FPGA‐Based Low‐Latency Semantic Feedback System for Real‐Time Instrument Localization in Telesurgery
The International Journal of Medical Robotics and Computer Assisted Surgery · 2026-06 · 22 (3)
DOI: 10.1002/rcs.70180 ↗Kurz gefasst
Fachartikel in der Zeitschrift The International Journal of Medical Robotics and Computer Assisted Surgery, online erschienen am 20.05.2026. Verfasst von Zhikang Ma und 8 weiteren Autor:innen. Beteiligte Einrichtung: School of Mechanical Engineering, Tianjin University, Tianjin, China. Hauptthemen laut MEDLINE: Robotic Surgical Procedures, Surgery, Computer-Assisted. Die Arbeit nennt 27 Literaturangaben. Gefördert von: National Natural Science Foundation of China, Research Project on Teaching Reform in Ordinary Undergraduate Universities in Hunan Province, Graduate Education and Teaching Reform Project of Central South University.
Abstract (Originalsprache)
Background
Telesurgery is limited by visual latency, compromising master–slave synchronization and safety. This study proposes a real‐time semantic feedback framework that reduces latency while maintaining millimetre‐level localization accuracy.
Methods
We developed a low‐latency semantic feedback system for real‐time instrument localization using a custom visual marker and FPGA‐based extraction. The system was evaluated through phantom experiments, accuracy benchmarking against the NDI system, latency measurement, and task‐based evaluation under delayed visual feedback conditions.
Results
The system achieved a 99.2% marker extraction success rate, with a maximum mean error of 0.68 0.72 mm and an end‐to‐end latency of 31.2 1.4 ms. Task‐based evaluation further showed significantly reduced task completion time under delayed visual feedback.
Conclusions
The proposed system provides robust, precise, and low‐latency semantic feedback and demonstrates potential to mitigate the impact of visual latency on surgical performance, supporting its application in future telesurgical systems pending further in vivo validation.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- School of Mechanical Engineering, Tianjin University, Tianjin, China.
- Institute of Medical Robotics and Intelligent Systems, Tianjin University, Tianjin, China.
- School of Mechanical Engineering, The Open University of China, Beijing, China.
- Department of General Surgery, The Third Xiangya Hospital, Central South University, Hunan, China.
ORCID-Profile: Jianchang Zhao, Jinhua Li
Förderung
- National Natural Science Foundation of China
- Research Project on Teaching Reform in Ordinary Undergraduate Universities in Hunan Province
- Graduate Education and Teaching Reform Project of Central South University
- Emerging Frontiers Cultivation Program of Tianjin University Interdisciplinary Center
Zitieren
APA
Ma, Z., Sun, X., Yi, B., Wang, J., Zhang, F., Wang, S., Pan, L., Zhao, J., & Li, J. (2026). FPGA‐Based Low‐Latency Semantic Feedback System for Real‐Time Instrument Localization in Telesurgery. The International Journal of Medical Robotics and Computer Assisted Surgery, 22(3). https://doi.org/10.1002/rcs.70180
Vancouver
Ma Z, Sun X, Yi B, Wang J, Zhang F, Wang S, et al. FPGA‐Based Low‐Latency Semantic Feedback System for Real‐Time Instrument Localization in Telesurgery. The International Journal of Medical Robotics and Computer Assisted Surgery. 2026;22(3). doi:10.1002/rcs.70180
BibTeX
@article{Ma2026,
author = {Ma, Zhikang and Sun, Xinan and Yi, Bo and Wang, Jiale and Zhang, Fengpu and Wang, Shuxin and Pan, Lizhi and Zhao, Jianchang and Li, Jinhua},
title = {FPGA‐Based Low‐Latency Semantic Feedback System for Real‐Time Instrument Localization in Telesurgery},
journal = {The International Journal of Medical Robotics and Computer Assisted Surgery},
year = {2026},
volume = {22},
number = {3},
doi = {10.1002/rcs.70180}
}