Balancing Radiation Dose and Image Quality: Protocol Optimization for Mobile Head CT in Neurointensive Care Unit Patients
Diagnostics · 13.01.2026 · 16 (2) · 256
DOI: 10.3390/diagnostics16020256 ↗Kurz gefasst
Studienprotokoll in der Zeitschrift Diagnostics, online erschienen am 13.01.2026. Verfasst von Damian Mialkowskyj und 8 weiteren Autor:innen. Beteiligte Einrichtung: Institute for Diagnostic and Interventional Neuroradiology, University Hospital, LMU Munich, Marchioninistr. 15, 81377 Munich, Germany. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 32 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Objective
Mobile head CT enables bedside neuroimaging in critically ill patients, reducing risks associated with intrahospital transport. Despite increasing clinical use, evidence on dose optimization for mobile CT systems remains limited. This study evaluated whether an optimized CT protocol can reduce radiation exposure without compromising diagnostic image quality in neurointensive care unit patients.
Methods
In this retrospective single-center study, twenty-two non-contrast head CT examinations were acquired with a second-generation mobile CT scanner between March and May 2023. Patients underwent either a default (group A, n = 14; volumetric computed tomography dose index (CTDIvol) 44.1 mGy) or low-dose CT protocol (group B, n = 8; CTDIvol 32.1 mGy). Regarding dosimetry analysis, we recorded dose length product (DLP) and effective dose (ED). Quantitative image quality was assessed by manually placing ROIs at the basal ganglia and cerebellar levels to determine signal, noise, signal-to-noise ratio, and contrast-to-noise ratio. Two neuroradiologists independently rated qualitative image quality using a four-point Likert scale. Statistical comparisons were performed using a significance threshold of 0.05.
Results
Median DLP and ED were significantly lower for group B (592 mGy·cm, 1.12 mSv) than for group A (826 mGy·cm, 1.57 mSv; each p < 0.0001). Quantitative image quality parameters did not differ significantly between groups (p > 0.05). Qualitative image quality was rated excellent (median score 4).
Conclusions
The optimized mobile head CT protocol achieved a 28.7% reduction in radiation exposure while maintaining high diagnostic image quality. These findings support the adoption of low-dose strategies in mobile CT imaging in line with established radiation protection standards.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Institute for Diagnostic and Interventional Neuroradiology, University Hospital, LMU Munich, Marchioninistr. 15, 81377 Munich, Germany.
- Department of Neurology, University Hospital, LMU Munich, Marchioninistr. 15, 81377 Munich, Germany.
- Radiologie Augsburg Friedberg ÜBAG, Hermanstraße 15, 86150 Augsburg, Germany.
ORCID-Profile: Robert Stahl, Konstantinos Dimitriadis, Christoph G Trumm, Robert Forbrig
Zitieren
APA
Mialkowskyj, D., Stahl, R., Heck, S., Dimitriadis, K., Fischer, T. D., Liebig, T., Trumm, C. G., Wesemann, T., & Forbrig, R. (2026). Balancing Radiation Dose and Image Quality: Protocol Optimization for Mobile Head CT in Neurointensive Care Unit Patients. Diagnostics, 16(2), 256. https://doi.org/10.3390/diagnostics16020256
Vancouver
Mialkowskyj D, Stahl R, Heck S, Dimitriadis K, Fischer TD, Liebig T, et al. Balancing Radiation Dose and Image Quality: Protocol Optimization for Mobile Head CT in Neurointensive Care Unit Patients. Diagnostics. 2026;16(2):256. doi:10.3390/diagnostics16020256
BibTeX
@article{Mialkowskyj2026,
author = {Mialkowskyj, Damian and Stahl, Robert and Heck, Suzette and Dimitriadis, Konstantinos and Fischer, Thomas David and Liebig, Thomas and Trumm, Christoph G and Wesemann, Tim and Forbrig, Robert},
title = {Balancing Radiation Dose and Image Quality: Protocol Optimization for Mobile Head CT in Neurointensive Care Unit Patients},
journal = {Diagnostics},
year = {2026},
volume = {16},
number = {2},
pages = {256},
doi = {10.3390/diagnostics16020256}
}
EUVIMED