Deep learning image reconstruction technique for improving image quality and radiation dose reduction compared to iterative reconstruction technique in non-contrast CT head imaging
Frontiers in Radiology · 16.06.2026 · 6
DOI: 10.3389/fradi.2026.1824489 ↗Kurz gefasst
Fachartikel in der Zeitschrift Frontiers in Radiology, online erschienen am 16.06.2026. Verfasst von Obhuli Chandran M und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Medical Imaging Technology, Manipal College of Health Professions, Manipal Academy of Higher Education, Manipal, India. Die Arbeit nennt 28 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Introduction
Non-contrast CT (NCCT) head is a standard imaging modality for evaluating central nervous system pathologies. Repeated CT brain examinations are associated with a cumulative risk of cancer development. Deep learning image reconstruction (DLIR) techniques help improve image quality (IQ) while reducing radiation dose (RD) compared with iterative reconstruction techniques (iDose 4 ). Therefore, our study compared IQ parameters among low-dose (LD) CT with DLIR (Precise Image), LD CT with iterative reconstruction (iDose 4 ), and standard-dose (SD) CT with iDose 4 and further evaluated radiation dose reduction between SD-iDose 4 and LD-DLIR in NCCT head imaging.
Methods
Group A (SD with iDose 4 ) and Group B (LD with DLIR; Precise Image) each included 96 patients. All NCCT brain scans were performed using a 128-slice incisive CT scanner (Philips Healthcare Systems) with iDose 4 and DLIR. Qualitative and quantitative IQ analyses and radiation dose indices were compared between the groups. Lesion conspicuity was also assessed between LD-DLIR and LD-iDose 4 .
Results
There was a significant reduction in RD (effective dose: 1.01 vs. 2.4 mSv; p <0.05) with LD-DLIR. LD-DLIR demonstrated superior IQ compared with both SD-iDose 4 and LD-iDose 4 . Subjectively, gray–white matter differentiation improved from scores of 3 (SD-iDose 4 ) and 2.5 (LD-iDose 4 ) to 4.5 (LD-DLIR), while overall image quality and subjective image noise increased from 3–3.5 (SD-iDose 4 ) and 2.5–3 (LD-iDose 4 ) to 4–4.5 with LD-DLIR ( p < 0.05). Image noise was lowest with LD-DLIR in gray matter thalamus (GMT: 2.30–2.31) compared with SD-iDose 4 (4.80–4.85) and LD-iDose 4 (6.15–6.25); similar results were seen in the white matter posterior limb of the internal capsule (WMPIC: 1.98–2.21 vs. 4.36–4.38 vs. 5.19–5.32) and adjacent cortical gray matter (ACGM: 2.17–2.19 vs. 4.14–4.22 vs. 5.04–5.08). SNR was highest with LD-DLIR [GMT: 16.29–16.36 vs. 7.32–7.34 (SD-iDose 4 ) and 5.61–5.75 (LD-iDose 4 ); WMPIC: 14.69–14.75 vs. 6.49–6.51 vs. 5.18–5.26], and contrast-to-noise ratio (CNR) was also markedly improved [GMT–WMPIC: 2.59–2.61 (LD-DLIR) vs. 0.98–0.99 (SD-iDose 4 ) and 0.82–0.83 (LD-iDose 4 ); ACGM–FWM: 2.15–2.29 vs. 0.93–0.95 vs. 0.85–0.86]. Lesion conspicuity was superior with LD-DLIR for all brain lesions compared with LD-iDose 4 .
Conclusion
Our study demonstrated that LD NCCT head imaging with DLIR provides superior noise reduction and improved GWMD, SNR, and CNR compared with SD and LD with iDose 4 and supports the clinical utilization of DLIR.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Medical Imaging Technology, Manipal College of Health Professions, Manipal Academy of Higher Education, Manipal, India.
- Department of Radiodiagnosis and Imaging, Kasturba Medical College, Manipal Academy of Higher Education, Manipal, India.
- Philips Research and Development, Philips Innovation Campus, Yelahanka, Karnataka, India.
Zitieren
APA
Chandran M, O., Pendem, S., Kadavigere, R., Priyanka, Ps, P., & Chacko, C. (2026). Deep learning image reconstruction technique for improving image quality and radiation dose reduction compared to iterative reconstruction technique in non-contrast CT head imaging. Frontiers in Radiology, 6. https://doi.org/10.3389/fradi.2026.1824489
Vancouver
Chandran M O, Pendem S, Kadavigere R, Priyanka, Ps P, Chacko C. Deep learning image reconstruction technique for improving image quality and radiation dose reduction compared to iterative reconstruction technique in non-contrast CT head imaging. Frontiers in Radiology. 2026;6. doi:10.3389/fradi.2026.1824489
BibTeX
@article{ChandranM2026,
author = {Chandran M, Obhuli and Pendem, Saikiran and Kadavigere, Rajagopal and Priyanka and Ps, Priya and Chacko, Cijo},
title = {Deep learning image reconstruction technique for improving image quality and radiation dose reduction compared to iterative reconstruction technique in non-contrast CT head imaging},
journal = {Frontiers in Radiology},
year = {2026},
volume = {6},
doi = {10.3389/fradi.2026.1824489}
}
EUVIMED