Does Duty Cycle Modification Affect Temperature During Thulium Fiber Laser Lithotripsy?
Lasers in Surgery and Medicine · 2025
DOI: 10.1002/lsm.70067 ↗Kurz gefasst
Vergleichsstudie in der Zeitschrift Lasers in Surgery and Medicine, 2025. Verfasst von Kallan Richards und 8 weiteren Autor:innen. Beteiligte Einrichtung: Department of Urology, Loma Linda University Health, Loma Linda, California, USA. Hauptthemen laut MEDLINE: Lithotripsy, Laser, Thulium, Hot Temperature, Lasers, Solid-State. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 12 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Objectives
Although the thulium fiber laser (TFL, 1940 nm) efficiently fragments stones, concerns have been raised regarding heat production. Recently, a TFL with a modified duty cycle (MDC) was designed to reduce heat generation. The purpose of this study was to compare heat generation between the MDC and conventional TFL.
Methods
Ten millimeter BegoStone phantoms were placed in the proximal ureter of a kidney and ureter model. The model was submerged in a 35°C saline bath, with continuous saline irrigation (22°C) maintained at 15 mL/min through a ureteroscope. Temperature was measured using a needle thermocouple. Five trials of 60 s continuous laser activation were performed for each power setting: 3 W (0.3 J/10 Hz), 10 W (1 J/10 Hz), 20 W (1 J/20 Hz) and 30 W (0.6 J/50 Hz). Thermal dose was calculated as cumulative equivalent minutes at 43°C (CEM₄₃; injury threshold> 20). Kruskal‐Wallis and Mann‐Whitney U tests were used for statistical analysis with p < 0.05 being significant.
Results
Both lasers produced maximum temperatures ( T max ) of 27°C at 3 W. The MDC TFL generated a significantly lower Tmax than the conventional TFL at 10 W (36.9°C vs 42.1°C, p < 0.001), 20 W (38.1°C vs 44.6°C, p < 0.001), and 30 W (52.5°C vs 63.6°C, p < 0.05). CEM₄₃ for the MDC and conventional TFL at 20 W were (0.00 vs 0.90, respectively), and at 30 W (425 vs 275,919, respectively).
Conclusions
The MDC TFL generated lower temperatures at 10, 20, and 30 W. Future studies are necessary to evaluate stone fragmentation efficiency of the MDC TFL.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Urology, Loma Linda University Health, Loma Linda, California, USA.
ORCID-Profile: Katya Hanessian, Daniel Jhang
Zitieren
APA
Richards, K., Hanessian, K., Albaghli, A., Martin, G., Jhang, D., Battikha, A., Ghoulian, J., Okhunov, Z., & Baldwin, D. D. (2025). Does Duty Cycle Modification Affect Temperature During Thulium Fiber Laser Lithotripsy?. Lasers in Surgery and Medicine. https://doi.org/10.1002/lsm.70067
Vancouver
Richards K, Hanessian K, Albaghli A, Martin G, Jhang D, Battikha A, et al. Does Duty Cycle Modification Affect Temperature During Thulium Fiber Laser Lithotripsy?. Lasers in Surgery and Medicine. 2025. doi:10.1002/lsm.70067
BibTeX
@article{Richards2025,
author = {Richards, Kallan and Hanessian, Katya and Albaghli, Ali and Martin, Gabriel and Jhang, Daniel and Battikha, Adel and Ghoulian, Joshua and Okhunov, Zham and Baldwin, D Duane},
title = {Does Duty Cycle Modification Affect Temperature During Thulium Fiber Laser Lithotripsy?},
journal = {Lasers in Surgery and Medicine},
year = {2025},
doi = {10.1002/lsm.70067}
}
EUVIMED