Robustness of radiomics among photon-counting detector CT and dual-energy CT systems: a texture phantom study

Lan Zhu, Haipeng Dong, Jing Sun, Lingyun Wang, Yue Xing, Yangfan Hu, Junjie Lu, Jiarui Yang, Jingshen Chu, Chao Yan, Fei Yuan, Jingyu Zhong

European Radiology · 24.07.2024 · 35 (2) · 871-884

DOI: 10.1007/s00330-024-10976-1 ↗

Kurz gefasst

Fachartikel in der Zeitschrift European Radiology, online erschienen am 24.07.2024. Verfasst von Lan Zhu und 11 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiology, Ruijin Hospital, Shanghai Jiao Tong University School of Medicine, Shanghai, 200025, China. Hauptthemen laut MEDLINE: Phantoms, Imaging, Tomography, X-Ray Computed, Radiography, Dual-Energy Scanned Projection. Die Arbeit wurde bisher 19-mal zitiert (Crossref) und nennt 63 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: National Natural Science Foundation of China, Yangfan Project of Science and Technology Commission of Shanghai Municipality.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objectives

To evaluate the robustness of radiomics features among photon-counting detector CT (PCD-CT) and dual-energy CT (DECT) systems.

Methods

A texture phantom consisting of twenty-eight materials was scanned with one PCD-CT and four DECT systems (dual-source, rapid kV-switching, dual-layer, and sequential scanning) at three dose levels twice. Thirty sets of virtual monochromatic images at 70 keV were reconstructed. Regions of interest were delineated for each material with a rigid registration. Ninety-three radiomics were extracted per PyRadiomics. The test-retest repeatability between repeated scans was assessed by Bland-Altman analysis. The intra-system reproducibility between dose levels, and inter-system reproducibility within the same dose level, were evaluated by intraclass correlation coefficient (ICC) and concordance correlation coefficient (CCC). Inter-system variability among five scanners was assessed by coefficient of variation (CV) and quartile coefficient of dispersion (QCD).

Results

The test–retest repeatability analysis presented that 97.1% of features were repeatable between scan–rescans. The mean ± standard deviation ICC and CCC were 0.945 ± 0.079 and 0.945 ± 0.079 for intra-system reproducibility, respectively, and 86.0% and 85.7% of features were with ICC > 0.90 and CCC > 0.90, respectively, between different dose levels. The mean ± standard deviation ICC and CCC were 0.157 ± 0.174 and 0.157 ± 0.174 for inter-system reproducibility, respectively, and none of the features were with ICC > 0.90 or CCC > 0.90 within the same dose level. The inter-system variability suggested that 6.5% and 12.8% of features were with CV < 10% and QCD < 10%, respectively, among five CT systems.

Conclusion

The radiomics features were non-reproducible with significant variability in values among different CT techniques. Clinical relevance statement Radiomics features are non-reproducible with significant variability in values among photon-counting detector CT and dual-energy CT systems, necessitating careful attention to improve the cross-system generalizability of radiomic features before implementation of radiomics analysis in clinical routine. Key Points CT radiomics stability should be guaranteed before the implementation in the clinical routine. Radiomics robustness was on a low level among photon-counting detectors and dual-energy CT techniques. Limited inter-system robustness of radiomic features may impact the generalizability of models.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Phantoms, ImagingTomography, X-Ray ComputedRadiography, Dual-Energy Scanned ProjectionReproducibility of ResultsRadiation DosagePhotonsHumansRadiomics

Stichwörter der Autor:innen

Multidetector computed tomographyPhantoms (imaging)RadiomicsReproducibility of results

Autor:innen und Institutionen

  • Department of Radiology, Ruijin Hospital, Shanghai Jiao Tong University School of Medicine, Shanghai, 200025, China.
  • Department of General Surgery, Shanghai Minimally Invasive Surgery Center, Ruijin Hospital, Shanghai Jiao Tong University School of Medicine, Shanghai, 200025, China.
  • Department of Imaging, Tongren Hospital, Shanghai Jiao Tong University School of Medicine, Shanghai, 200336, China.
  • Department of Epidemiology and Population Health, Stanford University School of Medicine, Stanford, CA, 94305, USA.
  • Department of Biomedical Engineering, Boston University, Boston, MA, 02215, USA.
  • Department of Science and Technology Development, Ruijin Hospital, Shanghai Jiao Tong University School of Medicine, Shanghai, 200025, China.

Förderung

  • National Natural Science Foundation of China
  • Yangfan Project of Science and Technology Commission of Shanghai Municipality

Zitieren

APA

Zhu, L., Dong, H., Sun, J., Wang, L., Xing, Y., Hu, Y., Lu, J., Yang, J., Chu, J., Yan, C., Yuan, F., & Zhong, J. (2024). Robustness of radiomics among photon-counting detector CT and dual-energy CT systems: a texture phantom study. European Radiology, 35(2), 871-884. https://doi.org/10.1007/s00330-024-10976-1

Vancouver

Zhu L, Dong H, Sun J, Wang L, Xing Y, Hu Y, et al. Robustness of radiomics among photon-counting detector CT and dual-energy CT systems: a texture phantom study. European Radiology. 2024;35(2):871-884. doi:10.1007/s00330-024-10976-1

BibTeX
@article{Zhu2024,
  author = {Zhu, Lan and Dong, Haipeng and Sun, Jing and Wang, Lingyun and Xing, Yue and Hu, Yangfan and Lu, Junjie and Yang, Jiarui and Chu, Jingshen and Yan, Chao and Yuan, Fei and Zhong, Jingyu},
  title = {Robustness of radiomics among photon-counting detector CT and dual-energy CT systems: a texture phantom study},
  journal = {European Radiology},
  year = {2024},
  volume = {35},
  number = {2},
  pages = {871--884},
  doi = {10.1007/s00330-024-10976-1}
}
Kostenlos starten →