Mobile head CT scanner provides superior image quality compared to conventional stationary CT

Henrik Andersson, Björn M Hansen, Emma Hall, Amir Hillal, Helena Mellander Oxholm, Veronica Fransson, Johan Wassélius

Neuroradiology · 30.06.2026

DOI: 10.1007/s00234-026-04083-9 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Neuroradiology, online erschienen am 30.06.2026. Verfasst von Henrik Andersson und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiology, Skåne University Hospital, Lund, Sweden. Die Arbeit nennt 20 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Lund University.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Introduction

We compared image quality of first- and second generation SOMATOM On.site mobile head CT with vendor-matched stationary head CT in neurosurgical ICU patients.

Methods

Retrospective paired study of 100 ICU patients, 50 mobile CT v1 and 50 mobile CT v2, paired with stationary CT, yielding 200 head CT exams. Four radiologists blinded to image characteristics scored five tasks on a four-point scale: supratentorial cortical and deep image quality, infratentorial image quality and artifact severity. Quantitative metrics were attenuation and noise from four standardized ROIs, and grey-white matter contrast-to-noise ratio.

Results

MobCT v2 received higher reader scores than paired stationary CT for overall image quality, supratentorial cortical and deep image quality, and equal to stationary CT for infratentorial image quality and artifact severity. Mobile CT v1 was rated lower than paired stationary CT, and lower than mobile CT v2, for all tasks. Attenuation, noise, and CNR were similar across groups. Radiation dose favoured mobile CT v2 with a median Dose Length Product (DLP) of 39.6 mGy·cm compared to 50.1 mGy·cm for mobile CTv1.

Conclusion

The second generation On.site mobile CT yielded superior image quality versus the first generation and comparable to contemporary stationary CT with similar quantitative metrics. These findings support a broader adoption of point-of-care head CT.

Schlagwörter

Stichwörter der Autor:innen

Image qualityMobile CTNon-contrast head CTObserver performancePoint-of-care systemsPortable CTTomography, X-Ray computed

Autor:innen und Institutionen

  • Department of Radiology, Skåne University Hospital, Lund, Sweden.
  • Stroke Imaging Research Group, Radiology, Lund University, Lund, Sweden.
  • Radiation Physics, Department of Hematology, Oncology and Radiation Physics, Skåne University Hospital, Lund, Sweden.
  • Medical Radiation Physics, Department of Translational Medicine, Lund University, Lund, Sweden.
  • Department of Radiology, Skåne University Hospital, Lund, Sweden. johan.wasselius@med.lu.se.
  • Stroke Imaging Research Group, Radiology, Lund University, Lund, Sweden. johan.wasselius@med.lu.se.

Förderung

  • Lund University

Zitieren

APA

Andersson, H., M Hansen, B., Hall, E., Hillal, A., Mellander Oxholm, H., Fransson, V., & Wassélius, J. (2026). Mobile head CT scanner provides superior image quality compared to conventional stationary CT. Neuroradiology. https://doi.org/10.1007/s00234-026-04083-9

Vancouver

Andersson H, M Hansen B, Hall E, Hillal A, Mellander Oxholm H, Fransson V, et al. Mobile head CT scanner provides superior image quality compared to conventional stationary CT. Neuroradiology. 2026. doi:10.1007/s00234-026-04083-9

BibTeX
@article{Andersson2026,
  author = {Andersson, Henrik and M Hansen, Björn and Hall, Emma and Hillal, Amir and Mellander Oxholm, Helena and Fransson, Veronica and Wassélius, Johan},
  title = {Mobile head CT scanner provides superior image quality compared to conventional stationary CT},
  journal = {Neuroradiology},
  year = {2026},
  doi = {10.1007/s00234-026-04083-9}
}
Kostenlos starten →