Ultra-low-dose vs. standard-of-care-dose CT of the chest in patients with post-COVID-19 conditions—a prospective intra-patient multi-reader study
European Radiology · 2024
DOI: 10.1007/s00330-024-10754-z ↗Kurz gefasst
Vergleichsstudie in der Zeitschrift European Radiology, 2024. Verfasst von Christian Wassipaul und 15 weiteren Autor:innen. Beteiligte Einrichtung: Department of Biomedical Imaging and Image-guided Therapy, Medical University of Vienna, Vienna, Austria. Hauptthemen laut MEDLINE: COVID-19, Radiation Dosage, Tomography, X-Ray Computed, SARS-CoV-2. Die Arbeit wurde bisher 7-mal zitiert (Crossref) und nennt 51 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Objectives
To conduct an intrapatient comparison of ultra-low-dose computed tomography (ULDCT) and standard-of-care-dose CT (SDCT) of the chest in terms of the diagnostic accuracy of ULDCT and intrareader agreement in patients with post-COVID conditions.
Methods
We prospectively included 153 consecutive patients with post-COVID-19 conditions. All participants received an SDCT and an additional ULDCT scan of the chest. SDCTs were performed with standard imaging parameters and ULDCTs at a fixed tube voltage of 100 kVp (with tin filtration), 50 ref. mAs (dose modulation active), and iterative reconstruction algorithm level 5 of 5. All CT scans were separately evaluated by four radiologists for the presence of lung changes and their consistency with post-COVID lung abnormalities. Radiation dose parameters and the sensitivity, specificity, and accuracy of ULDCT were calculated.
Results
Of the 153 included patients (mean age 47.4 ± 15.3 years; 48.4% women), 45 (29.4%) showed post-COVID lung abnormalities. In those 45 patients, the most frequently detected CT patterns were ground-glass opacities (100.0%), reticulations (43.5%), and parenchymal bands (37.0%). The accuracy, sensitivity, and specificity of ULDCT compared to SDCT for the detection of post-COVID lung abnormalities were 92.6, 87.2, and 94.9%, respectively. The median total dose length product (DLP) of ULDCTs was less than one-tenth of the radiation dose of our SDCTs (12.6 mGy*cm [9.9; 15.5] vs. 132.1 mGy*cm [103.9; 160.2]; p < 0.001).
Conclusion
ULDCT of the chest offers high accuracy in the detection of post-COVID lung abnormalities compared to an SDCT scan at less than one-tenth the radiation dose, corresponding to only twice the dose of a standard chest radiograph in two views. Clinical relevance statement Ultra-low-dose CT of the chest may provide a favorable, radiation-saving alternative to standard-dose CT in the long-term follow-up of the large patient cohort of post-COVID-19 patients.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Biomedical Imaging and Image-guided Therapy, Medical University of Vienna, Vienna, Austria.
- Department of Radiology, UMass Memorial Medical Center and University of Massachusetts Chan Medical School, Worcester, MA, USA.
- Imaging Verbund, Vienna, Austria.
- contextflow GmbH, Vienna, Austria.
- Center for Medical Physics and Biomedical Engineering, Medical University of Vienna, Vienna, Austria.
- Department of Diagnostic and Interventional Radiology, Clinic Donaustadt, Vienna Healthcare Group, Vienna, Austria.
Zitieren
APA
Wassipaul, C., Kifjak, D., Milos, R. I., Prayer, F., Roehrich, S., Winter, M., Beer, L., Watzenboeck, M. L., Pochepnia, S., Weber, M., Tamandl, D., Homolka, P., Birkfellner, W., Ringl, H., Prosch, H., & Heidinger, B. H. (2024). Ultra-low-dose vs. standard-of-care-dose CT of the chest in patients with post-COVID-19 conditions—a prospective intra-patient multi-reader study. European Radiology. https://doi.org/10.1007/s00330-024-10754-z
Vancouver
Wassipaul C, Kifjak D, Milos RI, Prayer F, Roehrich S, Winter M, et al. Ultra-low-dose vs. standard-of-care-dose CT of the chest in patients with post-COVID-19 conditions—a prospective intra-patient multi-reader study. European Radiology. 2024. doi:10.1007/s00330-024-10754-z
BibTeX
@article{Wassipaul2024,
author = {Wassipaul, Christian and Kifjak, Daria and Milos, Ruxandra-Iulia and Prayer, Florian and Roehrich, Sebastian and Winter, Melanie and Beer, Lucian and Watzenboeck, Martin L and Pochepnia, Svitlana and Weber, Michael and Tamandl, Dietmar and Homolka, Peter and Birkfellner, Wolfgang and Ringl, Helmut and Prosch, Helmut and Heidinger, Benedikt H},
title = {Ultra-low-dose vs. standard-of-care-dose CT of the chest in patients with post-COVID-19 conditions—a prospective intra-patient multi-reader study},
journal = {European Radiology},
year = {2024},
doi = {10.1007/s00330-024-10754-z}
}
EUVIMED