmology},
  year = {2026},
  doi = {10.1007/s00417-026-07121-7}
}
Kostenlos starten →