Comparative strain‑gauge analysis of Gamma3 and Gamma3 reconstruction nails in unstable basicervical femur fractures: a pilot biomechanical study
International Orthopaedics · 2026-07 · 50 (7) · 1731-1739
DOI: 10.1007/s00264-026-06872-6 ↗Kurz gefasst
Vergleichsstudie in der Zeitschrift International Orthopaedics, online erschienen am 02.06.2026. Verfasst von Jose Antonio Valle-Cruz und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Traumatology and Orthopaedic Surgery, Hospital Clínico San Carlos, Madrid, Spain. Hauptthemen laut MEDLINE: Bone Nails, Fracture Fixation, Intramedullary, Fracture Fixation, Internal. Die Arbeit nennt 21 Literaturangaben.
Abstract (Originalsprache)
PURPOSE: Evidence supporting the mechanical superiority of the Gamma3 RC nail over the standard Gamma3 configuration remains limited. This pilot biomechanical study compared strain distribution and load-transfer behavior between both implants in a standardized unstable basicervical proximal femoral fracture model. METHODS: Six fourth-generation composite femora were allocated into three groups: one intact reference specimen, three fractures stabilized with a standard Gamma3 nail, and three fractures stabilized with a Gamma3 RC nail incorporating a U-blade lag screw. Mechanical testing consisted of controlled low-load axial-oblique loading under standardized conditions. Four synchronized strain-gauge rosettes recorded cortical deformation at the proximal femur. RESULTS: The Gamma3 RC construct demonstrated consistently lower strain magnitudes at the medial cortices and preserved a more linear strain-load response throughout the tested loading range compared with the standard Gamma3 configuration (Δε = 18.6 με and - 26.5 με). Differences between constructs became progressively more evident with increasing load (Δε = 180.9 με and - 62.9 με), particularly at the proximal medial cortex, where the standard Gamma3 showed greater strain concentration and early nonlinearity. CONCLUSION: In this exploratory synthetic-femur model, the Gamma3 RC nail demonstrated a mechanical trend toward more homogeneous load distribution and improved rotational control compared with the standard Gamma3 nail.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Traumatology and Orthopaedic Surgery, Hospital Clínico San Carlos, Madrid, Spain.
- Department of Surgery, Complutense University of Madrid, Madrid, Spain.
- Department of Traumatology and Orthopaedic Surgery, Hospital Clínico San Carlos, Madrid, Spain. danielgarriguez@gmail.com.
- Department of Mechanical Engineering, ETSII, Technical University of Madrid, Madrid, Spain.
Zitieren
APA
Valle-Cruz, J. A., Lopiz, Y., Garríguez-Pérez, D., Francés-Borrego, A., Ros Felip, A., & Marco, F. (2026). Comparative strain‑gauge analysis of Gamma3 and Gamma3 reconstruction nails in unstable basicervical femur fractures: a pilot biomechanical study. International Orthopaedics, 50(7), 1731-1739. https://doi.org/10.1007/s00264-026-06872-6
Vancouver
Valle-Cruz JA, Lopiz Y, Garríguez-Pérez D, Francés-Borrego A, Ros Felip A, Marco F. Comparative strain‑gauge analysis of Gamma3 and Gamma3 reconstruction nails in unstable basicervical femur fractures: a pilot biomechanical study. International Orthopaedics. 2026;50(7):1731-1739. doi:10.1007/s00264-026-06872-6
BibTeX
@article{ValleCruz2026,
author = {Valle-Cruz, Jose Antonio and Lopiz, Yaiza and Garríguez-Pérez, Daniel and Francés-Borrego, Alberto and Ros Felip, Antonio and Marco, Fernando},
title = {Comparative strain‑gauge analysis of Gamma3 and Gamma3 reconstruction nails in unstable basicervical femur fractures: a pilot biomechanical study},
journal = {International Orthopaedics},
year = {2026},
volume = {50},
number = {7},
pages = {1731--1739},
doi = {10.1007/s00264-026-06872-6}
}