Validierungsstudie in der Zeitschrift Journal of Medical Systems, online erschienen am 03.01.2025. Verfasst von Jesse A M van Doormaal und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Neurosurgery, University Medical Centre Utrecht, Utrecht, The Netherlands. j.a.m.vandoormaal-2@umcutrecht.nl. Hauptthemen laut MEDLINE: Phantoms, Imaging. Die Arbeit wurde bisher 6-mal zitiert (Crossref) und nennt 29 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
This study aimed to develop and validate a cost-effective, customizable patient-specific phantom for simulating external ventricular drain placement, combining image segmentation, 3-D printing and molding techniques. Two variations of the phantom were created based on patient MRI data, integrating a realistic skin layer with anatomical landmarks, a 3-D printed skull, an agarose polysaccharide gel brain, and a ventricular cavity. To validate the phantom, 15 neurosurgeons, residents, and physician assistants performed 30 EVD placements. The effectiveness of the phantom as a training tool was assessed through a standardized user experience questionnaire, which evaluated the physical attributes, realism, and overall satisfaction. The mechanical properties of the phantom brain were quantified by measuring catheter insertion forces using a linear force tester to compare them to those experienced in real brain tissue. The study participants successfully completed EVD placements with a 76.7% optimal placement rate, which aligns with rates observed in clinical practice. Feedback highlighted the anatomical accuracy of the phantom and its value in enhancing surgical skills, though it also identified areas for improvement, particularly in the realism of the skin layer. Mechanical testing demonstrated that the insertion forces required were comparable to those encountered in actual brain tissue. The developed phantom offers a realistic, low-cost, and adaptable model for EVD simulation. This tool is particularly beneficial for both training and research, with future enhancements planned to improve the realism of the skin and incorporate more anatomical features to increase the fidelity of the simulation.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Neurosurgery, University Medical Centre Utrecht, Utrecht, The Netherlands. j.a.m.vandoormaal-2@umcutrecht.nl.
- Department of Neurosurgery, University Medical Centre Utrecht, Utrecht, The Netherlands.
- Department of Medical Technology and Clinical Physics, University Medical Centre Utrecht, Utrecht, The Netherlands.
- Department of Neuro-Oncology, Princess Máxima Centre for Pediatric Oncology, Utrecht, The Netherlands.
- Department of Neurosurgery and Klinisches Neurozentrum, Universitätsspital Zürich, Zurich, Switzerland.
Zitieren
APA
van Doormaal, J. A. M., Fick, T., Boskovic, E., Hoving, E. W., Robe, P. A. J. T., & van Doormaal, T. P. C. (2025). Development and Validation of a Neurosurgical Phantom for Simulating External Ventricular Drain Placement. Journal of Medical Systems, 49(1). https://doi.org/10.1007/s10916-024-02133-4
Vancouver
van Doormaal JAM, Fick T, Boskovic E, Hoving EW, Robe PAJT, van Doormaal TPC. Development and Validation of a Neurosurgical Phantom for Simulating External Ventricular Drain Placement. Journal of Medical Systems. 2025;49(1). doi:10.1007/s10916-024-02133-4
BibTeX
@article{vanDoormaal2025,
author = {van Doormaal, Jesse A M and Fick, Tim and Boskovic, Ernest and Hoving, Eelco W and Robe, Pierre A J T and van Doormaal, Tristan P C},
title = {Development and Validation of a Neurosurgical Phantom for Simulating External Ventricular Drain Placement},
journal = {Journal of Medical Systems},
year = {2025},
volume = {49},
number = {1},
doi = {10.1007/s10916-024-02133-4}
}
EUVIMED