Accelerated flat panel computed tomography for pre-operative temporal bone imaging: Image quality and dosimetry comparison to conventional high resolution multislice computed tomography
Neuroradiology · 2025
DOI: 10.1007/s00234-025-03592-3 ↗Kurz gefasst
Vergleichsstudie in der Zeitschrift Neuroradiology, 2025. Verfasst von Elie Diamandis und 7 weiteren Autor:innen. Beteiligte Einrichtung: University Clinic for Neuroradiology, Otto-Von-Guericke-University Magdeburg, Magdeburg, Germany. elie.diamandis@med.ovgu.de. Hauptthemen laut MEDLINE: Temporal Bone, Multidetector Computed Tomography, Tomography, X-Ray Computed, X-Ray Intensifying Screens. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 31 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Otto-von-Guericke-Universität Magdeburg.
Abstract (Originalsprache)
Purpose
High-resolution multislice CT (HR-MSCT) and cone beam CT (CBCT) are commonly used for preoperative temporal bone imaging, with HR-MSCT often preferred due to its shorter scan duration and lower susceptibility to motion artifacts. However, recent advancements in accelerated flat panel CT (Acc-FPCT) available with the latest generation angiography systems have addressed traditional limitations of CBCT by significantly decreasing scan time. This cadaver-based study evaluates the diagnostic performance and radiation dose of Acc-FPCT compared to HR-MSCT in preoperative temporal bone imaging.
Methods
Six different Acc-FPCT protocols were acquired on five whole-head cadaveric specimens (ten temporal bones). Three neuroradiologists experienced in temporal bone imaging assessed the image quality of Acc-FPCT protocols in comparison to that of HR-MSCT for the visualization of 31 landmarks of middle and inner ear using a 5-point Likert scale. We also compared radiation dose parameters (CT dose index and dose length product) among the protocols.
Results
Two high-Resolution Acc-FPCT protocols were found to be superior to HR-MSCT by all raters ( p < 0.001). There were no significant differences between the two HR-FPCT protocols ( p = 0.25). The remaining Acc-FPCT protocols were rated significantly inferior to HR-MSCT. The inter-rater reliability was excellent (ICC (2,k) = 0.925; CI [0.92–0.93]). The dose length product was significantly lower in all Acc-FPCT protocols compared to HR-MSCT.
Conclusion
The results of our cadaver-based study highlight the utility of certain Acc-FPCT protocols as a viable alternative to HR-MSCT in preoperative temporal bone imaging, improving the visualization of critical anatomical landmarks without increasing radiation exposure.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- University Clinic for Neuroradiology, Otto-Von-Guericke-University Magdeburg, Magdeburg, Germany. elie.diamandis@med.ovgu.de.
- University Clinic for Neuroradiology, Otto-Von-Guericke-University Magdeburg, Magdeburg, Germany.
- German Center for Neurodegenerative Diseases (DZNE), Magdeburg, Germany.
- University Clinic of Otolaryngology, Head and Neck Surgery, Otto-Von-Guericke-University Magdeburg, Magdeburg, Germany.
- Institute of Anatomy, Otto-Von-Guericke-University Magdeburg, Magdeburg, Germany.
Förderung
- Otto-von-Guericke-Universität Magdeburg
Zitieren
APA
Diamandis, E., Müller, S. J., Khadhraoui, E., Klebingat, S., Einspänner, E., Durisin, M., Albrecht, A., & Behme, D. (2025). Accelerated flat panel computed tomography for pre-operative temporal bone imaging: Image quality and dosimetry comparison to conventional high resolution multislice computed tomography. Neuroradiology. https://doi.org/10.1007/s00234-025-03592-3
Vancouver
Diamandis E, Müller SJ, Khadhraoui E, Klebingat S, Einspänner E, Durisin M, et al. Accelerated flat panel computed tomography for pre-operative temporal bone imaging: Image quality and dosimetry comparison to conventional high resolution multislice computed tomography. Neuroradiology. 2025. doi:10.1007/s00234-025-03592-3
BibTeX
@article{Diamandis2025,
author = {Diamandis, Elie and Müller, Sebastian Johannes and Khadhraoui, Eya and Klebingat, Stefan and Einspänner, Eric and Durisin, Martin and Albrecht, Anne and Behme, Daniel},
title = {Accelerated flat panel computed tomography for pre-operative temporal bone imaging: Image quality and dosimetry comparison to conventional high resolution multislice computed tomography},
journal = {Neuroradiology},
year = {2025},
doi = {10.1007/s00234-025-03592-3}
}