Intraoperative ABR measurements via the active middle ear implant vibrant soundbridge using different types of couplers
European Archives of Oto-Rhino-Laryngology · 2025-07 · 282 (7) · 3507-3512
DOI: 10.1007/s00405-025-09248-5 ↗Kurz gefasst
Fachartikel in der Zeitschrift European Archives of Oto-Rhino-Laryngology, online erschienen am 01.02.2025. Verfasst von Mario Cebulla und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Otorhinolaryngology, Head and Neck Surgery, University Hospital Würzburg, Würzburg, Germany. cebulla_m@ukw.de. Hauptthemen laut MEDLINE: Ossicular Prosthesis, Evoked Potentials, Auditory, Brain Stem. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 16 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Universitätsklinikum Würzburg.
Abstract (Originalsprache)
Purpose
The Vibrant Soundbridge active middle ear implant is indicated for the treatment of different types of hearing loss. Recently, a novel system for determining implant performance based on auditory brainstem response (ABR) was developed. A reference value for the expected ABR hearing thresholds is determined on the base of a substantial number of measurements in this work.
Method
ABRs were recorded in patients following VSB surgery. A standard ABR system was employed, utilizing an implemented optimized chirp stimulus, which was transmitted directly to the VSB via the AcoustiAP transmission unit (MED-EL). The study included 104 subjects with different types of hearing loss. Six distinct VSB couplers were utilized in this cohort.
Results
ABR were recorded in all patients. The intraoperative ABR thresholds were 20.1 dB higher compared to the preoperative BC thresholds, 15.9 dB higher compared to the postoperative vibrogram, and 13.4 dB higher compared with the vibrogram thresholds determined at the first fitting. No significant differences were observed between the different couplers in either the investigated methods.
Conclusion
It can be posited that the presented intraoperative measurement reference value enhances the interpretability of ABR in VSB and should be considered, when using this method in VSB surgery.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Otorhinolaryngology, Head and Neck Surgery, University Hospital Würzburg, Würzburg, Germany. cebulla_m@ukw.de.
- Department of Otorhinolaryngology, Head and Neck Surgery, University Hospital Würzburg, Würzburg, Germany.
Förderung
- Universitätsklinikum Würzburg
Zitieren
APA
Cebulla, M., Köstler, C., Herrmann, D. P., & Rak, K. (2025). Intraoperative ABR measurements via the active middle ear implant vibrant soundbridge using different types of couplers. European Archives of Oto-Rhino-Laryngology, 282(7), 3507-3512. https://doi.org/10.1007/s00405-025-09248-5
Vancouver
Cebulla M, Köstler C, Herrmann DP, Rak K. Intraoperative ABR measurements via the active middle ear implant vibrant soundbridge using different types of couplers. European Archives of Oto-Rhino-Laryngology. 2025;282(7):3507-3512. doi:10.1007/s00405-025-09248-5
BibTeX
@article{Cebulla2025,
author = {Cebulla, Mario and Köstler, Carolina and Herrmann, David P and Rak, Kristen},
title = {Intraoperative ABR measurements via the active middle ear implant vibrant soundbridge using different types of couplers},
journal = {European Archives of Oto-Rhino-Laryngology},
year = {2025},
volume = {282},
number = {7},
pages = {3507--3512},
doi = {10.1007/s00405-025-09248-5}
}