Ultrahigh‐resolution K‐edge imaging for guidance of microrobots in minimally invasive interventions: A feasibility study
Medical Physics · 2025
DOI: 10.1002/mp.70105 ↗Kurz gefasst
Fachartikel in der Zeitschrift Medical Physics, 2025. Verfasst von Arnaud R Brian-Choux und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Biomedical Engineering, Georgia Institute of Technology, Department of Radiology and Imaging Sciences, Winship Cancer Institute, Emory …. Hauptthemen laut MEDLINE: Tomography, X-Ray Computed, Minimally Invasive Surgical Procedures, Robotics, Robotic Surgical Procedures. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 17 Literaturangaben.
Abstract (Originalsprache)
Background
Photon‐counting detector (PCD) CT offers enhanced spatial resolution, improved image contrast, reduced radiation dose, and material differentiation through K‐edge imaging. These features may be of value for guiding microrobots and surgical instruments during minimally invasive brain surgery.
Purpose
This study evaluates the feasibility of using K‐edge imaging with deep silicon (dSi) PCD CT to estimate the pose and location of microrobots within the human head, utilizing materials like neodymium (Nd) and tungsten (W) with distinct K‐edge energies.
Methods
A micro‐driller robot with a cubic Nd magnet (K‐edge energy = 43.5 keV) and a 3D‐printed drill bit was used. W (K‐edge energy = 69.5 keV) was added to aid in orientation detection. These materials were positioned at distances of 0–1 mm within a 3D‐printed setup and placed in a human skull filled with ballistic gel, alongside other metallic parts. The setup was scanned using a prototype dSi CT scanner with eight energy bins. Scanning parameters included 120 kVp, 300 mAs, 0.5 s rotation time, and 4‐cm z‐collimation.
Results
Nd and W components were distinguishable, with Nd showing higher contrast in low‐energy virtual monoenergetic images (VMIs) and W being more radiopaque in higher VMIs. The signal‐to‐noise ratios (SNR) at 70 keV and 1‐mm distance were 31.0 for Nd and 132.7 for W. The error in estimating the distance between components (Δ L ) ranged from 82 to 263 µm comparable to one image voxel.
Conclusions
Deep silicon PCD‐CT with K‐edge imaging successfully differentiated microrobot components in a human head phantom, indicating its potential for precision image‐guided robotic surgery. Further studies are needed to optimize radiation doses for clinical use.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Biomedical Engineering, Georgia Institute of Technology, Department of Radiology and Imaging Sciences, Winship Cancer Institute, Emory University, Atlanta, Georgia, USA.
- CT Research, GE HealthCare, Niskayuna, New York, USA.
- Department of Medical Physics, University of Wisconsin, Madison, Wisconsin, USA.
Zitieren
APA
Brian-Choux, A. R., Holmes, T. W., Yin, Z., Hoang, K. B., Salyapongse, A., Ansari, A., & Pourmorteza, A. (2025). Ultrahigh‐resolution K‐edge imaging for guidance of microrobots in minimally invasive interventions: A feasibility study. Medical Physics. https://doi.org/10.1002/mp.70105
Vancouver
Brian-Choux AR, Holmes TW, Yin Z, Hoang KB, Salyapongse A, Ansari A, et al. Ultrahigh‐resolution K‐edge imaging for guidance of microrobots in minimally invasive interventions: A feasibility study. Medical Physics. 2025. doi:10.1002/mp.70105
BibTeX
@article{BrianChoux2025,
author = {Brian-Choux, Arnaud R and Holmes, Thomas Wesley and Yin, Zhye and Hoang, Kimberly B and Salyapongse, Aria and Ansari, Azadeh and Pourmorteza, Amir},
title = {Ultrahigh‐resolution K‐edge imaging for guidance of microrobots in minimally invasive interventions: A feasibility study},
journal = {Medical Physics},
year = {2025},
doi = {10.1002/mp.70105}
}
EUVIMED