Patient‐specific synthetic magnetic resonance imaging generation from cone beam computed tomography for image guidance in liver stereotactic body radiation therapy
Precision Radiation Oncology · 2022-06 · 6 (2) · 110-118
DOI: 10.1002/pro6.1163 ↗Kurz gefasst
Fachartikel in der Zeitschrift Precision Radiation Oncology, online erschienen am 11.06.2022. Verfasst von Zeyu Zhang und 5 weiteren Autor:innen. Beteiligte Einrichtung: Duke University Medical Center, Durham, North Carolina, USA. Die Arbeit wurde bisher 10-mal zitiert (Crossref) und nennt 21 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0). Gefördert von: NCI NIH HHS, NIBIB NIH HHS.
Abstract (Originalsprache)
Objective
Despite its prevalence, cone beam computed tomography (CBCT) has poor soft‐tissue contrast, making it challenging to localize liver tumors. We propose a patient‐specific deep learning model to generate synthetic magnetic resonance imaging (MRI) from CBCT to improve tumor localization.
Methods
A key innovation is using patient‐specific CBCT‐MRI image pairs to train a deep learning model to generate synthetic MRI from CBCT. Specifically, patient planning CT was deformably registered to prior MRI, and then used to simulate CBCT with simulated projections and Feldkamp, Davis, and Kress reconstruction. These CBCT‐MRI images were augmented using translations and rotations to generate enough patient‐specific training data. A U‐Net‐based deep learning model was developed and trained to generate synthetic MRI from CBCT in the liver, and then tested on a different CBCT dataset. Synthetic MRIs were quantitatively evaluated against ground‐truth MRI.
Results
The synthetic MRI demonstrated superb soft‐tissue contrast with clear tumor visualization. On average, the synthetic MRI achieved 28.01, 0.025, and 0.929 for peak signal‐to‐noise ratio, mean square error, and structural similarity index, respectively, outperforming CBCT images. The model performance was consistent across all three patients tested.
Conclusion
Our study demonstrated the feasibility of a patient‐specific model to generate synthetic MRI from CBCT for liver tumor localization, opening up a potential to democratize MRI guidance in clinics with conventional LINACs.
Autor:innen und Institutionen
- Duke University Medical Center, Durham, North Carolina, USA.
- Medical College of Wisconsin, Milwaukee, Wisconsin, USA.
- University of Maryland School of Medicine, Baltimore, Maryland, USA.
Förderung
- NCI NIH HHS
- NIBIB NIH HHS
Zitieren
APA
Zhang, Z., Jiang, Z., Zhong, H., Lu, K., Yin, F. F., & Ren, L. (2022). Patient‐specific synthetic magnetic resonance imaging generation from cone beam computed tomography for image guidance in liver stereotactic body radiation therapy. Precision Radiation Oncology, 6(2), 110-118. https://doi.org/10.1002/pro6.1163
Vancouver
Zhang Z, Jiang Z, Zhong H, Lu K, Yin FF, Ren L. Patient‐specific synthetic magnetic resonance imaging generation from cone beam computed tomography for image guidance in liver stereotactic body radiation therapy. Precision Radiation Oncology. 2022;6(2):110-118. doi:10.1002/pro6.1163
BibTeX
@article{Zhang2022,
author = {Zhang, Zeyu and Jiang, Zhuoran and Zhong, Hualiang and Lu, Ke and Yin, Fang-Fang and Ren, Lei},
title = {Patient‐specific synthetic magnetic resonance imaging generation from cone beam computed tomography for image guidance in liver stereotactic body radiation therapy},
journal = {Precision Radiation Oncology},
year = {2022},
volume = {6},
number = {2},
pages = {110--118},
doi = {10.1002/pro6.1163}
}
EUVIMED