EUVIMEDDescobrir EUVIMED
VergleichsstudieOpen access · CC BY 4.0

Comparative Analysis of Analog and Digital Chest Tube Drainage Systems Using a High‐Fidelity 3D‐Printed Neonatal Chest Model

F-X Anzinger, T J Hashagen, P Palaniappan, A Lindner, M Riboldi, J Gödeke, O J Muensterer

Pediatric Pulmonology · 2025-04 · 60 (4)

DOI: 10.1002/ppul.71090 ↗

At a glance

Vergleichsstudie in Pediatric Pulmonology, published online on 2025-04-11. Written by F-X Anzinger and 6 co-authors. Institution: Department of Pediatric Surgery, Ludwig-Maximilians-University of Munich | LMU, Munich, Germany. Major MEDLINE topics: Printing, Three-Dimensional, Chest Tubes, Drainage, Models, Anatomic. Cited 2 times so far (Crossref); 21 references. The full text is freely available (open access, CC BY 4.0). Funded by: The authors received no specific funding for this work..

ResumoSubjectsDados bibliográficosCiteRelated studies

Abstract (original language)

Background

In recent years, digital chest tube drainage systems have been introduced. Limited studies address their benefits and risks in pediatric patients, particularly neonates. This study compares a three‐chamber chest drainage system with a digital system using a high‐fidelity 3D‐printed model.

Methods

We conducted direct measurements and 3D‐printed model tests with both systems at different suction pressures (−1 to −20 cmH 2 O) to assess the actual pressures. The effects of siphon and automatic flushes in the digital system were also studied.

Results

At −20 and −10 cmH 2 O, significant differences were found between the digital and analog systems in direct and model measurements. The analog system became unreliable below −10 cmH 2 O. For the digital system, most measurements remained within the set pressures, with outliers up to −30 cmH 2 O due to regular flushing.

Conclusion

This experimental study evaluates the suitability of digital drainage systems for the pediatric and neonatal populations. Our model demonstrated reliable simulation of thoracic conditions, making it a useful tool for pre‐clinical testing where patient testing may be limited. Both systems yielded satisfactory results at −20 and −10 cmH 2 O, but the digital system showed greater flexibility, maintaining pressures as low as −5 cmH 2 O. The analog system was consistent but less adaptable, which may limit its use in dynamic situations. The digital system's ability to simulate more flexible scenarios offers potential clinical advantages, though further investigation is needed to assess its impact on neonatal safety. The increase in suction during flushing may pose a risk for neonatal patients.

Subjects

MeSH terms (MEDLINE)

Printing, Three-DimensionalChest TubesDrainageModels, AnatomicThoraxHumansInfant, NewbornEquipment DesignPressure

Author keywords

3D printinganalog drainage systemchest modelchest tube drainage systemdigital drainage system

Authors and institutions

  • Department of Pediatric Surgery, Ludwig-Maximilians-University of Munich | LMU, Munich, Germany.
  • Ludwig-Maximilians-University of Munich | LMU, Chair for Experimental Physics, Munich, Germany.

Funding

  • The authors received no specific funding for this work.

Cite

APA

Anzinger, F. X., Hashagen, T. J., Palaniappan, P., Lindner, A., Riboldi, M., Gödeke, J., & Muensterer, O. J. (2025). Comparative Analysis of Analog and Digital Chest Tube Drainage Systems Using a High‐Fidelity 3D‐Printed Neonatal Chest Model. Pediatric Pulmonology, 60(4). https://doi.org/10.1002/ppul.71090

Vancouver

Anzinger FX, Hashagen TJ, Palaniappan P, Lindner A, Riboldi M, Gödeke J, et al. Comparative Analysis of Analog and Digital Chest Tube Drainage Systems Using a High‐Fidelity 3D‐Printed Neonatal Chest Model. Pediatric Pulmonology. 2025;60(4). doi:10.1002/ppul.71090

BibTeX
@article{Anzinger2025,
  author = {Anzinger, F-X and Hashagen, T J and Palaniappan, P and Lindner, A and Riboldi, M and Gödeke, J and Muensterer, O J},
  title = {Comparative Analysis of Analog and Digital Chest Tube Drainage Systems Using a High‐Fidelity 3D‐Printed Neonatal Chest Model},
  journal = {Pediatric Pulmonology},
  year = {2025},
  volume = {60},
  number = {4},
  doi = {10.1002/ppul.71090}
}
Start for free →