Esteem® Active Middle Ear Implant versus Conventional Hearing Aids: Long-Term Performance
Audiology and Neurotology · 2024
DOI: 10.1159/000542158 ↗Kurz gefasst
Vergleichsstudie in der Zeitschrift Audiology and Neurotology, 2024. Verfasst von Maurizio Barbara und 5 weiteren Autor:innen. Beteiligte Einrichtung: Otolaryngology Unit, NESMOS Department, Faculty of Medicine and Psychology, Sapienza University Hospital Sant'Andrea, Rome, Italy. Hauptthemen laut MEDLINE: Hearing Aids, Hearing Loss, Sensorineural, Ossicular Prosthesis. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 22 Literaturangaben.
Abstract (Originalsprache)
Introduction
The totally implantable active middle ear implant Esteem® may be considered an effective alternative to conventional hearing aids (cHAs) to manage moderate-to-severe forms of sensorineural hearing loss. This study aimed to provide long-term comparative data of Esteem performances with those achieved by cHA.
Methods
From a total of 46 subjects who received unilateral application of Esteem®, and were followed up over the years, ten underwent an audiological assessment that compared the outcomes with those achieved in the contralateral ear by a cHA, considering the initially symmetric auditory thresholds in both ears. Other than pure tone audiometry and speech audiometry in quiet, the assessment was performed by using the adaptive speech in noise, i.e., Matrix test.
Results
The mean speech intelligibility in quiet shows in the unaided situation a recognition of 50.7% at 71 dBHL, 71% at 69 dBHL with only contralateral cHA, 92% at 66 dBHL with only Esteem device and 94% at 61 dBHL with Esteem® device and contralateral cHA. The mean speech intelligibility in noise shows in the unaided situation a recognition of 36% at 71 dBHL, 56% at 69 dBHL with only contralateral cHA, 79% at 66 dBHL with only Esteem® device and 84% at 61 dBHL with Esteem® device and contralateral cHA. At Matrix test in the unaided condition, 4 patients reached 50% of intelligibility and the 50% threshold was obtained with a mean sound/noise ratio of +10 dBHL. In the contralaterally aided condition, 10 patients reached a 50% threshold in a condition of mean S/N ratio of +10.6 dBHL. In the Esteem® only and Esteem® plus cHA condition, all patients reached the 50% threshold with a mean S/N ratio of +3.4 dBHL with the Esteem® device and +0.92 dBHL with Esteem® plus a contralateral cHA, with a statistically nonsignificant difference. The mean deviation from the reference value (7.1 dB in the normal hearing population) was 17.1 dBHL, in unaided situation; this condition did not change with only the contralateral cHA (17.6 dBHL), whilst a significant improvement could be identified with only Esteem® device, where the mean deviation was 10.5 dBHL, and mostly with Esteem® device associated with the contralateral cHA, with a value of 8.02 dBHL.
Conclusions
The adaptive speech audiometry in noise (Matrix Test) showed that binaural stimulation provides greater benefits in the speech recognition in noise test in comparison to monaural stimulation, especially when this is carried out only by the cHA. However, the Esteem® device allowed to obtain audiological benefits that are significantly superior to those offered by cHAs, especially in cases where the hearing loss is severe and, in some cases, profound, achieving performances almost comparable to those of a cochlear implant.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Otolaryngology Unit, NESMOS Department, Faculty of Medicine and Psychology, Sapienza University Hospital Sant'Andrea, Rome, Italy.
Zitieren
APA
Barbara, M., Filippi, C., Tarentini, S., Bandiera, G., Monini, S., & Covelli, E. (2024). Esteem® Active Middle Ear Implant versus Conventional Hearing Aids: Long-Term Performance. Audiology and Neurotology. https://doi.org/10.1159/000542158
Vancouver
Barbara M, Filippi C, Tarentini S, Bandiera G, Monini S, Covelli E. Esteem® Active Middle Ear Implant versus Conventional Hearing Aids: Long-Term Performance. Audiology and Neurotology. 2024. doi:10.1159/000542158
BibTeX
@article{Barbara2024,
author = {Barbara, Maurizio and Filippi, Chiara and Tarentini, Silvia and Bandiera, Giorgio and Monini, Simonetta and Covelli, Edoardo},
title = {Esteem® Active Middle Ear Implant versus Conventional Hearing Aids: Long-Term Performance},
journal = {Audiology and Neurotology},
year = {2024},
doi = {10.1159/000542158}
}
EUVIMED