Integrated Needle Guide on Smartphone for Percutaneous Interventions Using Augmented Reality
CardioVascular and Interventional Radiology · 2025
DOI: 10.1007/s00270-025-04044-4 ↗Kurz gefasst
Fachartikel in der Zeitschrift CardioVascular and Interventional Radiology, 2025. Verfasst von Laetitia Saccenti und 14 weiteren Autor:innen. Beteiligte Einrichtung: Center for Interventional Oncology, Radiology and Imaging Sciences, Clinical Center, National Institutes of Health, 10 Center Drive, Bethesda, MD, 20…. Hauptthemen laut MEDLINE: Smartphone, Augmented Reality, Needles, Radiography, Interventional. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 31 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: NIH Clinical Center, Intramural NIH HHS, National Institutes of Health Clinical Center.
Abstract (Originalsprache)
Purpose
This study aimed to describe the workflow and evaluate the accuracy of a novel smartphone augmented reality (AR) application that includes an integrated needle guide, in a phantom. Materials and Methods A smartphone cover with an integrated needle guide was designed and 3D-printed. An AR application for percutaneous application was developed, which integrated a projected needle path based on the rigid needle guide. After planning the needle path using this novel tool, the operator could place the needle through the guide to reach the target. Six lesions with out-of-plane entry points were targeted on an abdominal phantom. Timing and accuracy of needle placements were measured on post-procedural CT both using smartphone AR guidance and with a freehand approach. Results were compared using Wilcoxon rank sum and Pearson’s chi-squared tests.
Results
A total of 108 needle placements were performed by 9 physicians with widely varying experience. The median accuracy was 4 mm (IQR 3-6 mm) using the smartphone versus 18 mm (IQR 9-27 mm) for freehand ( P < 0.001). Using the smartphone AR application, planning time was 91 s (IQR 71-151 s), and puncture time was 68 s (IQR 57-77 s). There was no difference in accuracy, planning, or puncture times according to experience level when using the AR tool.
Conclusion
This smartphone application with an integrated needle guide allows path planning and accurate needle placement on phantoms with real-time angular feedback in less than 3 min. This technology could promote standardization, reduce experience requirements, or provide accurate low-cost guidance in environments without procedural imaging for percutaneous interventions. Graphical Abstract
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Center for Interventional Oncology, Radiology and Imaging Sciences, Clinical Center, National Institutes of Health, 10 Center Drive, Bethesda, MD, 20892, USA.
- Henri Mondor's Institute of Biomedical Research - Inserm, U955 Team N°18, Creteil, France.
- Philips Research of North America, Cambridge, MA, USA.
- Department of Radiology, Henri Mondor University Hospital, Assistance Publique - Hopitaux de Paris (AP-HP), Creteil, France.
- Center for Interventional Oncology, Radiology and Imaging Sciences, Clinical Center, National Institutes of Health, 10 Center Drive, Bethesda, MD, 20892, USA. bwood@cc.nih.gov.
Förderung
- NIH Clinical Center
- Intramural NIH HHS
- National Institutes of Health Clinical Center
Zitieren
APA
Saccenti, L., Varble, N., Borde, T., Huth, H., Kassin, M., Ukeh, I., Bakhutashvili, I., Golbus, A., Hazen, L., Xu, S., Tacher, V., Kobeiter, H., Horton, K., Li, M., & Wood, B. (2025). Integrated Needle Guide on Smartphone for Percutaneous Interventions Using Augmented Reality. CardioVascular and Interventional Radiology. https://doi.org/10.1007/s00270-025-04044-4
Vancouver
Saccenti L, Varble N, Borde T, Huth H, Kassin M, Ukeh I, et al. Integrated Needle Guide on Smartphone for Percutaneous Interventions Using Augmented Reality. CardioVascular and Interventional Radiology. 2025. doi:10.1007/s00270-025-04044-4
BibTeX
@article{Saccenti2025,
author = {Saccenti, Laetitia and Varble, Nicole and Borde, Tabea and Huth, Hannah and Kassin, Michael and Ukeh, Ifechi and Bakhutashvili, Ivan and Golbus, Ashley and Hazen, Lindsey and Xu, Sheng and Tacher, Vania and Kobeiter, Hicham and Horton, Keith and Li, Ming and Wood, Bradford},
title = {Integrated Needle Guide on Smartphone for Percutaneous Interventions Using Augmented Reality},
journal = {CardioVascular and Interventional Radiology},
year = {2025},
doi = {10.1007/s00270-025-04044-4}
}
EUVIMED