A Hybrid Compliant–Tendon Laparoscopic Instrument for Distal Articulation and Grasping: Design and Experimental Validation
The International Journal of Medical Robotics and Computer Assisted Surgery · 2026
DOI: 10.1002/rcs.70206 ↗Kurz gefasst
Fachartikel in der Zeitschrift The International Journal of Medical Robotics and Computer Assisted Surgery, 2026. Verfasst von Prabhat Kumar und 1 weiteren Autor:innen. Beteiligte Einrichtung: Department of Neurological Surgery, University of California, San Francisco, California, USA. Hauptthemen laut MEDLINE: Tendons, Laparoscopy, Robotic Surgical Procedures, Laparoscopes. Die Arbeit nennt 16 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0).
Abstract (Originalsprache)
Background
Conventional laparoscopic instruments provide limited dexterity in confined anatomical environments, whereas existing articulated designs rely on complex rigid joints or motorised actuation, increasing mechanical complexity and limiting miniaturisation.
Methods
A hybrid compliant–tendon laparoscopic instrument was developed to integrate articulated distal motion and active grasping within a 5 mm form factor. The compliant segment was fabricated from super‐elastic Nitinol, and grasping was achieved using a tendon–pulley mechanism. The system was evaluated using nonlinear finite element analysis and experimental testing, including digital image correlation and benchtop force measurements.
Results
The mechanism achieved up to 75° articulation with mean model errors below 8%. The grasper produced a maximum force of 13.2 ± 0.47 N and maintained performance across articulation angles. Pull‐out tests confirmed stable interaction with soft‐tissue analogues.
Conclusions
The proposed design demonstrates a compact, low‐complexity approach for combining compliant articulation and tendon‐driven grasping within a 5 mm surgical instrument.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Neurological Surgery, University of California, San Francisco, California, USA.
- Mechanical Engineering Department, Indian Institute of Technology, Bombay, India.
ORCID-Profile: Prabhat Kumar
Zitieren
APA
Kumar, P., & Ravi, B. (2026). A Hybrid Compliant–Tendon Laparoscopic Instrument for Distal Articulation and Grasping: Design and Experimental Validation. The International Journal of Medical Robotics and Computer Assisted Surgery. https://doi.org/10.1002/rcs.70206
Vancouver
Kumar P, Ravi B. A Hybrid Compliant–Tendon Laparoscopic Instrument for Distal Articulation and Grasping: Design and Experimental Validation. The International Journal of Medical Robotics and Computer Assisted Surgery. 2026. doi:10.1002/rcs.70206
BibTeX
@article{Kumar2026,
author = {Kumar, Prabhat and Ravi, Bhallamudi},
title = {A Hybrid Compliant–Tendon Laparoscopic Instrument for Distal Articulation and Grasping: Design and Experimental Validation},
journal = {The International Journal of Medical Robotics and Computer Assisted Surgery},
year = {2026},
doi = {10.1002/rcs.70206}
}