Comparative evaluation of additively manufactured PLA models as cost-effective, non-inferior alternatives to composite bones for pull-out testing
Archives of Orthopaedic and Trauma Surgery · 28.03.2026 · 146 (1)
DOI: 10.1007/s00402-026-06278-4 ↗Kurz gefasst
Vergleichsstudie in der Zeitschrift Archives of Orthopaedic and Trauma Surgery, online erschienen am 28.03.2026. Verfasst von Utku Demirtaş und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Orthopaedics and Traumatology, Sultan 1. Murat State Hospital, Edirne, Turkey. opdrutkudemirtas@gmail.com. Hauptthemen laut MEDLINE: Polyesters, Printing, Three-Dimensional, Materials Testing, Models, Anatomic. Die Arbeit nennt 34 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Sakarya University.
Abstract (Originalsprache)
Introduction
The purpose of this study was to perform biomechanical pull-out comparisons between composite bone blocks and additively manufactured bone models using locked plate and screw systems. The hypothesis posited that the mean pull-out force of the 3D-printed models would demonstrate non-inferior pull-out strength compared to standard composite bones within a clinically acceptable 15% margin. A standardized pull-out mechanism was developed to evaluate its biomechanical reliability and its potential as a customizable alternative. Materials and methods The study compared three groups: standard composite (Sawbones) blocks and two types of additively manufactured polylactic acid (PLA) blocks (grid and gyroid infill pattern), each measuring the same dimensions with 30% infill and mimicking a Sawbones block with 2 mm cortical thickness. A 6-hole LC-DCP titanium plate was fixed onto bone block models with dimensions of 29 × 170 × 42 mm using four screws, each measuring 36 mm in length, which were inserted perpendicularly and tightened to the same torque levels. Steel cables placed under the plate were connected to a mechanical testing setup, and pull-out tests were conducted to evaluate fixation strength. Seven replicate specimens were fabricated to ensure repeatability and reliability of the results.
Results
Mean pull-out forces were 1784.4 ± 79.9 N for Sawbones, 1698.5 ± 134.3 N for Grid PLA ( n = 7), and 1908.7 ± 335.2 N for Gyroid PLA ( n = 7). Statistical analysis using a one-sample t-test showed that both 3D-printed models were non-inferior to the standard Sawbones model within a clinical acceptance margin of 15% ( p < 0.05).
Conclusions
Additively manufactured models exhibit mechanical pull-out performance that is statistically non-inferior to traditional composite bone block models. These findings confirm that 3D-printed PLA-based grid and gyroid structures serve as reliable, cost-effective alternatives for orthopaedic mechanical research without compromising mechanical integrity.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Orthopaedics and Traumatology, Sultan 1. Murat State Hospital, Edirne, Turkey. opdrutkudemirtas@gmail.com.
- Mechanical Engineering Department, Sakarya University, Sakarya, Turkey. ibilgin@sakarya.edu.tr.
- Biomaterials, Energy, Photocatalysis, Enzyme Technology, Nano and Advanced Materials, Additive Manufacturing, Environmental Applications and Sustainability Research and Development Group, Sakarya University, Sakarya, Tu…
- Department of Orthopaedics and Traumatology, Istanbul Medipol University, Istanbul, Turkey.
Förderung
- Sakarya University
Zitieren
APA
Demirtaş, U., İyibilgin, O., Bayam, L., & Genç, E. (2026). Comparative evaluation of additively manufactured PLA models as cost-effective, non-inferior alternatives to composite bones for pull-out testing. Archives of Orthopaedic and Trauma Surgery, 146(1). https://doi.org/10.1007/s00402-026-06278-4
Vancouver
Demirtaş U, İyibilgin O, Bayam L, Genç E. Comparative evaluation of additively manufactured PLA models as cost-effective, non-inferior alternatives to composite bones for pull-out testing. Archives of Orthopaedic and Trauma Surgery. 2026;146(1). doi:10.1007/s00402-026-06278-4
BibTeX
@article{Demirta2026,
author = {Demirtaş, Utku and İyibilgin, Osman and Bayam, Levent and Genç, Erdinç},
title = {Comparative evaluation of additively manufactured PLA models as cost-effective, non-inferior alternatives to composite bones for pull-out testing},
journal = {Archives of Orthopaedic and Trauma Surgery},
year = {2026},
volume = {146},
number = {1},
doi = {10.1007/s00402-026-06278-4}
}
EUVIMED