Paediatric high-pitch lung imaging with photon-counting detector computed tomography: a dose reduction phantom study

Michael Zellner, Thomas Sartoretti, Thomas Flohr, Thomas Frauenfelder, Hatem Alkadhi, Christian J Kellenberger, Victor Mergen

Pediatric Radiology · 2025

DOI: 10.1007/s00247-025-06235-0 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Pediatric Radiology, 2025. Verfasst von Michael Zellner und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Diagnostic Imaging, University Children's Hospital Zurich, Lenggstrasse 30, Zurich, 8008, Switzerland. michael.zellner@kispi.uzh.ch. Hauptthemen laut MEDLINE: Phantoms, Imaging, Radiation Dosage, Tomography, X-Ray Computed, Lung. Die Arbeit wurde bisher 7-mal zitiert (Crossref) und nennt 25 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: University of Zurich.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Photon-counting detector computed tomography (PCD-CT) can reduce radiation dose in paediatric lung imaging.

Objective

The aim of this study was to determine the lowest radiation dose maintaining adequate image quality for high-pitch lung imaging using a PCD-CT in a chest phantom replicating the characteristics of a 5-year-old child. Materials and methods The phantom was imaged on a dual-source PCD-CT with five different volume CT dose indices (CTDI vol ): 0.45 mGy, 0.30 mGy, 0.15 mGy, 0.07 mGy, and 0.01 mGy. Scans were acquired with Sn100 kV in standard and ultra-high resolution modes. Polychromatic images were reconstructed with a 1-mm slice thickness, lung kernel Bl60, without quantum iterative reconstruction and with quantum iterative reconstruction at strengths 2 and 4. Two paediatric radiologists rated reconstructions subjectively, defining adequate image quality as the visibility of small peripheral structures. Objective evaluation included global noise index and global signal-to-noise ratio index.

Results

Exposure times were 0.42 s and 0.84 s for standard and ultra-high resolution modes, respectively. Subjective assessments showed no significant differences across scan modes or quantum iterative reconstruction strengths for both readers at all doses (all, P > 0.05). Scans at 0.07 mGy with quantum iterative reconstruction 4 were deemed to maintain adequate image quality at the lowest dose. Global noise index was always lower and global signal-to-noise ratio index always higher in ultra-high resolution compared with standard mode, underscoring noise reduction achieved via ultra-high resolution mode’s small pixel effect.

Conclusions

PCD-CT enables high-pitch lung imaging while maintaining adequate image quality at a radiation dose as low as 0.07 mGy, with quantum iterative reconstruction 4, in a paediatric phantom representing a 5-year-old child. Graphical Abstract

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Phantoms, ImagingRadiation DosageTomography, X-Ray ComputedLungRadiation ProtectionHumansChild, PreschoolPhotonsRadiographic Image Interpretation, Computer-AssistedSensitivity and SpecificityReproducibility of Results

Stichwörter der Autor:innen

Paediatric imagingPaediatric lungPhantom studyPhoton-counting detectorRadiation dose

Autor:innen und Institutionen

  • Department of Diagnostic Imaging, University Children's Hospital Zurich, Lenggstrasse 30, Zurich, 8008, Switzerland. michael.zellner@kispi.uzh.ch.
  • Diagnostic and Interventional Radiology, University Hospital Zurich, University Zurich, Zurich, Switzerland.
  • Department of Diagnostic Imaging, University Children's Hospital Zurich, Lenggstrasse 30, Zurich, 8008, Switzerland.

Förderung

  • University of Zurich

Zitieren

APA

Zellner, M., Sartoretti, T., Flohr, T., Frauenfelder, T., Alkadhi, H., Kellenberger, C. J., & Mergen, V. (2025). Paediatric high-pitch lung imaging with photon-counting detector computed tomography: a dose reduction phantom study. Pediatric Radiology. https://doi.org/10.1007/s00247-025-06235-0

Vancouver

Zellner M, Sartoretti T, Flohr T, Frauenfelder T, Alkadhi H, Kellenberger CJ, et al. Paediatric high-pitch lung imaging with photon-counting detector computed tomography: a dose reduction phantom study. Pediatric Radiology. 2025. doi:10.1007/s00247-025-06235-0

BibTeX
@article{Zellner2025,
  author = {Zellner, Michael and Sartoretti, Thomas and Flohr, Thomas and Frauenfelder, Thomas and Alkadhi, Hatem and Kellenberger, Christian J and Mergen, Victor},
  title = {Paediatric high-pitch lung imaging with photon-counting detector computed tomography: a dose reduction phantom study},
  journal = {Pediatric Radiology},
  year = {2025},
  doi = {10.1007/s00247-025-06235-0}
}
Kostenlos starten →