EUVIMEDEUVIMED entdecken
Vergleichsstudie

Ho: YAG and thulium fiber lasers exhibit comparable temperature distribution: an in vitro evaluation of thermal safety distances for endourological applications—section of EAU endourology

Daniele Robesti, Christian Corsini, Davide Perri, Jia-Lun Kwok, Arman Tsaturyan, Carlotta Nedbal, Luca Villa, Nicola Fossati, Andrea Gallina, Ioannis Kartalas Goumas, Andrea Salonia, Vineet Gauhar und 4 weitere

World Journal of Urology · 2025

DOI: 10.1007/s00345-025-05754-7 ↗

Kurz gefasst

Vergleichsstudie in der Zeitschrift World Journal of Urology, 2025. Verfasst von Daniele Robesti und 15 weiteren Autor:innen. Beteiligte Einrichtung: Vita-Salute San Raffaele University, Milan, Italy. Hauptthemen laut MEDLINE: Thulium, Lasers, Solid-State, Kidney Calices. Die Arbeit wurde bisher 7-mal zitiert (Crossref) und nennt 27 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

INTRODUCTION: The concept of thermal safety distance upon laser activation has been recently introduced. We aimed to compare the thermal safety distance of Ho: YAG and thulium fiber laser (TFL) activation at different setups using an in vitro model. METHODS: The experimental setup used a glass pipette with an open roof bulb in a 37 °C water bath to simulate a renal calyx. We analyzed two scenarios: (1) Laser activation without irrigation; (2) Laser activation at the center of the calyx with irrigation (saline at 40 cmH2O) via a flexible ureteroscope with a 12-14 Fr access sheath. Thermal camera tests were conducted using a Ho: YAG and TFL lasers. A thermal camera monitored the temperature for 60 s of laser activation at 5, 10, and 20 W, with and without irrigation. Each test was repeated three times. The Dewey and Sapareto formula was used to calculate cumulative equivalent minutes at 43 °C (CEM43), with a safety threshold of 120 min. The safety distance was defined as the minimum distance required to overcome the safety threshold. RESULTS: We observed no differences both in recorded temperature values and in their temporal/spatial distribution (all p > 0.05). The safety threshold was not exceeded for either Ho: YAG or TFL lasers at 5 W, 10 W, and 20 W when irrigation was applied, with no statistically significant difference in safety distance between the two lasers (p > 0.05). Similarly, without irrigation, only the 20 W (Ho: YAG: 0.93 ± 0.11 mm, TFL: 0.84 ± 0.09 mm) settings exceeded the safety threshold, and no significant difference in safety distance was found between the lasers (p > 0.05). CONCLUSIONS: No differences in safety distance were observed between TFL and Ho: YAG lasers, regardless of irrigation or power settings. It is recommended for both lasers to remain below 20 W in absence of proper irrigation.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

ThuliumLasers, Solid-StateKidney CalicesHumansIn Vitro TechniquesTemperature

Stichwörter der Autor:innen

In vitro modelLaser temperatureStoneThermal doseThermal safety

Substanzen

Thulium

Autor:innen und Institutionen

  • Vita-Salute San Raffaele University, Milan, Italy.
  • Division of Experimental Oncology/Unit of Urology, URI, IRCCS Ospedale San Raffaele, Milan, Italy.
  • Division of Urology, Sant'Anna Hospital, Via Ravona 20, 22042, San Fermo Della Battaglia, Italy.
  • Department of Urology, Tan Tock Seng Hospital, Singapore, Singapore.
  • Department of Urology and Andrology, Astana Medical University, Astana, Kazakhstan.
  • Urology Unit, Azienda Ospedaliero-Universitaria Delle Marche, Università Politecnica Delle Marche, Ancona, Italy.

Zitieren

APA

Robesti, D., Corsini, C., Perri, D., Kwok, J. L., Tsaturyan, A., Nedbal, C., Villa, L., Fossati, N., Gallina, A., Kartalas Goumas, I., Salonia, A., Gauhar, V., Somani, B. K., Traxer, O., Saccomandi, P., & Ventimiglia, E. (2025). Ho: YAG and thulium fiber lasers exhibit comparable temperature distribution: an in vitro evaluation of thermal safety distances for endourological applications—section of EAU endourology. World Journal of Urology. https://doi.org/10.1007/s00345-025-05754-7

Vancouver

Robesti D, Corsini C, Perri D, Kwok JL, Tsaturyan A, Nedbal C, et al. Ho: YAG and thulium fiber lasers exhibit comparable temperature distribution: an in vitro evaluation of thermal safety distances for endourological applications—section of EAU endourology. World Journal of Urology. 2025. doi:10.1007/s00345-025-05754-7

BibTeX
@article{Robesti2025,
  author = {Robesti, Daniele and Corsini, Christian and Perri, Davide and Kwok, Jia-Lun and Tsaturyan, Arman and Nedbal, Carlotta and Villa, Luca and Fossati, Nicola and Gallina, Andrea and Kartalas Goumas, Ioannis and Salonia, Andrea and Gauhar, Vineet and Somani, Bhaskar K and Traxer, Olivier and Saccomandi, Paola and Ventimiglia, Eugenio},
  title = {Ho: YAG and thulium fiber lasers exhibit comparable temperature distribution: an in vitro evaluation of thermal safety distances for endourological applications—section of EAU endourology},
  journal = {World Journal of Urology},
  year = {2025},
  doi = {10.1007/s00345-025-05754-7}
}
Kostenlos starten →