EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Photon-counting computed tomography for paediatric congenital heart defects yields images of high diagnostic quality with low radiation doses at both 70 kV and 90 kV

Fredrik Stålhammar, Marie-Louise Aurumskjöld iD, Sofie Meyer, Marie Wiklund, Pär Wingren, Petru Liuba iD, Erik Hedström iD

Pediatric Radiology · 2024

DOI: 10.1007/s00247-024-05939-z ↗

Kurz gefasst

Fachartikel in der Zeitschrift Pediatric Radiology, 2024. Verfasst von Fredrik Stålhammar und 6 weiteren Autor:innen. Beteiligte Einrichtung: Diagnostic Radiology, Department of Clinical Sciences Lund, Lund University, Lund, Sweden. Hauptthemen laut MEDLINE: Heart Defects, Congenital, Radiation Dosage, Tomography, X-Ray Computed. Die Arbeit wurde bisher 17-mal zitiert (Crossref) und nennt 18 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Hjärt-Lungfonden, the Swedish governmental funding of clinical research, Region Skåne.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Photon-counting computed tomography (PCCT) is a new clinical method that may show better diagnostic quality at lower radiation doses than conventional CT.

Objective

To investigate the diagnostic quality and radiation dose of paediatric cardiovascular PCCT for diagnosis of congenital heart defects at 70 kV and 90 kV. Materials and methods This retrospective assessment included clinical non-gated paediatric PCCT examinations for assessment of congenital heart defects. Radiation doses were recorded, and overall and specific diagnostic quality (1–4) were scored by four paediatric radiologists. Agreement, differences, and trends were assessed by percent rater agreement, intraclass correlation, Mann–Whitney tests, and Jonckheere-Terpstra tests.

Results

Seventy children with congenital heart defects were examined at 70 kV ( n = 35; age 2 days–16 years; 63% boys) or 90 kV ( n = 35; age 2 days–17 years; 51% boys). All observers gave a median score of 4 (high diagnostic quality) for both 70 kV and 90 kV, with no difference in median values between tube voltages (all P > 0.06). Agreement for overall scores was 66–94% for 70 kV and 60–77% for 90 kV. Agreement for specific scores was 80–97% for 70 kV and 83–89% for 90 kV. Size-dependent dose estimate was 0.68 mGy (0.25–2.02 mGy) for 70 kV and 1.10 mGy (0.58–2.71 mGy; P < 0.001) for 90 kV. Effective dose was 0.30 mSv (0.15–0.82 mSv) for 70 kV and 0.39 mSv (0.22–1.51 mSv; P = 0.01) for 90 kV.

Conclusion

Paediatric cardiovascular PCCT yields images for congenital heart defects of high diagnostic quality with low radiation dose at both 70 kV and 90 kV.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Heart Defects, CongenitalRadiation DosageTomography, X-Ray ComputedHumansFemaleMaleChildInfantChild, PreschoolInfant, NewbornAdolescentRetrospective StudiesPhotonsReproducibility of ResultsSensitivity and Specificity

Stichwörter der Autor:innen

DiagnosisHeart defects, congenitalPaediatricsPhoton-counting computed tomographyRadiation dosage

Autor:innen und Institutionen

  • Diagnostic Radiology, Department of Clinical Sciences Lund, Lund University, Lund, Sweden.
  • Department of Radiology, Skåne University Hospital, S-22185, Lund, Sweden.
  • Medical Radiation Physics, Department of Clinical Sciences Malmö, Lund University, Lund, Sweden.
  • Radiation Physics, Department of Haematology, Oncology and Radiation Physics, Skåne University Hospital, Lund, Sweden.
  • Paediatric Cardiology, Department of Clinical Sciences Lund, Lund University, Lund, Sweden.
  • Department of Paediatric Cardiology, Skåne University Hospital, Lund, Sweden.

ORCID-Profile: Marie-Louise Aurumskjöld, Petru Liuba, Erik Hedström

Förderung

  • Hjärt-Lungfonden
  • the Swedish governmental funding of clinical research
  • Region Skåne
  • Skånes universitetssjukhus
  • Lund University

Zitieren

APA

Stålhammar, F., Aurumskjöld, M. L., Meyer, S., Wiklund, M., Wingren, P., Liuba, P., & Hedström, E. (2024). Photon-counting computed tomography for paediatric congenital heart defects yields images of high diagnostic quality with low radiation doses at both 70 kV and 90 kV. Pediatric Radiology. https://doi.org/10.1007/s00247-024-05939-z

Vancouver

Stålhammar F, Aurumskjöld ML, Meyer S, Wiklund M, Wingren P, Liuba P, et al. Photon-counting computed tomography for paediatric congenital heart defects yields images of high diagnostic quality with low radiation doses at both 70 kV and 90 kV. Pediatric Radiology. 2024. doi:10.1007/s00247-024-05939-z

BibTeX
@article{Stlhammar2024,
  author = {Stålhammar, Fredrik and Aurumskjöld, Marie-Louise and Meyer, Sofie and Wiklund, Marie and Wingren, Pär and Liuba, Petru and Hedström, Erik},
  title = {Photon-counting computed tomography for paediatric congenital heart defects yields images of high diagnostic quality with low radiation doses at both 70 kV and 90 kV},
  journal = {Pediatric Radiology},
  year = {2024},
  doi = {10.1007/s00247-024-05939-z}
}
Kostenlos starten →