Biomechanical Performance of Total Wrist Arthrodesis Plates With and Without Arthrodesis of the Carpometacarpal Joint
HAND · 2025
DOI: 10.1177/15589447231198263 ↗Kurz gefasst
Fachartikel in der Zeitschrift HAND, 2025. Verfasst von David H Owen und 8 weiteren Autor:innen. Beteiligte Einrichtung: Canberra Hospital, ACT, Australia. Hauptthemen laut MEDLINE: Arthrodesis, Bone Plates, Carpometacarpal Joints, Wrist Joint. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 26 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: medical device companies.
Abstract (Originalsprache)
Background
It is unknown whether total wrist arthrodesis (TWA) should be performed with or without arthrodesis of the carpometacarpal joint (CMCJ). The aim of this study is to compare CMCJ-spanning TWA plates using 3D printed wrist arthrodesis model with and without arthrodesis of the CMCJ.
Methods
Total wrist arthrodesis plates mounted to 3D printed models were tested under a 4-N bending load at 4 Hz for 50 000 cycles, increased by 15% every 10 000 cycles until failure.
Results
Plates with arthrodesis CMCJ were stiffer and failed at a significantly greater load and number of cycles than plates mounted to models without CMCJ arthrodesis. The Synthes stainless steel locking TWA plate performed better than the Trimed plate applied to the model without CMCJ arthrodesis and the Acumed plate applied to the model with CMCJ arthrodesis. Based on these findings, we recommend arthrodesis of the CMCJ in TWA.
Conclusions
Incorporation of the CMCJ in TWA may protect against plate failure. If arthrodesis of the CMCJ is not performed, plate removal should be considered before breakage occurs. Level of Evidence: IV
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Canberra Hospital, ACT, Australia.
- Australian National University Medical School, Canberra, ACT, Australia.
- The University of Melbourne, VIC, Australia.
- Lismore Base Hospital, NSW, Australia.
- Bond University, Robina, QLD, Australia.
ORCID-Profile: David H Owen
Förderung
- medical device companies
Zitieren
APA
Owen, D. H., Wang, D., Cong, X., Mowbray, C., Perriman, D. M., Roberts, C. J., Smith, P. N., Drobetz, H., & Ackland, D. (2025). Biomechanical Performance of Total Wrist Arthrodesis Plates With and Without Arthrodesis of the Carpometacarpal Joint. HAND. https://doi.org/10.1177/15589447231198263
Vancouver
Owen DH, Wang D, Cong X, Mowbray C, Perriman DM, Roberts CJ, et al. Biomechanical Performance of Total Wrist Arthrodesis Plates With and Without Arthrodesis of the Carpometacarpal Joint. HAND. 2025. doi:10.1177/15589447231198263
BibTeX
@article{Owen2025,
author = {Owen, David H and Wang, Dongyang and Cong, Xu and Mowbray, Cameron and Perriman, Diana M and Roberts, Chris J and Smith, Paul N and Drobetz, Herwig and Ackland, David},
title = {Biomechanical Performance of Total Wrist Arthrodesis Plates With and Without Arthrodesis of the Carpometacarpal Joint},
journal = {HAND},
year = {2025},
doi = {10.1177/15589447231198263}
}
EUVIMED