EUVIMEDEUVIMED entdecken
VergleichsstudieOpen Access · CC BY 4.0

Comparison of the Dynamic Cut‐Out Failure Modes of Common Proximal Femoral Fixation Devices Using a Mesh‐Free Computational Method

Erica Ueda Boles, Sloan Kulper, Katie Whiffin, Marilyn Janice Oentaryo iD, Kerstin Schneider, Frankie Leung, Christian X Fang iD

Journal of Orthopaedic Research · 2026-02 · 44 (2)

DOI: 10.1002/jor.70159 ↗

Kurz gefasst

Vergleichsstudie in der Zeitschrift Journal of Orthopaedic Research, online erschienen am 10.02.2026. Verfasst von Erica Ueda Boles und 6 weiteren Autor:innen. Beteiligte Einrichtung: Lifespans Ltd., Hong Kong, China. Hauptthemen laut MEDLINE: Fracture Fixation, Internal, Proximal Femoral Fractures, Hip Fractures. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 73 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Migration and cut‐out are commo n failures observed in internal fixation of intertrochanteric fractures. This study compares the performance of four typical devices under dynamic loading using benchtop and computer‐simulated fracture models. The dynamic hip screw (DHS), Gamma3, proximal femoral nail anti‐rotation (PFNA‐II), and TFN‐Advanced (TFNA) were inserted into a solid rigid 10 pounds per cubic foot (PCF) polyurethane bone foam mimicking a reduced unstable intertrochanteric fracture model. Static and dynamic loading tests using a double‐peak loading curve based on natural walking gait were conducted. The blade devices had higher resistance to the onset of implant migration under physical dynamic loading in comparison to the screw devices. However, as dynamic loading progressed, no implant showed clear superior performance. Mesh‐free simulations of the physical tests were then conducted to visualize stress and failure patterns within the bone foam during migration and cut‐out, using an experimentally validated porous foam material model. The mean concordance correlation coefficient between load to cut‐out for the physical and simulated tests was 0.953 and 0.858 under static and dynamic loading, respectively. Under simulated dynamic loading, the volumes of yielding and failed bone foam were 15%, 32%, and 25% higher for Gamma3, PFNA‐II, and TFNA than for DHS, suggesting that devices with similar forces at cut‐out may produce relatively large differences in the volume of damaged bone tissue. Accurate simulation of material damage during dynamic loading may offer a useful alternative method for differentiating device performance in support of clinical decision‐making around implant selection.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Fracture Fixation, InternalProximal Femoral FracturesHip FracturesBone ScrewsHumansComputer SimulationBone NailsProsthesis FailureBiomechanical Phenomena

Stichwörter der Autor:innen

cut‐outdynamicinternal fixationmesh‐freemigrationproximal femursimulation

Autor:innen und Institutionen

  • Lifespans Ltd., Hong Kong, China.
  • Department of Orthopaedics & Traumatology, LKS Faculty of Medicine, The University of Hong Kong, Hong Kong, China.
  • School of Biomedical Engineering, The University of Hong Kong, Hong Kong, China.
  • Orthopaedic Surgeon, Zurich, Switzerland.

ORCID-Profile: Marilyn Janice Oentaryo, Christian X Fang

Zitieren

APA

Boles, E. U., Kulper, S., Whiffin, K., Oentaryo, M. J., Schneider, K., Leung, F., & Fang, C. X. (2026). Comparison of the Dynamic Cut‐Out Failure Modes of Common Proximal Femoral Fixation Devices Using a Mesh‐Free Computational Method. Journal of Orthopaedic Research, 44(2). https://doi.org/10.1002/jor.70159

Vancouver

Boles EU, Kulper S, Whiffin K, Oentaryo MJ, Schneider K, Leung F, et al. Comparison of the Dynamic Cut‐Out Failure Modes of Common Proximal Femoral Fixation Devices Using a Mesh‐Free Computational Method. Journal of Orthopaedic Research. 2026;44(2). doi:10.1002/jor.70159

BibTeX
@article{Boles2026,
  author = {Boles, Erica Ueda and Kulper, Sloan and Whiffin, Katie and Oentaryo, Marilyn Janice and Schneider, Kerstin and Leung, Frankie and Fang, Christian X},
  title = {Comparison of the Dynamic Cut‐Out Failure Modes of Common Proximal Femoral Fixation Devices Using a Mesh‐Free Computational Method},
  journal = {Journal of Orthopaedic Research},
  year = {2026},
  volume = {44},
  number = {2},
  doi = {10.1002/jor.70159}
}
Kostenlos starten →