Novel Design of a Surgical Tool Insertion Robot Using a Chebyshev Lambda Mechanism

SeongHyeon Won iD, Chanwoo Kim, Yeonho Ko, Junho Hong iD, Jisu Jeon, Daehie Hong iD

The International Journal of Medical Robotics and Computer Assisted Surgery · 2025

DOI: 10.1002/rcs.70095 ↗

Kurz gefasst

Fachartikel in der Zeitschrift The International Journal of Medical Robotics and Computer Assisted Surgery, 2025. Verfasst von SeongHyeon Won und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Mechanical Engineering, Korea University, Seoul, South Korea. Hauptthemen laut MEDLINE: Robotic Surgical Procedures, Robotics. Die Arbeit nennt 29 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0). Gefördert von: Ministry of Health & Welfare, Republic of Korea, National Research Foundation of Korea.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Endoscopic Retrograde Cholangiopancreatography (ERCP) and Percutaneous Coronary Intervention (PCI) are minimally invasive procedures that treat diseases with minimal scarring and rapid recovery. However, catheterisation under fluoroscopy exposes medical staff to radiation, and navigating flexible guidewires through complex anatomical pathways requires precise control.

Methods

To address these challenges, we propose a novel insertion robot based on a Chebyshev lambda mechanism that makes a linear grasp path and an arc‐shaped quick return path using a single motor. An integrated roll module enables continuous insertion and rotation, yielding a robot with two degrees of freedom.

Results

Experiments show robots can make sufficient insertion force for real procedures, with robot's maximum mean translational and rotational errors of 0.11 mm and 1.47°, respectively. Phantom model tests show its applicability to human anatomy.

Conclusions

The proposed robotic system demonstrated sufficient precision and force for real surgical applications, indicating its potential for clinical use in minimally invasive procedures.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Robotic Surgical ProceduresRoboticsHumansEquipment DesignPhantoms, ImagingMinimally Invasive Surgical ProceduresPercutaneous Coronary InterventionFluoroscopySurgery, Computer-Assisted

Stichwörter der Autor:innen

cardiovascular robotsmechanismrobot‐assisted catheterisation systemrobot‐assisted surgery

Autor:innen und Institutionen

  • Department of Mechanical Engineering, Korea University, Seoul, South Korea.
  • Department of Mechanical Engineering, KAIST, Daejeon, South Korea.

ORCID-Profile: SeongHyeon Won, Junho Hong, Daehie Hong

Förderung

  • Ministry of Health & Welfare, Republic of Korea
  • National Research Foundation of Korea

Zitieren

APA

Won, S., Kim, C., Ko, Y., Hong, J., Jeon, J., & Hong, D. (2025). Novel Design of a Surgical Tool Insertion Robot Using a Chebyshev Lambda Mechanism. The International Journal of Medical Robotics and Computer Assisted Surgery. https://doi.org/10.1002/rcs.70095

Vancouver

Won S, Kim C, Ko Y, Hong J, Jeon J, Hong D. Novel Design of a Surgical Tool Insertion Robot Using a Chebyshev Lambda Mechanism. The International Journal of Medical Robotics and Computer Assisted Surgery. 2025. doi:10.1002/rcs.70095

BibTeX
@article{Won2025,
  author = {Won, SeongHyeon and Kim, Chanwoo and Ko, Yeonho and Hong, Junho and Jeon, Jisu and Hong, Daehie},
  title = {Novel Design of a Surgical Tool Insertion Robot Using a Chebyshev Lambda Mechanism},
  journal = {The International Journal of Medical Robotics and Computer Assisted Surgery},
  year = {2025},
  doi = {10.1002/rcs.70095}
}
Kostenlos starten →