EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Deep learning super-resolution reconstruction for fast and high-quality cine cardiovascular magnetic resonance

Dmitrij Kravchenko, Alexander Isaak, Narine Mesropyan, Johannes M Peeters, Daniel Kuetting, Claus C Pieper, Christoph Katemann, Ulrike Attenberger, Tilman Emrich, Akos Varga-Szemes, Julian A Luetkens

European Radiology · 23.10.2024 · 35 (5) · 2877-2887

DOI: 10.1007/s00330-024-11145-0 ↗

Kurz gefasst

Fachartikel in der Zeitschrift European Radiology, online erschienen am 23.10.2024. Verfasst von Dmitrij Kravchenko und 10 weiteren Autor:innen. Beteiligte Einrichtung: Department of Diagnostic and Interventional Radiology, University Hospital Bonn, Bonn, Germany. Hauptthemen laut MEDLINE: Deep Learning, Magnetic Resonance Imaging, Cine, Image Interpretation, Computer-Assisted, Image Processing, Computer-Assisted. Die Arbeit wurde bisher 34-mal zitiert (Crossref) und nennt 36 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objectives

To compare standard-resolution balanced steady-state free precession (bSSFP) cine images with cine images acquired at low resolution but reconstructed with a deep learning (DL) super-resolution algorithm. Materials and methods Cine cardiovascular magnetic resonance (CMR) datasets (short-axis and 4-chamber views) were prospectively acquired in healthy volunteers and patients at normal (cine NR : 1.89 × 1.96 mm 2 , reconstructed at 1.04 × 1.04 mm 2 ) and at a low-resolution (2.98 × 3.00 mm 2 , reconstructed at 1.04 × 1.04 mm 2 ). Low-resolution images were reconstructed using compressed sensing DL denoising and resolution upscaling (cine DL ). Left ventricular ejection fraction (LVEF), end-diastolic volume index (LVEDVi), and strain were assessed. Apparent signal-to-noise (aSNR) and contrast-to-noise ratios (aCNR) were calculated. Subjective image quality was assessed on a 5-point Likert scale. Student’s paired t -test, Wilcoxon matched-pairs signed-rank-test, and intraclass correlation coefficient (ICC) were used for statistical analysis.

Results

Thirty participants were analyzed (37 ± 16 years; 20 healthy volunteers and 10 patients). Short-axis views whole-stack acquisition duration of cine DL was shorter than cine NR (57.5 ± 8.7 vs 98.7 ± 12.4 s; p < 0.0001). No differences were noted for: LVEF (59 ± 7 vs 59 ± 7%; ICC: 0.95 [95% confidence interval: 0.94, 0.99]; p = 0.17), LVEDVi (85.0 ± 13.5 vs 84.4 ± 13.7 mL/m 2 ; ICC: 0.99 [0.98, 0.99]; p = 0.12), longitudinal strain (−19.5 ± 4.3 vs −19.8 ± 3.9%; ICC: 0.94 [0.88, 0.97]; p = 0.52), short-axis aSNR (81 ± 49 vs 69 ± 38; p = 0.32), aCNR (53 ± 31 vs 45 ± 27; p = 0.33), or subjective image quality (5.0 [IQR 4.9, 5.0] vs 5.0 [IQR 4.7, 5.0]; p = 0.99).

Conclusion

Deep-learning reconstruction of cine images acquired at a lower spatial resolution led to a decrease in acquisition times of 42% with shorter breath-holds without affecting volumetric results or image quality. Key Points Question Cine CMR acquisitions are time-intensive and vulnerable to artifacts .

Findings

Low-resolution upscaled reconstructions using DL super-resolution decreased acquisition times by 35–42% without a significant difference in volumetric results or subjective image quality . Clinical relevance DL super-resolution reconstructions of bSSFP cine images acquired at a lower spatial resolution reduce acquisition times while preserving diagnostic accuracy, improving the clinical feasibility of cine imaging by decreasing breath hold duration . Graphical Abstract

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Deep LearningMagnetic Resonance Imaging, CineImage Interpretation, Computer-AssistedImage Processing, Computer-AssistedHumansMaleFemaleAdultProspective StudiesMiddle AgedSignal-To-Noise RatioAlgorithms

Stichwörter der Autor:innen

Artificial intelligence (AI)CardiovascularMagnetic resonance imaging (cine)

Autor:innen und Institutionen

  • Department of Diagnostic and Interventional Radiology, University Hospital Bonn, Bonn, Germany.
  • Quantitative Imaging Laboratory Bonn, Bonn, Germany.
  • Division of Cardiovascular Imaging, Department of Radiology and Radiological Science, Medical University of South Carolina, Charleston, SC, USA.
  • Philips MR Clinical Science, Best, Netherlands.
  • Philips GmbH Market DACH, Hamburg, Germany.
  • Department of Diagnostic and Interventional Radiology, University Medical Center of the Johannes Gutenberg-University, Mainz, Germany.

Zitieren

APA

Kravchenko, D., Isaak, A., Mesropyan, N., Peeters, J. M., Kuetting, D., Pieper, C. C., Katemann, C., Attenberger, U., Emrich, T., Varga-Szemes, A., & Luetkens, J. A. (2024). Deep learning super-resolution reconstruction for fast and high-quality cine cardiovascular magnetic resonance. European Radiology, 35(5), 2877-2887. https://doi.org/10.1007/s00330-024-11145-0

Vancouver

Kravchenko D, Isaak A, Mesropyan N, Peeters JM, Kuetting D, Pieper CC, et al. Deep learning super-resolution reconstruction for fast and high-quality cine cardiovascular magnetic resonance. European Radiology. 2024;35(5):2877-2887. doi:10.1007/s00330-024-11145-0

BibTeX
@article{Kravchenko2024,
  author = {Kravchenko, Dmitrij and Isaak, Alexander and Mesropyan, Narine and Peeters, Johannes M and Kuetting, Daniel and Pieper, Claus C and Katemann, Christoph and Attenberger, Ulrike and Emrich, Tilman and Varga-Szemes, Akos and Luetkens, Julian A},
  title = {Deep learning super-resolution reconstruction for fast and high-quality cine cardiovascular magnetic resonance},
  journal = {European Radiology},
  year = {2024},
  volume = {35},
  number = {5},
  pages = {2877--2887},
  doi = {10.1007/s00330-024-11145-0}
}
Kostenlos starten →