Hybrid Parallel‐Serial Tool Exchange Robot for Ophthalmic Precision Surgery: Mechanical Design
The International Journal of Medical Robotics and Computer Assisted Surgery · 2026
DOI: 10.1002/rcs.70208 ↗Kurz gefasst
Fachartikel in der Zeitschrift The International Journal of Medical Robotics and Computer Assisted Surgery, 2026. Verfasst von Yu-Hsiu Lee und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Mechanical Engineering, National Taiwan University, Taipei, Taiwan. Hauptthemen laut MEDLINE: Robotic Surgical Procedures, Ophthalmologic Surgical Procedures. Die Arbeit nennt 30 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: National Science and Technology Council, National Taiwan University.
Abstract (Originalsprache)
Background
Robot‐assisted ophthalmic surgery requires precise manipulation about a remote centre of motion (RCM). Conventional systems often employ mechanically fixed RCMs, limiting alignment flexibility, motion compensation, and tool exchange capability. This work presents HYPER‐OPS, a hybrid parallel‐serial ophthalmic robot featuring a software‐controlled RCM and tool exchange capability.
Methods
A compact six‐degree‐of‐freedom manipulator was developed using a dual‐layer five‐bar linkage. The system features software‐controlled RCM functionality and passive magnetic tool exchange. Kinematic, workspace and structural analyses were performed, followed by experimental validation.
Results
The robot achieved the workspace required for ophthalmic procedures while enabling configurable RCM placement. Finite element analysis predicted a maximum deformation of 0.045 mm and a first natural frequency of 56 Hz. Experiments demonstrated RCM position errors below 0.2 mm and tool exchange repeatability of 0.39 mm.
Conclusion
HYPER‐OPS provides a compact platform for ophthalmic surgery with improved flexibility for alignment, motion compensation and workflow automation.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Mechanical Engineering, National Taiwan University, Taipei, Taiwan.
ORCID-Profile: Yu-Hsiu Lee
Förderung
- National Science and Technology Council
- National Taiwan University
Zitieren
APA
Lee, Y. H., Wu, Y. C., Chen, Y. A., Chou, C., & Lin, C. A. (2026). Hybrid Parallel‐Serial Tool Exchange Robot for Ophthalmic Precision Surgery: Mechanical Design. The International Journal of Medical Robotics and Computer Assisted Surgery. https://doi.org/10.1002/rcs.70208
Vancouver
Lee YH, Wu YC, Chen YA, Chou C, Lin CA. Hybrid Parallel‐Serial Tool Exchange Robot for Ophthalmic Precision Surgery: Mechanical Design. The International Journal of Medical Robotics and Computer Assisted Surgery. 2026. doi:10.1002/rcs.70208
BibTeX
@article{Lee2026,
author = {Lee, Yu-Hsiu and Wu, Yung-Chen and Chen, Yu-An and Chou, Ching and Lin, Che-An},
title = {Hybrid Parallel‐Serial Tool Exchange Robot for Ophthalmic Precision Surgery: Mechanical Design},
journal = {The International Journal of Medical Robotics and Computer Assisted Surgery},
year = {2026},
doi = {10.1002/rcs.70208}
}